中华超声影像学杂志
中華超聲影像學雜誌
중화초성영상학잡지
CHINESE JOURNAL OF ULTRASONOGRAPHY
2012年
1期
75-78
,共4页
区文财%吴盛正%刘政%柳建华%何景光%萧淑宜%杜洪%邱春花%马穗红
區文財%吳盛正%劉政%柳建華%何景光%蕭淑宜%杜洪%邱春花%馬穗紅
구문재%오성정%류정%류건화%하경광%소숙의%두홍%구춘화%마수홍
超声疗法%微气泡%脾破裂%止血
超聲療法%微氣泡%脾破裂%止血
초성요법%미기포%비파렬%지혈
Ultrasonic therapy%Microbubbles%Splenic rupture%Hemostasis
目的 探讨脉冲式治疗超声联合微泡对脾创伤出血的止血作用.方法 14只健康家犬随机分为3组,超声微泡组(MEUS组)6只,单纯超声组(TUS组)4只,单纯微泡组(MB组)4只.开腹切割脾建立脾破裂出血模型,测量初始出血速度.静脉缓慢推注凝血酶原复合物后,MEUS组用脉冲式超声治疗仪辐照伤口,同时静脉匀速推注微泡;TUS组超声治疗时静脉推注生理盐水;MB组超声治疗仪假照的同时静脉推注微泡.治疗完毕测量出血速度,观察止血效果,并进行超声造影评价.最后进行病理检查.结果 治疗前各组出血速度无显著差异.治疗后MEUS组脾脏伤口仅缓慢渗血,出血速度显著低于TUS组和MB(P<0.05).MEUS组造影示治疗区造影增强缺损或者低灌注,病理见脾窦扩张充血、血小板聚集,出血和血栓形成.结论 脉冲式超声联合微泡可损伤脾脏微小血管,阻断局部血液循环,可应用于脾脏创伤的止血治疗.
目的 探討脈遲式治療超聲聯閤微泡對脾創傷齣血的止血作用.方法 14隻健康傢犬隨機分為3組,超聲微泡組(MEUS組)6隻,單純超聲組(TUS組)4隻,單純微泡組(MB組)4隻.開腹切割脾建立脾破裂齣血模型,測量初始齣血速度.靜脈緩慢推註凝血酶原複閤物後,MEUS組用脈遲式超聲治療儀輻照傷口,同時靜脈勻速推註微泡;TUS組超聲治療時靜脈推註生理鹽水;MB組超聲治療儀假照的同時靜脈推註微泡.治療完畢測量齣血速度,觀察止血效果,併進行超聲造影評價.最後進行病理檢查.結果 治療前各組齣血速度無顯著差異.治療後MEUS組脾髒傷口僅緩慢滲血,齣血速度顯著低于TUS組和MB(P<0.05).MEUS組造影示治療區造影增彊缺損或者低灌註,病理見脾竇擴張充血、血小闆聚集,齣血和血栓形成.結論 脈遲式超聲聯閤微泡可損傷脾髒微小血管,阻斷跼部血液循環,可應用于脾髒創傷的止血治療.
목적 탐토맥충식치료초성연합미포대비창상출혈적지혈작용.방법 14지건강가견수궤분위3조,초성미포조(MEUS조)6지,단순초성조(TUS조)4지,단순미포조(MB조)4지.개복절할비건립비파렬출혈모형,측량초시출혈속도.정맥완만추주응혈매원복합물후,MEUS조용맥충식초성치료의복조상구,동시정맥균속추주미포;TUS조초성치료시정맥추주생리염수;MB조초성치료의가조적동시정맥추주미포.치료완필측량출혈속도,관찰지혈효과,병진행초성조영평개.최후진행병리검사.결과 치료전각조출혈속도무현저차이.치료후MEUS조비장상구부완만삼혈,출혈속도현저저우TUS조화MB(P<0.05).MEUS조조영시치료구조영증강결손혹자저관주,병리견비두확장충혈、혈소판취집,출혈화혈전형성.결론 맥충식초성연합미포가손상비장미소혈관,조단국부혈액순배,가응용우비장창상적지혈치료.
Objective To investigate the haemostatic effect of combining pulsed therapeutic ultrasound (TUS) and microbubble (MB) on splenic trauma.Methods Forteen healthy dogs were divided into 3 groups.Six animals were treated by microbubble-enhanced therapeutic ultrasound (MEUS),the other eight animals were treated with TUS only ( n =4) and the MB only ( n =4) served aS the controls.The spleens of all animals were surgically exposed and a 20 mm long,5 mm deep incision was created on the spleens using scalpel.Human prothrombin complex (HPC) was slowly injected prior to the treatment.During the procedure,the initial bleeding rate and the bleeding rate after the treatment were calculated using gauze weighing.Contrast enhanced ultrasound (CEUS) was performed to assess the blocking effects of splenic circulation.The targeted spleens were harvested for pathological examination.Results There were no differences in initial bleeding rate before treatment.After treatment,in the MEUS group oozed blood was found and the bleeding rate were lower than that of the TUS and the MB groups ( P <0.05).A nonenhanced or perfusion defect region was formed within the treated area.The histological results showed splenic sinus hyperemia,platelet aggregation and intravascular thrombosis.Conclusions Microbubbles enhanced pulsed therapeutic ultrasound can arrest the regional blood circulation of spleen by disrupting the microvasculature and this can be used in splenic heamostasis for treating spleen trauma.