Intrauterine devices,medicated%Intrauterines device,copper%Contrast clinical trial
目的 观察释放吲哚美辛的V形含铜含药宫内节育器(MYCu IUD)的临床效果及安全性.方法 采用临床多中心,按统一标准选择2000例要求使用IUD避孕的妇女,随机放置MYCu IUD与T形含铜(TCu380A)IUD各1000例,放置后1、3、6、12个月时定期随访,监测效果及安全性.结果 放置12个月时,MYCu IUD组与TCu380A IUD组分别使用11 985.9、11 753.6个妇女月,累积继续使用率分别为97.80%、94.70%;累积带器妊娠率分别为0.10%、0.31%(P>0.05);累积脱落(部分脱落)率MYCu IUD组为0.10%,低于TCu380A IUD组的1.73%(P<0.01);累积因医疗原因[出血和(或)疼痛]取出率分别为1.51%、2.94%(P<0.05);与IUD使用有关的终止、总终止率两组差异有统计学意义(P<0.01);置器后1、3、6、12个月时不规则出血、经期延长、经量增多及疼痛的发生率含吲哚美辛的MYCu IUD组明显少于不含吲哚美辛的TCu380A IUD组(P<0.01).结论 MYCu IUD脱落率极低,抗生育效果好,置取方便、安全,置器后不良反应小,是比较理想的IUD,应推广使用.
目的 觀察釋放吲哚美辛的V形含銅含藥宮內節育器(MYCu IUD)的臨床效果及安全性.方法 採用臨床多中心,按統一標準選擇2000例要求使用IUD避孕的婦女,隨機放置MYCu IUD與T形含銅(TCu380A)IUD各1000例,放置後1、3、6、12箇月時定期隨訪,鑑測效果及安全性.結果 放置12箇月時,MYCu IUD組與TCu380A IUD組分彆使用11 985.9、11 753.6箇婦女月,纍積繼續使用率分彆為97.80%、94.70%;纍積帶器妊娠率分彆為0.10%、0.31%(P>0.05);纍積脫落(部分脫落)率MYCu IUD組為0.10%,低于TCu380A IUD組的1.73%(P<0.01);纍積因醫療原因[齣血和(或)疼痛]取齣率分彆為1.51%、2.94%(P<0.05);與IUD使用有關的終止、總終止率兩組差異有統計學意義(P<0.01);置器後1、3、6、12箇月時不規則齣血、經期延長、經量增多及疼痛的髮生率含吲哚美辛的MYCu IUD組明顯少于不含吲哚美辛的TCu380A IUD組(P<0.01).結論 MYCu IUD脫落率極低,抗生育效果好,置取方便、安全,置器後不良反應小,是比較理想的IUD,應推廣使用.
목적 관찰석방신타미신적V형함동함약궁내절육기(MYCu IUD)적림상효과급안전성.방법 채용림상다중심,안통일표준선택2000례요구사용IUD피잉적부녀,수궤방치MYCu IUD여T형함동(TCu380A)IUD각1000례,방치후1、3、6、12개월시정기수방,감측효과급안전성.결과 방치12개월시,MYCu IUD조여TCu380A IUD조분별사용11 985.9、11 753.6개부녀월,루적계속사용솔분별위97.80%、94.70%;루적대기임신솔분별위0.10%、0.31%(P>0.05);루적탈락(부분탈락)솔MYCu IUD조위0.10%,저우TCu380A IUD조적1.73%(P<0.01);루적인의료원인[출혈화(혹)동통]취출솔분별위1.51%、2.94%(P<0.05);여IUD사용유관적종지、총종지솔량조차이유통계학의의(P<0.01);치기후1、3、6、12개월시불규칙출혈、경기연장、경량증다급동통적발생솔함신타미신적MYCu IUD조명현소우불함신타미신적TCu380A IUD조(P<0.01).결론 MYCu IUD탈락솔겁저,항생육효과호,치취방편、안전,치기후불량반응소,시비교이상적IUD,응추엄사용.
Objective To observe safety and efficacy of intrauterine device MYCu IUD that releases indomethacin. Methods In total, 2000 women requiring IUD for contraception were chosen from domestic multiple clinical centers according to an unified standard. MYCu IUD and TCu380A IUD were randomly inserted for 1000 women each, respectively. All women were regularly followed-up one, three, six and 12 months after insertion. Results All the women were followed-up for 11 985.9 and 11 753.6 person-months in MYCu IUD and TCu380A IUD groups, respectively, with cumulative IUD application rates of 97. 80% and 94. 70%, and cumulative pregnancy rates with IUD of 0. 10% and 0. 31%, respectively. Cumulative expulsion (or partial expulsion) rate in MYCu IUD group was 0. 10%, significantly lower than that in TCu380A IUD one (1.73%), P <0. 01, and cumulative removal rate due to medical reasons (bleeding or pain) was 1.51 % and 2. 94%, respectively, P < 0. 05. Rate of IUD-associated termination and overall rate of termination differed significantly between the two groups, P <0. 01. Adverse reactions one, three, six and twelve months after IUD insertion included irregular bleeding, prolonged menstruation, increased menstrual blood volume and pain, less in indomethacin-containing MYCu IUD group than those in TCu380A IUD without containing it, P < 0. 01. Conclusions MYCu IUD is a comparatively ideal IUD that should be promoted for extensive use, with extremely low expulsion rate, good contraceptive effect, convenience and safety in insertion and removal, and less adverse effects.