目的 评估VERIS mfERG (EDI Inc)和Espion multifocal imager (Diagnosys)两种不同系统测量正常眼多焦视网膜电图(mfERG)的电位波反应的差异及其相关性,并用两个不同系统分别测量黄斑前膜患眼的mfERG并进行分析.方法 这是一项横断面研究,在VERIS系统使用Burian-Allen (BA)电极,Espion系统使用BA和DTL电极分别测量25只正常眼的mfERG,将3次测得的一阶反应的相应数据进行配对t检验及Pearson相关分析.两个系统用BA电极分别测量19只黄斑前膜患眼的mfERG,对一阶反应的相应数据进行Fisher's exact test分析.结果 VERIS BA测得正常眼N1波、P1波振幅密度值最大,其次为Espion BA,Espion DTL最小;VERIS BA测得N1波、P1波潜伏期最长,Espion BA与Espion DTL差异无统计学意义.Espion BA与VERIS BA测得的P1波振幅密度、潜伏期皆有明显的相关性.黄斑前膜患眼中,VERIS BA比Espion BA能检测到更多P1波振幅密度和P1波潜伏期的异常.结论 不论VERIS或Espion系统都能成功记录多焦视网膜电图,当两种系统都使用Burian-Allen电极时,两者mfERG一阶反应P1波振幅密度及潜伏期有着良好的相关性.
目的 評估VERIS mfERG (EDI Inc)和Espion multifocal imager (Diagnosys)兩種不同繫統測量正常眼多焦視網膜電圖(mfERG)的電位波反應的差異及其相關性,併用兩箇不同繫統分彆測量黃斑前膜患眼的mfERG併進行分析.方法 這是一項橫斷麵研究,在VERIS繫統使用Burian-Allen (BA)電極,Espion繫統使用BA和DTL電極分彆測量25隻正常眼的mfERG,將3次測得的一階反應的相應數據進行配對t檢驗及Pearson相關分析.兩箇繫統用BA電極分彆測量19隻黃斑前膜患眼的mfERG,對一階反應的相應數據進行Fisher's exact test分析.結果 VERIS BA測得正常眼N1波、P1波振幅密度值最大,其次為Espion BA,Espion DTL最小;VERIS BA測得N1波、P1波潛伏期最長,Espion BA與Espion DTL差異無統計學意義.Espion BA與VERIS BA測得的P1波振幅密度、潛伏期皆有明顯的相關性.黃斑前膜患眼中,VERIS BA比Espion BA能檢測到更多P1波振幅密度和P1波潛伏期的異常.結論 不論VERIS或Espion繫統都能成功記錄多焦視網膜電圖,噹兩種繫統都使用Burian-Allen電極時,兩者mfERG一階反應P1波振幅密度及潛伏期有著良好的相關性.
목적 평고VERIS mfERG (EDI Inc)화Espion multifocal imager (Diagnosys)량충불동계통측량정상안다초시망막전도(mfERG)적전위파반응적차이급기상관성,병용량개불동계통분별측량황반전막환안적mfERG병진행분석.방법 저시일항횡단면연구,재VERIS계통사용Burian-Allen (BA)전겁,Espion계통사용BA화DTL전겁분별측량25지정상안적mfERG,장3차측득적일계반응적상응수거진행배대t검험급Pearson상관분석.량개계통용BA전겁분별측량19지황반전막환안적mfERG,대일계반응적상응수거진행Fisher's exact test분석.결과 VERIS BA측득정상안N1파、P1파진폭밀도치최대,기차위Espion BA,Espion DTL최소;VERIS BA측득N1파、P1파잠복기최장,Espion BA여Espion DTL차이무통계학의의.Espion BA여VERIS BA측득적P1파진폭밀도、잠복기개유명현적상관성.황반전막환안중,VERIS BA비Espion BA능검측도경다P1파진폭밀도화P1파잠복기적이상.결론 불론VERIS혹Espion계통도능성공기록다초시망막전도,당량충계통도사용Burian-Allen전겁시,량자mfERG일계반응P1파진폭밀도급잠복기유착량호적상관성.
Objective To investigate the differences and correlation of the first order kernels (FOK) of mfERG from normal eyes recorded by the VERIS mfERG (EDI Inc) and Espion multifocal imager (Diagnosys) systems and to analyze the mfERG parameters in eyes based on recordings of the epiretinal membrane (ERM) recorded by the two systems.Methods This was a cross-sectional study.The mfERG of 25 normal eyes were recorded separately with the VERIS system using a Burian-Allen (BA) contact lens electrode,and with the Espion system using BA and DTL electrodes.The FOK were analyzed with paired t test and Pearson correlation analysis. The FOK mfERG of 19eyes with epiretinal membrane were also recorded by the VERIS and Espion systems using a BA electrode and the results were compared with Fisher's exact test.Results In normal eyes,the response densities of the N1 and P1 waveforms recorded by VERIS BA were the highest,followed by Espion BA and finally Espion DTL.The latencies of N1 and P1 waveforms recorded by VERIS BA were longer than latencies recorded by both Espion BA and Espion DTL,while there were no significant differences between Espion BA and Espion DTL. There were significant correlations in response densities and latencies of P1 waveforms between VERIS BA and Espion BA.In ERM eyes, there were more abnormalities in amplitude densities and latencies of the FOK P1 waveforms detected by VERIS BA than by Espion BA.Conclusion Both the VERIS and Espion systems can record mfERG well.There are significant correlations in response amplitudes and latencies of FOK P1 waveforms between VERIS BA and Espion BA.