张晓峰%黄飙%虞竞峰%祝炳方%强新展%肖华龙
張曉峰%黃飆%虞競峰%祝炳方%彊新展%肖華龍
장효봉%황표%우경봉%축병방%강신전%초화룡
细胞黏附分子%荧光免疫测定%方法
細胞黏附分子%熒光免疫測定%方法
세포점부분자%형광면역측정%방법
Cell adhesion molecules%Fluoroimmunoassay%Methods
目的 评估所建立的可溶性细胞间黏附分子-1生物素-亲和素系统(BAS)时间分辨荧光免疫分析方法(sICAM-1 BSA-TRFIA)的技术指标及应用价值.方法 建立sICAM-1 BsA-TR-FIA,分析其检测曲线的性质参数,研究该方法的灵敏度、精密度和回收率、特异性,稳定性、平行性,以及与酶联免疫吸附试验分析法(ELlSA)间的相关性.结果 此法特异性较高,同批试剂连续6个月应用分析结果稳定,灵敏度为1.32 μg/L,批内和批间CV分别为4.24%和6.83%,平均回收率为99.87%,线性范围为(2.5~1 933)μg/L,与ELISA所测结果比较高度相关(r=0.939),与TNF-а、IL-8、IL-1、IL-2和IL-6等免疫分子无交叉反应.结论 sICAM一1 BSA-TRFIA是一种特异性强、稳定性好、灵敏度高、可测范围较广的检测方法,可以方便地应用于临床,为相关疾病的诊断和监测提供一项良好的实验诊断指标.
目的 評估所建立的可溶性細胞間黏附分子-1生物素-親和素繫統(BAS)時間分辨熒光免疫分析方法(sICAM-1 BSA-TRFIA)的技術指標及應用價值.方法 建立sICAM-1 BsA-TR-FIA,分析其檢測麯線的性質參數,研究該方法的靈敏度、精密度和迴收率、特異性,穩定性、平行性,以及與酶聯免疫吸附試驗分析法(ELlSA)間的相關性.結果 此法特異性較高,同批試劑連續6箇月應用分析結果穩定,靈敏度為1.32 μg/L,批內和批間CV分彆為4.24%和6.83%,平均迴收率為99.87%,線性範圍為(2.5~1 933)μg/L,與ELISA所測結果比較高度相關(r=0.939),與TNF-а、IL-8、IL-1、IL-2和IL-6等免疫分子無交扠反應.結論 sICAM一1 BSA-TRFIA是一種特異性彊、穩定性好、靈敏度高、可測範圍較廣的檢測方法,可以方便地應用于臨床,為相關疾病的診斷和鑑測提供一項良好的實驗診斷指標.
목적 평고소건립적가용성세포간점부분자-1생물소-친화소계통(BAS)시간분변형광면역분석방법(sICAM-1 BSA-TRFIA)적기술지표급응용개치.방법 건립sICAM-1 BsA-TR-FIA,분석기검측곡선적성질삼수,연구해방법적령민도、정밀도화회수솔、특이성,은정성、평행성,이급여매련면역흡부시험분석법(ELlSA)간적상관성.결과 차법특이성교고,동비시제련속6개월응용분석결과은정,령민도위1.32 μg/L,비내화비간CV분별위4.24%화6.83%,평균회수솔위99.87%,선성범위위(2.5~1 933)μg/L,여ELISA소측결과비교고도상관(r=0.939),여TNF-а、IL-8、IL-1、IL-2화IL-6등면역분자무교차반응.결론 sICAM일1 BSA-TRFIA시일충특이성강、은정성호、령민도고、가측범위교엄적검측방법,가이방편지응용우림상,위상관질병적진단화감측제공일항량호적실험진단지표.
Objective To evaluate the methodology and clinical application about the newly es-tablished time-resolved fluoroimmunoassay(TRFIA), which has been used to detect soluble intercellu-lar adhesion molecule-1 (sICAM-1) captured by the surface BAS-biotin. Methods After sICAM-1BSA-TRFIA test system was established, the parameters of the detection curve were analyzed, and further to study its sensitivity, precision, recovery rate, specificity, stability, and parallelism. Then the correlation of the method and enzyme linked immunosorbent assay(ELISA) was investigated. Re-sults This method had higher specificity and stability than ELISA. After 6 months of successive ap-plication of the same batch of agents, the analytic results kept stable. The sensitivity was 1.32 μg/L.The intra- and inter-batch CV were 4. 24% and 6.83%, respectively. The average recovery rate was 99.87%. On the test curve, the linear ranger was 2. 5-1 933μg/L. The detection results were closely correlative with those with ELISA(r=0. 939). No cross reaction exist between the new method and FIA is characteristic of strong specificity, excellent stability, high sensitivity, and wide detection range. It can be widely applied and conveniently used in the clinic as a laboratory index on clinical diag-nosis and monitoring.