Iris localization%Wavefront aberration
目的 探讨虹膜定位联合波前像差引导个体化LASIK术前检查及手术过程中虹膜定位的正确方法.方法 采用博士伦公司Zywave波前像差仪,采集在暗室瞳孔自然散大状态下的虹膜数据及波前像差数据,通过网络传输,在切削角膜瓣后未掀开角膜瓣前,启动虹膜定位功能,进行虹膜定位.结果 采集虹膜图象时的要求:头位的额状面与波前像差仪的矢状面相垂直,虹膜定位后的正确图象显示为大圆圈与术眼角膜缘吻合,小圆圈与瞳孔吻合,小圆环内的6个反光白点清晰并分布均匀,瞳孔中心位的绿色“+”与坐标的黑色“+”中心位相吻,角膜瓣表面的水分要充足,虹膜纹理调至最清晰,捕捉时的瞳孔直径在3.5~5.0 mm最佳.结论 掌握好采集虹膜图象时的检查要求、作好虹膜定位的具体步骤,是保证虹膜定位联合波前像差引导的个体化LASIK效果的关键.
目的 探討虹膜定位聯閤波前像差引導箇體化LASIK術前檢查及手術過程中虹膜定位的正確方法.方法 採用博士倫公司Zywave波前像差儀,採集在暗室瞳孔自然散大狀態下的虹膜數據及波前像差數據,通過網絡傳輸,在切削角膜瓣後未掀開角膜瓣前,啟動虹膜定位功能,進行虹膜定位.結果 採集虹膜圖象時的要求:頭位的額狀麵與波前像差儀的矢狀麵相垂直,虹膜定位後的正確圖象顯示為大圓圈與術眼角膜緣吻閤,小圓圈與瞳孔吻閤,小圓環內的6箇反光白點清晰併分佈均勻,瞳孔中心位的綠色“+”與坐標的黑色“+”中心位相吻,角膜瓣錶麵的水分要充足,虹膜紋理調至最清晰,捕捉時的瞳孔直徑在3.5~5.0 mm最佳.結論 掌握好採集虹膜圖象時的檢查要求、作好虹膜定位的具體步驟,是保證虹膜定位聯閤波前像差引導的箇體化LASIK效果的關鍵.
목적 탐토홍막정위연합파전상차인도개체화LASIK술전검사급수술과정중홍막정위적정학방법.방법 채용박사륜공사Zywave파전상차의,채집재암실동공자연산대상태하적홍막수거급파전상차수거,통과망락전수,재절삭각막판후미흔개각막판전,계동홍막정위공능,진행홍막정위.결과 채집홍막도상시적요구:두위적액상면여파전상차의적시상면상수직,홍막정위후적정학도상현시위대원권여술안각막연문합,소원권여동공문합,소원배내적6개반광백점청석병분포균균,동공중심위적록색“+”여좌표적흑색“+”중심위상문,각막판표면적수분요충족,홍막문리조지최청석,포착시적동공직경재3.5~5.0 mm최가.결론 장악호채집홍막도상시적검사요구、작호홍막정위적구체보취,시보증홍막정위연합파전상차인도적개체화LASIK효과적관건.
Objective To study right methods to localize iris before and during iris localization combined with wavefront aberration guided LASIK.Methods The data of iris and wavefront aberration were collected by the Zywave wavefront analyzer ( USA Bausch & Lomb Company) under the dark room with naturally dilated pupil and were transmitted through the network.Iris localization started after cutting corneal flap and before openingit.Results Requirements of capturing iris images were stated as follow.Frontal plane of the head position and sagittal plane of the Zywave wavefront analyzer were perpendicular.The right images after iris localization showed large circle coincided with the edge of the cornea,small circle coincided with the pupil,six reflective white spots in small circle were clear and evenly distributed.The green “ + ” in the center position of the pupil coincided with the center of black “ +” in axis.The water on corneal flap surface should be sufficient,iris textures were the clearest,and pupil diameter from 3.5 to 5.0 mm during capturing images were optimal.Conclusions The keys to ensure the effect in iris localization combined with wavefront aberration guided LASIK were the inspection of requirements during capturing images,and the specific steps during iris localization were well made.