2012年
3期
285-288
,共4页
韩军%李勤%梁志文%陈秘
韓軍%李勤%樑誌文%陳祕
한군%리근%량지문%진비
剂量学,胶片%吸光度%冲洗与扫描条件
劑量學,膠片%吸光度%遲洗與掃描條件
제량학,효편%흡광도%충세여소묘조건
Dosimetry,film%Optical density%Flushing and scanning condition
目的 分析冲洗和扫描条件对胶片剂量与吸光度(A)值的影响,建立基于卤化银胶片剂量仪的临床质量保证程序.方法 选择不同洗片温度、不同批次胶片、不同批次药水及同批次药水和胶片分析剂量与A值的响应.测试扫描仪不均匀性、稳定工作时间和噪声对胶片剂量响应的影响.结果 同一剂量水平A值随温度升高而增大;不同批次胶片剂量与A值响应差异较大;不同批次药水、胶片剂量与A值响应差异明显;同一批次胶片剂量与A值响应与药水时间呈现一定规律.扫描仪扫描的不均匀性最大达0.03,相同剂量A值在预热10 min后稳定,噪声影响随剂量和分辨率增加而加大.结论 最佳洗片温度为29~31℃;不同批次胶片不能混用,更换胶片、药水时需重新获取刻度片.扫描仪需进行不均匀性校准,最佳扫描设置为72 dpi、16 bit,且要预热5~10 min.
目的 分析遲洗和掃描條件對膠片劑量與吸光度(A)值的影響,建立基于滷化銀膠片劑量儀的臨床質量保證程序.方法 選擇不同洗片溫度、不同批次膠片、不同批次藥水及同批次藥水和膠片分析劑量與A值的響應.測試掃描儀不均勻性、穩定工作時間和譟聲對膠片劑量響應的影響.結果 同一劑量水平A值隨溫度升高而增大;不同批次膠片劑量與A值響應差異較大;不同批次藥水、膠片劑量與A值響應差異明顯;同一批次膠片劑量與A值響應與藥水時間呈現一定規律.掃描儀掃描的不均勻性最大達0.03,相同劑量A值在預熱10 min後穩定,譟聲影響隨劑量和分辨率增加而加大.結論 最佳洗片溫度為29~31℃;不同批次膠片不能混用,更換膠片、藥水時需重新穫取刻度片.掃描儀需進行不均勻性校準,最佳掃描設置為72 dpi、16 bit,且要預熱5~10 min.
목적 분석충세화소묘조건대효편제량여흡광도(A)치적영향,건립기우서화은효편제량의적림상질량보증정서.방법 선택불동세편온도、불동비차효편、불동비차약수급동비차약수화효편분석제량여A치적향응.측시소묘의불균균성、은정공작시간화조성대효편제량향응적영향.결과 동일제량수평A치수온도승고이증대;불동비차효편제량여A치향응차이교대;불동비차약수、효편제량여A치향응차이명현;동일비차효편제량여A치향응여약수시간정현일정규률.소묘의소묘적불균균성최대체0.03,상동제량A치재예열10 min후은정,조성영향수제량화분변솔증가이가대.결론 최가세편온도위29~31℃;불동비차효편불능혼용,경환효편、약수시수중신획취각도편.소묘의수진행불균균성교준,최가소묘설치위72 dpi、16 bit,차요예열5~10 min.
Objective To explore the effects of different flushing and scanning processing on optical density (A) responding and set up the clinic quality assurance protocol based on silver halide emulsions radiographic films.Methods Setting different flushing temperature and choosing different batch's film and developer and fixer or fixer in the same batch with different analyzing dose were performed to analyze the effect on A value; The effect of light uniform,the stabilize work time and the noise of different scanning resolution were discussed.Results The A value at the same dose level would enhance as the temperature increased;the responding curve of dose and A value with different batches of films varied a lot;the responding curve of dose and A value with different batches of films and developer and fixer solution had marked variations;the responding of dose and A with the same batches would show some low.The heterogeneity of the scanner would achieve 0.03 ; the A value of the same dose would gradually steady while the time ofscanning was more than 10 minutes:the affect of noise would increase as the dose and resolution ratio increased.Conclusions The best processing temperature is 29-31 ℃.Different batches of film couldn't be confounded.A new calibration must be obtained when the film dosimetry to evaluate dose distribution is used.A 5 - 10 min warm up for stabilize work and the best setting resolution/depth lever are 72 dpi/16 bit for scanning films are determined.