目的 研究后房植入Ahmed引流管对术前伴浅前房人工晶状体眼的难治性青光眼的手术效果及安全性.方法 取5例(6只眼)入工晶状体眼性青光眼患者行Ahmed青光眼瓣膜阀植入,因前房较浅,术中将引流管植入后房,观察术后眼压控制情况及术中术后并发症,随访6~14个月.结果 术后6个月时,患者眼压平均为(17.8±1.8)mmHg,其中1例患者术后3周时眼压升高,经超声生物显微镜检查发现引流管退缩至巩膜内,二次手术后眼压恢复正常.1例患者术后4个月出现玻璃体阻塞引流管.术中及术后未见其他并发症发生.结论 对于术前发现浅前房的人工晶状体眼性青光眼者,术中将引流管植入后房是一种较为安全有效地手术方式.
目的 研究後房植入Ahmed引流管對術前伴淺前房人工晶狀體眼的難治性青光眼的手術效果及安全性.方法 取5例(6隻眼)入工晶狀體眼性青光眼患者行Ahmed青光眼瓣膜閥植入,因前房較淺,術中將引流管植入後房,觀察術後眼壓控製情況及術中術後併髮癥,隨訪6~14箇月.結果 術後6箇月時,患者眼壓平均為(17.8±1.8)mmHg,其中1例患者術後3週時眼壓升高,經超聲生物顯微鏡檢查髮現引流管退縮至鞏膜內,二次手術後眼壓恢複正常.1例患者術後4箇月齣現玻璃體阻塞引流管.術中及術後未見其他併髮癥髮生.結論 對于術前髮現淺前房的人工晶狀體眼性青光眼者,術中將引流管植入後房是一種較為安全有效地手術方式.
목적 연구후방식입Ahmed인류관대술전반천전방인공정상체안적난치성청광안적수술효과급안전성.방법 취5례(6지안)입공정상체안성청광안환자행Ahmed청광안판막벌식입,인전방교천,술중장인류관식입후방,관찰술후안압공제정황급술중술후병발증,수방6~14개월.결과 술후6개월시,환자안압평균위(17.8±1.8)mmHg,기중1례환자술후3주시안압승고,경초성생물현미경검사발현인류관퇴축지공막내,이차수술후안압회복정상.1례환자술후4개월출현파리체조새인류관.술중급술후미견기타병발증발생.결론 대우술전발현천전방적인공정상체안성청광안자,술중장인류관식입후방시일충교위안전유효지수술방식.
Objective To observe the surgical effectiveness and safety of Ahmed glaucoma valve (AGV) implant insertion in the posterior chamber sulcus for pseudophakic glaucoma with shallow anterior chamber preoperative. Methods The medical records of 5 patients (6 eyes) with pseudophakia accompanied shallow anterior chamber who underwent insertion of an Ahmed valve with the silicone tube placed in the posterior chamber sulcus were retrospectively reviewed. The postoperative IOP and surgical complications were observed with a follow-up of 6-14months. Results The mean IOP was (17.8± 1.8) mmHg at 6 months postoperatively. One patient presented a high pressure at 3 weeks after the operation and the silicone tube draw back to the sclera stroma were found by ultrasound biomicroscopy (UBM) and the IOP was normal after re-operative. The silicone tube of another patient was blocked by vitreous at 4 months postoperatively. No other complications were observed during surgery or follow-up. Conclusions Placement of the silicone tube in the posterior chamber is an effective and safe alternative technique for pseudophakic glaucoma with shallow anterior chamber.