中华麻醉学杂志
中華痳醉學雜誌
중화마취학잡지
CHINESE JOURNAL OF ANESTHESIOLOGY
2008年
12期
1114-1116
,共3页
祝胜美%吴健%何慧梁%诸绍君
祝勝美%吳健%何慧樑%諸紹君
축성미%오건%하혜량%제소군
热稀释法%肺%肺动脉%心排血量%监测,生理学
熱稀釋法%肺%肺動脈%心排血量%鑑測,生理學
열희석법%폐%폐동맥%심배혈량%감측,생이학
Thermodilution%Lung%Pulmonary artery%Cardiac output%Monitoring,physiologic
目的 评价经肺热稀释法与肺动脉热稀释法测定猪心排血量(CO)的一致性.方法对13只未成年猪,麻醉诱导后气管插管,机械通气,通过设定的方法依次进行血液动力学干预,于每次干预前及干预过程中血液动力学稳定至少10 min时,同时采用经肺热稀释法和肺动脉热稀释法测定CO(COTP,COPA).采用Bland-Altman法进行一致性分析,设定两种测定方法可以互换的界限为COTP 的相对误差<20%.结果 共收集205对CO数据,COTP范围为1.4~7.06 L/min,COPA范围为1.25~6.55 L/min,COTP值比COPA值高约2%;COTP的偏离度为0.07 L/min,95%可信区间(CI)为0.03~0.11 L/min;一致性范围为-0.46~0.60 L/min,其下限的95%CI为-0.53~-0.40 L/min,其上限的95%CI为0.54~0.66 L/min;COTP的相对误差为15.3%.结论 经肺热稀释法和肺动脉热稀释法测定猪CO具有良好的一致性.
目的 評價經肺熱稀釋法與肺動脈熱稀釋法測定豬心排血量(CO)的一緻性.方法對13隻未成年豬,痳醉誘導後氣管插管,機械通氣,通過設定的方法依次進行血液動力學榦預,于每次榦預前及榦預過程中血液動力學穩定至少10 min時,同時採用經肺熱稀釋法和肺動脈熱稀釋法測定CO(COTP,COPA).採用Bland-Altman法進行一緻性分析,設定兩種測定方法可以互換的界限為COTP 的相對誤差<20%.結果 共收集205對CO數據,COTP範圍為1.4~7.06 L/min,COPA範圍為1.25~6.55 L/min,COTP值比COPA值高約2%;COTP的偏離度為0.07 L/min,95%可信區間(CI)為0.03~0.11 L/min;一緻性範圍為-0.46~0.60 L/min,其下限的95%CI為-0.53~-0.40 L/min,其上限的95%CI為0.54~0.66 L/min;COTP的相對誤差為15.3%.結論 經肺熱稀釋法和肺動脈熱稀釋法測定豬CO具有良好的一緻性.
목적 평개경폐열희석법여폐동맥열희석법측정저심배혈량(CO)적일치성.방법대13지미성년저,마취유도후기관삽관,궤계통기,통과설정적방법의차진행혈액동역학간예,우매차간예전급간예과정중혈액동역학은정지소10 min시,동시채용경폐열희석법화폐동맥열희석법측정CO(COTP,COPA).채용Bland-Altman법진행일치성분석,설정량충측정방법가이호환적계한위COTP 적상대오차<20%.결과 공수집205대CO수거,COTP범위위1.4~7.06 L/min,COPA범위위1.25~6.55 L/min,COTP치비COPA치고약2%;COTP적편리도위0.07 L/min,95%가신구간(CI)위0.03~0.11 L/min;일치성범위위-0.46~0.60 L/min,기하한적95%CI위-0.53~-0.40 L/min,기상한적95%CI위0.54~0.66 L/min;COTP적상대오차위15.3%.결론 경폐열희석법화폐동맥열희석법측정저CO구유량호적일치성.
Objective To evaluate the agreement between transpulmonary thermodilution and pulmonary artery catheter thermodilution cardiac output(CO)measurements in pigs.Methods Thirteen immature pigs weighing 22-25 kg were studied.Tracheal intubation was performed and the animals were mechanically ventilated after induction of anesthesia.In each animal,some predetermined methods were used in turn to intervene in the hemodynamics.CO measurements were performed simultaneously by transpulmonary thermodilution(COTP)and pulmonary artery catheter thermodilution(COPA)immediately before intervention and at least 10 min following stable hemodynamics during each intervention.The agreement was assessed using Bland-Ahman analysis,and the interchangeable limits of the two methods were defined as the relative error of COTP less than 20%.Results CO measurements yielded a total of 205 data pairs with a range of 1.4-7.06 L/min for COTP and 1.25-6.55 L/min for COPA and the value of COTP was about 2% higher than that of COPA.The bias of COTP was 0.07 L/min and 95% confidence intenral(CI)was 0.03-0.11 L/min;the limit of agreement was-0.46-0.60 L/min and 95% CI for the lower and upper hmits of agreement was-0.53--0.40 L/min and 0.54-0.66 L/min respectively.The relative error of COTP was 15.3%.Conclusion There is a good agreement between transpulmonary thermodilution and pulmonary artery catheter thermodilution CO measurements in pigs.