COD测定仪%替代试剂%分光光度计%应用
COD測定儀%替代試劑%分光光度計%應用
COD측정의%체대시제%분광광도계%응용
COD detector%alternative reagent%spectrophotometer%application
对进口的COD测定试剂进行了替代品开发,研制出适用于该仪器的替代试剂。通过实验找到了替代试剂最佳配比。实验证实,自制试剂可完全替代HACH公司的消解试剂用于各种工业和生活废水的消解测定,而且该自制试剂可用于普通分光光度计测定。
對進口的COD測定試劑進行瞭替代品開髮,研製齣適用于該儀器的替代試劑。通過實驗找到瞭替代試劑最佳配比。實驗證實,自製試劑可完全替代HACH公司的消解試劑用于各種工業和生活廢水的消解測定,而且該自製試劑可用于普通分光光度計測定。
대진구적COD측정시제진행료체대품개발,연제출괄용우해의기적체대시제。통과실험조도료체대시제최가배비。실험증실,자제시제가완전체대HACH공사적소해시제용우각충공업화생활폐수적소해측정,이차해자제시제가용우보통분광광도계측정。
Through the research and development, the substitute reagent has been found for the COD detector. The best proportion of the substitute has been set by the experimental tests. It is testified that the digestive reagent of HACH Company could be replaced with the self - made reagent for digestive detection of both industrial and domestic wastewater. The alternative reagent could also be applied in the common spectrophotometer.