目的 探讨国产回弹式眼压计临床应用的可靠性、稳定性及对特殊病例测量的优越性.方法 分别使用国产SW-500 SUOER回弹式眼压计、ICARE回弹式眼压计及非接触眼压计(NCT)测量眼科门诊就诊124例患者(124只眼)眼压,并对三者测量结果的相关性进行对比分析.结果 SUOER、ICARE、NCT三种眼压计测量的均值分别为:(16.82±4.46) mm Hg、(16.90±5.50) mm Hg、(16.36±5.16) mm Hg,SUOER与ICARE测量结果相比较差异无统计学意义(P=0.67),NCT与其他两种眼压计相比差异均有统计学意义.三种眼压计测量结果彼此之间具有良好的相关性分别为0.917(NCT-ICARE)、0.927 (NCT-SUOER)、0.944 (ICARE-SUOER).不同眼压区间三种测量值在中眼压组中两两之间的差值较小,在低眼压组与高眼压组中差值较大.不同操作者使用SUOER回弹式眼压计测量结果差异无统计学意义.SUOER回弹式眼压计所需检查时间平均为6.12 s,小于NCT平均测量时间10.37秒,77%的被检查者认为其舒适度优于NCT.结论 国产SUOER回弹式眼压计与ICARE回弹式眼压计一样能方便准确舒适地应用于各种临床患者,尤其对于测量婴幼儿等不能固视、配合不良及瘫痪不能坐位测量的患者有一定的优越性.
目的 探討國產迴彈式眼壓計臨床應用的可靠性、穩定性及對特殊病例測量的優越性.方法 分彆使用國產SW-500 SUOER迴彈式眼壓計、ICARE迴彈式眼壓計及非接觸眼壓計(NCT)測量眼科門診就診124例患者(124隻眼)眼壓,併對三者測量結果的相關性進行對比分析.結果 SUOER、ICARE、NCT三種眼壓計測量的均值分彆為:(16.82±4.46) mm Hg、(16.90±5.50) mm Hg、(16.36±5.16) mm Hg,SUOER與ICARE測量結果相比較差異無統計學意義(P=0.67),NCT與其他兩種眼壓計相比差異均有統計學意義.三種眼壓計測量結果彼此之間具有良好的相關性分彆為0.917(NCT-ICARE)、0.927 (NCT-SUOER)、0.944 (ICARE-SUOER).不同眼壓區間三種測量值在中眼壓組中兩兩之間的差值較小,在低眼壓組與高眼壓組中差值較大.不同操作者使用SUOER迴彈式眼壓計測量結果差異無統計學意義.SUOER迴彈式眼壓計所需檢查時間平均為6.12 s,小于NCT平均測量時間10.37秒,77%的被檢查者認為其舒適度優于NCT.結論 國產SUOER迴彈式眼壓計與ICARE迴彈式眼壓計一樣能方便準確舒適地應用于各種臨床患者,尤其對于測量嬰幼兒等不能固視、配閤不良及癱瘓不能坐位測量的患者有一定的優越性.
목적 탐토국산회탄식안압계림상응용적가고성、은정성급대특수병례측량적우월성.방법 분별사용국산SW-500 SUOER회탄식안압계、ICARE회탄식안압계급비접촉안압계(NCT)측량안과문진취진124례환자(124지안)안압,병대삼자측량결과적상관성진행대비분석.결과 SUOER、ICARE、NCT삼충안압계측량적균치분별위:(16.82±4.46) mm Hg、(16.90±5.50) mm Hg、(16.36±5.16) mm Hg,SUOER여ICARE측량결과상비교차이무통계학의의(P=0.67),NCT여기타량충안압계상비차이균유통계학의의.삼충안압계측량결과피차지간구유량호적상관성분별위0.917(NCT-ICARE)、0.927 (NCT-SUOER)、0.944 (ICARE-SUOER).불동안압구간삼충측량치재중안압조중량량지간적차치교소,재저안압조여고안압조중차치교대.불동조작자사용SUOER회탄식안압계측량결과차이무통계학의의.SUOER회탄식안압계소수검사시간평균위6.12 s,소우NCT평균측량시간10.37초,77%적피검사자인위기서괄도우우NCT.결론 국산SUOER회탄식안압계여ICARE회탄식안압계일양능방편준학서괄지응용우각충림상환자,우기대우측량영유인등불능고시、배합불량급탄탄불능좌위측량적환자유일정적우월성.
Objective To approach the confidence,stability and superiority of SUOER rebound tonometer. Methods Intraocular pressure (IOP) measurements were performed on 124 cases (124 eyes) with SUOER rebound tonometer,ICARE rebound tonometer and NCT tonometer respectively.Correlation of the results measured by three types of tonometer was analyzed. Results The mean IOP measured by SUOER,ICARE,and NCT were 16.82±4.46mmHg,16.90±5.50mmHg,16.36±5.16mmHg,respectively.No significant difference was found between the result of SUOER and ICARE,and there was a significant difference between the results of NCT and other methods.Significant correlation could be found among all three methods with the correlation coefficient of 0.917 (NCT-ICARE),0.927(NCT-SUOER),0.944 (ICARE-SUOER).The differences of IOP among three types of tonometers were smaller in medium IOP group compared with high and low IOP group.No significant inter-observer difference was found by SUOER tonometer.The average examine duration was 6.12 seconds by SUOER tonometer,which was shorter than that of NCT (10.37 seconds).77% patients felt more comfortable with SOUER tonometer compared with NCT. Conclusion SOUER tonometer show well accuracy,convenience,and compliance for clinical use for various patients,especially for the special patients such as infants,patients without good fixation,bed-ridden patients.