CHINESE JOURNAL OF ANALYTICAL CHEMISTRY
2010年
3期
389-392
,共4页
甄生航%郑军%邹超世%王艳%朱洋%邓世雄%谢国明%王箭
甄生航%鄭軍%鄒超世%王豔%硃洋%鄧世雄%謝國明%王箭
견생항%정군%추초세%왕염%주양%산세웅%사국명%왕전
血乙醇%丝网印刷技术%多壁碳纳米管%生物传感器
血乙醇%絲網印刷技術%多壁碳納米管%生物傳感器
혈을순%사망인쇄기술%다벽탄납미관%생물전감기
Blood alcohol%Screen printing technique%Multi-wall carbon nanotube%Biosensor
采用丝网印刷技术在PVC基板上印制三电极,将多壁碳纳米管、麦尔多拉蓝、乙醇脱氢酶(ADH)以及氧化型辅酶Ⅰ(NAD~+)依次修饰在工作电极表面,然后在该三电极表面贴一层亲水膜,形成一个5 μL反应池,制成血乙醇检测新型生物传感器检测条.结果表明,此生物传感器具有良好的准确性和稳定性;检测线性范围为0.5~20 mmol/L,r=0.99493;检出限为0.22 mmol/L;电流达到95%稳态时间小于15 s.考察了pH值、温度及干扰物对生物传感器的影响.用此生物传感器和顶空气相色谱法对10份全血标本乙醇浓度平行测试,两者相关性良好r=0.97583.利用虹吸现象吸取微量全血直接定量测定乙醇浓度是此生物传感器的特点.
採用絲網印刷技術在PVC基闆上印製三電極,將多壁碳納米管、麥爾多拉藍、乙醇脫氫酶(ADH)以及氧化型輔酶Ⅰ(NAD~+)依次脩飾在工作電極錶麵,然後在該三電極錶麵貼一層親水膜,形成一箇5 μL反應池,製成血乙醇檢測新型生物傳感器檢測條.結果錶明,此生物傳感器具有良好的準確性和穩定性;檢測線性範圍為0.5~20 mmol/L,r=0.99493;檢齣限為0.22 mmol/L;電流達到95%穩態時間小于15 s.攷察瞭pH值、溫度及榦擾物對生物傳感器的影響.用此生物傳感器和頂空氣相色譜法對10份全血標本乙醇濃度平行測試,兩者相關性良好r=0.97583.利用虹吸現象吸取微量全血直接定量測定乙醇濃度是此生物傳感器的特點.
채용사망인쇄기술재PVC기판상인제삼전겁,장다벽탄납미관、맥이다랍람、을순탈경매(ADH)이급양화형보매Ⅰ(NAD~+)의차수식재공작전겁표면,연후재해삼전겁표면첩일층친수막,형성일개5 μL반응지,제성혈을순검측신형생물전감기검측조.결과표명,차생물전감기구유량호적준학성화은정성;검측선성범위위0.5~20 mmol/L,r=0.99493;검출한위0.22 mmol/L;전류체도95%은태시간소우15 s.고찰료pH치、온도급간우물대생물전감기적영향.용차생물전감기화정공기상색보법대10빈전혈표본을순농도평행측시,량자상관성량호r=0.97583.이용홍흡현상흡취미량전혈직접정량측정을순농도시차생물전감기적특점.
A novel disposable three electrodes blood alcohol biosensor strip was fabricated by a screen printing technique. Multi-wall carbon nanotube(MWCNT), Meldola′s(MB), alcohol dehydrogenase(ADH)and nicotinamide adenine dinucleotide cofactor (NAD+) were modified on the surface of the carbon working electrode. Then hydrophilic membrane was stuck in the outermost of the three electrodes to make a reaction camera of 5 μL. Experimental results indicated that the biosensor possessed good accuracy and stability, the linear response range was 0.5-20 mmol/L with correlation coefficient of 0.9949, detection limit was 0.22 mmol/L, and the response time was less than 15 s. Some influencing factors to the biosensor were investigated, such as the pH, temperature and interferences. Correlation analysis showed that there was a significant correlation between the methods of biosensor and the headspace vapor phase chromatography in 10 whole blood samples(r=0.97583). Small volume whole blood sucked using siphonage to detect blood alcohol directly and quantitatively was the obvious character of the biosensor.