中国实用护理杂志
中國實用護理雜誌
중국실용호리잡지
CHINESE JOURNAL OF PRACTICAL NURSING
2009年
11期
1-4
,共4页
游励红%熊子波%聂琼%张颖%李霞%隋红霞%吕超群
遊勵紅%熊子波%聶瓊%張穎%李霞%隋紅霞%呂超群
유려홍%웅자파%섭경%장영%리하%수홍하%려초군
血液透析%全自动%半自动%透析器复用
血液透析%全自動%半自動%透析器複用
혈액투석%전자동%반자동%투석기복용
Hemodialysis%Automatic%SemiI automatic%HemodiaIyzer reuse
目的 探讨半自动和全自动透析器复用方式对透析器复用效果的影响.方法 将1728支可重复使用的透析器随机分为半自动复用组和全自动复用组各864支,观察2组透析器冲洗、检测时间、使用灭菌剂成本及热原反应、透析器破膜发生率、透析器重复检测数.结果 半自动复用组透析器冲洗、血室容积检测和压力测试时间分别为(26.443±3.237),(2.172±0.128),(2.157±0.090)min,灌注3.5%伦拿连灭菌剂成本为(1.698±0.107)元,支,共有499支和243支透析器需重复进行血室容积检测及压力检测,重复检测率达57.755%和28.125%;全自动复用组透析器冲洗、血室容积检测和压力测试时间为(5.793±0.193),(1.257±0.118),(1.110±0.076)min,透析器检测均能1次完成,灭菌剂成本(9.330±0.138)元/支,2组透析器用于血液透析均未见热原反应及透析器破膜发生.结论 半自动复用组在灌注过程中能有效节省灭菌剂成本.但透析器冲洗时间长、缺乏配套检测装置,检测费时、数据不易读取,透析器重复检测率高;而全自动复用组透析器冲洗、检测方便,但消耗灭菌剂成本高,设备投入昂贵,透析中心需配置合理数量才能满足临床需求.
目的 探討半自動和全自動透析器複用方式對透析器複用效果的影響.方法 將1728支可重複使用的透析器隨機分為半自動複用組和全自動複用組各864支,觀察2組透析器遲洗、檢測時間、使用滅菌劑成本及熱原反應、透析器破膜髮生率、透析器重複檢測數.結果 半自動複用組透析器遲洗、血室容積檢測和壓力測試時間分彆為(26.443±3.237),(2.172±0.128),(2.157±0.090)min,灌註3.5%倫拿連滅菌劑成本為(1.698±0.107)元,支,共有499支和243支透析器需重複進行血室容積檢測及壓力檢測,重複檢測率達57.755%和28.125%;全自動複用組透析器遲洗、血室容積檢測和壓力測試時間為(5.793±0.193),(1.257±0.118),(1.110±0.076)min,透析器檢測均能1次完成,滅菌劑成本(9.330±0.138)元/支,2組透析器用于血液透析均未見熱原反應及透析器破膜髮生.結論 半自動複用組在灌註過程中能有效節省滅菌劑成本.但透析器遲洗時間長、缺乏配套檢測裝置,檢測費時、數據不易讀取,透析器重複檢測率高;而全自動複用組透析器遲洗、檢測方便,但消耗滅菌劑成本高,設備投入昂貴,透析中心需配置閤理數量纔能滿足臨床需求.
목적 탐토반자동화전자동투석기복용방식대투석기복용효과적영향.방법 장1728지가중복사용적투석기수궤분위반자동복용조화전자동복용조각864지,관찰2조투석기충세、검측시간、사용멸균제성본급열원반응、투석기파막발생솔、투석기중복검측수.결과 반자동복용조투석기충세、혈실용적검측화압력측시시간분별위(26.443±3.237),(2.172±0.128),(2.157±0.090)min,관주3.5%륜나련멸균제성본위(1.698±0.107)원,지,공유499지화243지투석기수중복진행혈실용적검측급압력검측,중복검측솔체57.755%화28.125%;전자동복용조투석기충세、혈실용적검측화압력측시시간위(5.793±0.193),(1.257±0.118),(1.110±0.076)min,투석기검측균능1차완성,멸균제성본(9.330±0.138)원/지,2조투석기용우혈액투석균미견열원반응급투석기파막발생.결론 반자동복용조재관주과정중능유효절성멸균제성본.단투석기충세시간장、결핍배투검측장치,검측비시、수거불역독취,투석기중복검측솔고;이전자동복용조투석기충세、검측방편,단소모멸균제성본고,설비투입앙귀,투석중심수배치합리수량재능만족림상수구.
Objective To explore the influence of automatic and semi-automatic hemodialyzer reuse method on hemodialyzer reuse effect. Methods 1728 dialyzers were randomly divided into automatichemodialyzer reuse group and semi- automatic hemodialyzer reuse group with 864 dialyzers in each group. Thetime of douching and testing, the cost of sterilization,the frequency of the pyrphgen reaction,the broken dialyzer membrane and re-examined dialyzer between the two groups were measured. Results The time of douching dialyzer, testing of total cell volume and pressure in the semi- automatic hemodialyzer reuse group was (26.443±3.237), (2.172±0.128) and (2.157±0.090) minutes respectively,while the automatic hemodialyzer reuse group was (5.793±0.193), (1.257±0.118) and (1.110±0.076) minutes respectively. The frequency of re-examined dialyzer in testing total cell volume and pressure was 499(57.755%), 243(28.125%) respectively. At the same time, all dialyzer in semi-automatic hemodialyzer reuse group could be examined successfully at a time. The cost of sterilization in automatic henmdialyzer reuse group was (9.330±0.138)yuan. No pyrogen reaction and broken dialyzer membrane happened. Conclusions The semi-automatic bemodialyzer reuse group can retrench cost during perfusion,but consumes long douching time, lacks matching detection equipment, difficult to detect, and is not easy to read data and has high re- examination rate. while in the automatic hemodialyzer group, it is convenient of douching and detection, but the cost of sterilization and equipment is high, and clinical demand can be fulfilled only when the dialysis center can allocate reasonable number of the machines.