CHINESE JOURNAL OF EXPERIMENTAL OPHTHALMOLOGY
2011年
5期
456-459
,共4页
Ieare回弹式眼压计%眼压%中央角膜厚度
Ieare迴彈式眼壓計%眼壓%中央角膜厚度
Ieare회탄식안압계%안압%중앙각막후도
Icare rebound tonometer%Intraocular pressure%Central corneal thickniss
背景 Icare回弹式眼压计作为一种新式眼压计,有必要对它的临床应用价值进行评估.目的 通过比较分析Icare回弹式眼压计和Goldmann压平眼压计(GAT)的眼压测量结果,探讨Icare的临床价值.方法可疑青光眼、青光眼、屈光不正及部分健康体检者78例共152眼同时接受Icare、GAT眼压测量,受检眼先行Icare测量,然后再进行GAT测量,2次测量间隔3~5 min.对比分析两种眼压计的测量结果.结果使用Ieare和GAT测得的眼压均值分别为(19.16±5.03)mmHg和(18.41±4.52)mmHg,96眼(63.2%)两者的眼压差值≤1 mmHg,二者的测量值差异虽有统计学意义,但二者的变化呈明显正相关(r=0.940,P<0.01).当Icare眼压测量值<16 mmHg时,Icare的眼压测量值低于GAT,而当Icare眼压测量值≥16 mmHg时恰好相反;CCT偏薄、正常以及偏厚的情况下,Icare的眼压测量值均高于GAT的眼压测量值.Icare、GAT的眼压测量值和CCT间呈正相关(r=0.341,P<0.01;r=0.333,P<0.01).结论 与GAT眼压计比较,Icare回弹式眼压计易操作,测量结果可靠.临床实用性更强.
揹景 Icare迴彈式眼壓計作為一種新式眼壓計,有必要對它的臨床應用價值進行評估.目的 通過比較分析Icare迴彈式眼壓計和Goldmann壓平眼壓計(GAT)的眼壓測量結果,探討Icare的臨床價值.方法可疑青光眼、青光眼、屈光不正及部分健康體檢者78例共152眼同時接受Icare、GAT眼壓測量,受檢眼先行Icare測量,然後再進行GAT測量,2次測量間隔3~5 min.對比分析兩種眼壓計的測量結果.結果使用Ieare和GAT測得的眼壓均值分彆為(19.16±5.03)mmHg和(18.41±4.52)mmHg,96眼(63.2%)兩者的眼壓差值≤1 mmHg,二者的測量值差異雖有統計學意義,但二者的變化呈明顯正相關(r=0.940,P<0.01).噹Icare眼壓測量值<16 mmHg時,Icare的眼壓測量值低于GAT,而噹Icare眼壓測量值≥16 mmHg時恰好相反;CCT偏薄、正常以及偏厚的情況下,Icare的眼壓測量值均高于GAT的眼壓測量值.Icare、GAT的眼壓測量值和CCT間呈正相關(r=0.341,P<0.01;r=0.333,P<0.01).結論 與GAT眼壓計比較,Icare迴彈式眼壓計易操作,測量結果可靠.臨床實用性更彊.
배경 Icare회탄식안압계작위일충신식안압계,유필요대타적림상응용개치진행평고.목적 통과비교분석Icare회탄식안압계화Goldmann압평안압계(GAT)적안압측량결과,탐토Icare적림상개치.방법가의청광안、청광안、굴광불정급부분건강체검자78례공152안동시접수Icare、GAT안압측량,수검안선행Icare측량,연후재진행GAT측량,2차측량간격3~5 min.대비분석량충안압계적측량결과.결과사용Ieare화GAT측득적안압균치분별위(19.16±5.03)mmHg화(18.41±4.52)mmHg,96안(63.2%)량자적안압차치≤1 mmHg,이자적측량치차이수유통계학의의,단이자적변화정명현정상관(r=0.940,P<0.01).당Icare안압측량치<16 mmHg시,Icare적안압측량치저우GAT,이당Icare안압측량치≥16 mmHg시흡호상반;CCT편박、정상이급편후적정황하,Icare적안압측량치균고우GAT적안압측량치.Icare、GAT적안압측량치화CCT간정정상관(r=0.341,P<0.01;r=0.333,P<0.01).결론 여GAT안압계비교,Icare회탄식안압계역조작,측량결과가고.림상실용성경강.
Background As a new tonometer,it is necessary to assess the clinical value of Icare rebound tonometer.Objective This study was to compare the intraocular pressure(IOP)values measured by Icare with that measured by GAT,and discuss the clinical value of leare rebound tonometer. Methods IOP measurement was performed on 152 eyes of 78 subjects with suspicious glaucoma,glaucoma,refractive error and normal examinnee by Icare and GAT respectively.The Icare IOP was measured firstly and then the GAT IOP was carried out with the 3-or 5-minute interval.The IOP values were compared between Ieare and GAT.This study was approved by Ethic Committee of Wuhan General Hospital of Chinese PLA.Written informed consent was obtained from each subject prior to this study. Results The mean IOP values of Icare and GAT were(19.16±5.03)mmHg and(18.41±4.52)mmHg respectively.The differences between Icare IOP and CAT IOP were less than or equal to 1 mmHg in 96 of 105 eyes(63.2%).The positive correlation was found between the Icare IOP and GAT IOP(r=0.940,P<0.01).The Ieare IOP was lower than that of GAT when IOPIcare<16 mmHg,however,the IOP of Icare were higher when IOPIcare≥6 mmHg;the IOP of Icare were higher than that of GAT in the total CCT range.The correlation coefficients of IOP of Icare or CAT with CCT were 0.341(P<0.01)and 0.333(P<0.01),respectively. Conclusion Compared with GAT,Icare is more feasible in clinic because it is practicable and reliable.