目的 研究BJ-6B加速器维修、保养方法.方法 回顾分析2002-2009年BJ-6B加速器故障现象、原因、处理方法记录资料.结果 记录的231次故障中运动控制故障64次(其中手控盒部分故障20次),高压脉冲调制器报故障41次,36次报反峰过荷,5次报充电过荷,均整器故障7次,机械故障21次,数显故障20次.其余故障出现在磁控管灯丝电源、钛泵电源、光学指示系统、剂量监测系统等部位.运动控制故障率最高,其次是高压调制器部分、机械部分.结论 BJ-6B加速器技术比较成熟,故障率较低.为了提高正常开机率,医院应重视物理工程人员培训,提高其业务水平;工程师应熟读图纸,按时检查线路以及机器元部件,根据维修经验提前购置可能要更换的元件.机房湿度、温度必须控制好.随时观察出束时机器相关参数,避免打火对磁控管、加速管等珍贵元件的损害.
目的 研究BJ-6B加速器維脩、保養方法.方法 迴顧分析2002-2009年BJ-6B加速器故障現象、原因、處理方法記錄資料.結果 記錄的231次故障中運動控製故障64次(其中手控盒部分故障20次),高壓脈遲調製器報故障41次,36次報反峰過荷,5次報充電過荷,均整器故障7次,機械故障21次,數顯故障20次.其餘故障齣現在磁控管燈絲電源、鈦泵電源、光學指示繫統、劑量鑑測繫統等部位.運動控製故障率最高,其次是高壓調製器部分、機械部分.結論 BJ-6B加速器技術比較成熟,故障率較低.為瞭提高正常開機率,醫院應重視物理工程人員培訓,提高其業務水平;工程師應熟讀圖紙,按時檢查線路以及機器元部件,根據維脩經驗提前購置可能要更換的元件.機房濕度、溫度必鬚控製好.隨時觀察齣束時機器相關參數,避免打火對磁控管、加速管等珍貴元件的損害.
목적 연구BJ-6B가속기유수、보양방법.방법 회고분석2002-2009년BJ-6B가속기고장현상、원인、처리방법기록자료.결과 기록적231차고장중운동공제고장64차(기중수공합부분고장20차),고압맥충조제기보고장41차,36차보반봉과하,5차보충전과하,균정기고장7차,궤계고장21차,수현고장20차.기여고장출현재자공관등사전원、태빙전원、광학지시계통、제량감측계통등부위.운동공제고장솔최고,기차시고압조제기부분、궤계부분.결론 BJ-6B가속기기술비교성숙,고장솔교저.위료제고정상개궤솔,의원응중시물리공정인원배훈,제고기업무수평;공정사응숙독도지,안시검사선로이급궤기원부건,근거유수경험제전구치가능요경환적원건.궤방습도、온도필수공제호.수시관찰출속시궤기상관삼수,피면타화대자공관、가속관등진귀원건적손해.
Objective To study the maintenance of BJ-6B accelerator. Methods Analyzed retrospectively the maintenance record of BJ-6B accelerator, including phenomena, causes and handle from 2002 to 2009. Results In 231 records, there were motion-controlling 64, hand-controlling-pendant 20.modulator 41, anti-peak overload 36, charging overload 5. Flatness 7, mechanical 21, digital-display 20,others lie in magnetron power, water-cooling system, light-indicator system, dose-monitor system and wedge system. Motion-controlling system is the highest among those, followed by high-voltage modulator and mechanical system. Conclusions The down time for BJ-6B accelerator is low because of its perfect technology. To keep its stability in clinic, the hospital authorities should emphasize training of engineer for improving their maintenance ability. The engineer must be familiar with the circuit diagram, check the electric wire and machine unit on time and prepare unit for maintenance. The temperature and humidity in machine house must be controlled on demand. The engineer must pay attention to machine parameters when beam is on for avoiding the spark's damage to magnetron and accelerating-tube