采用阵列式电极,通过空间滤波方法可提高表面肌电信号的MUAP分辨能力.本文实现的阵列式表面肌电信号采集仪由表面肌电电极阵列、信号调理电路和数据采集部分构成.表面肌电电极阵列实现了镀金圆盘式和弹簧探针式两种类型,信号调理电路对电极上的肌电信号进行放大和滤波,数据采集部分将调理后的信号转换成数据并进行显示和存储.电极阵列和前级信号调理电路集成在一起,有效的降低了微弱信号通过导线传输所引入的干扰.通过实验,验证了这种阵列式表面肌电采集仪在研究肌肉中动作电位传播和利用空间滤波提高MUAP分辨力的可行性,并证实了镀金圆盘式电极在降低噪声方面、弹簧探针式电极在缩短MUAP时长方面的优势.
採用陣列式電極,通過空間濾波方法可提高錶麵肌電信號的MUAP分辨能力.本文實現的陣列式錶麵肌電信號採集儀由錶麵肌電電極陣列、信號調理電路和數據採集部分構成.錶麵肌電電極陣列實現瞭鍍金圓盤式和彈簧探針式兩種類型,信號調理電路對電極上的肌電信號進行放大和濾波,數據採集部分將調理後的信號轉換成數據併進行顯示和存儲.電極陣列和前級信號調理電路集成在一起,有效的降低瞭微弱信號通過導線傳輸所引入的榦擾.通過實驗,驗證瞭這種陣列式錶麵肌電採集儀在研究肌肉中動作電位傳播和利用空間濾波提高MUAP分辨力的可行性,併證實瞭鍍金圓盤式電極在降低譟聲方麵、彈簧探針式電極在縮短MUAP時長方麵的優勢.
채용진렬식전겁,통과공간려파방법가제고표면기전신호적MUAP분변능력.본문실현적진렬식표면기전신호채집의유표면기전전겁진렬、신호조리전로화수거채집부분구성.표면기전전겁진렬실현료도금원반식화탄황탐침식량충류형,신호조리전로대전겁상적기전신호진행방대화려파,수거채집부분장조리후적신호전환성수거병진행현시화존저.전겁진렬화전급신호조리전로집성재일기,유효적강저료미약신호통과도선전수소인입적간우.통과실험,험증료저충진렬식표면기전채집의재연구기육중동작전위전파화이용공간려파제고MUAP분변력적가행성,병증실료도금원반식전겁재강저조성방면、탄황탐침식전겁재축단MUAP시장방면적우세.
The discrimination of MUAP from surface Electromyogram (sEMG )can be improved by using elec-trodes array and space filtering method.According to this principle,an array acquisition instrument for sEMG was de-veloped in this paper,containing sEMG electrodes array,signal conditioning circuits and a data acquisition part.Gold plated disk shaped electrodes and spring probe electrodes built up two different kinds of sEMG electrode array;The conditioning circuit amplifies and filters the signal form electrode array;The data acquisition part converted the condi-tioned signal to digital for display and storage.The electrode array is integrated with the first-stage conditioning circuit,in order to prevent interferences of weak signal transmission through wires.It is proved that the array acquisition in-strument can be used in the research of action potential transmission in muscles,and it can improve the discrimination of MUAP by using space filtering method.It is also proved that the gold plated disk shaped electrodes work better in noise reduction and the spring probe electrodes work better in reducing the MUAP duration.