谢娜%干芸根%王宏伟%曾洪武%曹卫国%孙龙伟
謝娜%榦蕓根%王宏偉%曾洪武%曹衛國%孫龍偉
사나%간예근%왕굉위%증홍무%조위국%손룡위
体层摄影术,X线计算机%辐射剂量%头%婴幼儿
體層攝影術,X線計算機%輻射劑量%頭%嬰幼兒
체층섭영술,X선계산궤%복사제량%두%영유인
Tomography,X-ray computed%Radiation dose%Head%Infant
目的 探讨使用不同管电流时间积行婴幼儿头颅CT扫描时,影响图像质量的客观因素.方法 随机将90例行头颅CT扫描的婴幼儿分成3组,测量每个婴幼儿的头颅最大前后径(MAPD),选择性研究3个参考层面(幕下小脑、幕上基底节和半卵圆中心层面),每个组只研究一个层面,每个层面用150、100及80 mAs管电流时间积进行扫描,对各层面图像进行主观质量评分,并测量其客观量化噪声值.结果 90例婴幼儿平均MAPD为(148.0±17.4)mm.在小脑层面,主观图像质量得分较其他2个层面低.管电流时间积为150、100和80 mAs时,小脑层面的量化噪声值分别较幕上层面高6.3%、9.4%和22.9%.MAPD与图像主观质量评分及客观量化噪声值之间均呈显著性相关.结论 在婴幼儿头颅CT扫描时,小脑层面固有高噪声及MAPD是影响婴幼儿头颅低剂量CT扫描图像质量的客观因素.
目的 探討使用不同管電流時間積行嬰幼兒頭顱CT掃描時,影響圖像質量的客觀因素.方法 隨機將90例行頭顱CT掃描的嬰幼兒分成3組,測量每箇嬰幼兒的頭顱最大前後徑(MAPD),選擇性研究3箇參攷層麵(幕下小腦、幕上基底節和半卵圓中心層麵),每箇組隻研究一箇層麵,每箇層麵用150、100及80 mAs管電流時間積進行掃描,對各層麵圖像進行主觀質量評分,併測量其客觀量化譟聲值.結果 90例嬰幼兒平均MAPD為(148.0±17.4)mm.在小腦層麵,主觀圖像質量得分較其他2箇層麵低.管電流時間積為150、100和80 mAs時,小腦層麵的量化譟聲值分彆較幕上層麵高6.3%、9.4%和22.9%.MAPD與圖像主觀質量評分及客觀量化譟聲值之間均呈顯著性相關.結論 在嬰幼兒頭顱CT掃描時,小腦層麵固有高譟聲及MAPD是影響嬰幼兒頭顱低劑量CT掃描圖像質量的客觀因素.
목적 탐토사용불동관전류시간적행영유인두로CT소묘시,영향도상질량적객관인소.방법 수궤장90례행두로CT소묘적영유인분성3조,측량매개영유인적두로최대전후경(MAPD),선택성연구3개삼고층면(막하소뇌、막상기저절화반란원중심층면),매개조지연구일개층면,매개층면용150、100급80 mAs관전류시간적진행소묘,대각층면도상진행주관질량평분,병측량기객관양화조성치.결과 90례영유인평균MAPD위(148.0±17.4)mm.재소뇌층면,주관도상질량득분교기타2개층면저.관전류시간적위150、100화80 mAs시,소뇌층면적양화조성치분별교막상층면고6.3%、9.4%화22.9%.MAPD여도상주관질량평분급객관양화조성치지간균정현저성상관.결론 재영유인두로CT소묘시,소뇌층면고유고조성급MAPD시영향영유인두로저제량CT소묘도상질량적객관인소.
Objective To investigate the objective factors that affect the image quality of infant cranial CT using different mAs. Materials and Methods Ninety infants were divided into three groups randomly. The maximum anteroposterior diameter (MAPD) of skull of each infant was measured. Three reference levels, cerebellar, basal ganglia and centrum semiovale levels were selected respectively. Only one level was studied in each group and scanned with 150, 100 and 80 mAs. The subjective quality grade and the objective noise of all images were recorded and analysed statistically. Result The average MAPD of ninety patients was (148.0 ± 17.4)mm. On the cerebellar level, the subjective quality grade was lower than the other two levels, which were 6.3% , 9.4% and 22.9% respectively when mAs were 150, 100 and 80 mAs. Both quality grade of image and objective noise were significantly correlated with MAPD.Conclusion The inherent high noise of cerebellar level and MAPD were the objective factors that affect the image quality of low-dose cranial CT of infant.