中华超声影像学杂志
中華超聲影像學雜誌
중화초성영상학잡지
CHINESE JOURNAL OF ULTRASONOGRAPHY
2011年
3期
234-236
,共3页
李俊来%鲁媛媛%王国昌%鲁通%韩治宇%唐杰
李俊來%魯媛媛%王國昌%魯通%韓治宇%唐傑
리준래%로원원%왕국창%로통%한치우%당걸
超声检查%乳腺肿瘤%钙质沉着症%乳腺X线摄影术
超聲檢查%乳腺腫瘤%鈣質沉著癥%乳腺X線攝影術
초성검사%유선종류%개질침착증%유선X선섭영술
Ultrasonography%Breast neoplasms%Calcinosis%Mammography
目的 评价MicroPure辅助诊断技术对钙化超声显示率及提高乳腺癌诊断准确率的应用价值.方法 对我院63例患者的68个病灶行常规超声、X线钼靶及配有MicroPure辅助诊断技术的超声检查,分别描述病灶钙化及BI-RADS分级情况,全部病灶行穿刺活检或手术得以病理确诊.结果 ①以钼靶显示病灶钙化作为对照,常规超声的钙化显示率为80.6%,MicroPure技术的钙化显示率为100%.②常规超声、钼靶、MicroPure辅助诊断技术对诊断乳腺良恶性病灶的ROC曲线下面积分别为0.916、0.857和0.970.③常规超声BI-RADS分级为3级的3个病灶经MicroPure技术诊断为4级,病理结果为恶性.结论 MicroPure辅助诊断系统可以显著提高微小钙化的超声显示率,同时可以提高超声诊断乳腺癌的准确率,尤其对常规超声BI-RADS分级3~4级之间的患者,提高倾向性比率.
目的 評價MicroPure輔助診斷技術對鈣化超聲顯示率及提高乳腺癌診斷準確率的應用價值.方法 對我院63例患者的68箇病竈行常規超聲、X線鉬靶及配有MicroPure輔助診斷技術的超聲檢查,分彆描述病竈鈣化及BI-RADS分級情況,全部病竈行穿刺活檢或手術得以病理確診.結果 ①以鉬靶顯示病竈鈣化作為對照,常規超聲的鈣化顯示率為80.6%,MicroPure技術的鈣化顯示率為100%.②常規超聲、鉬靶、MicroPure輔助診斷技術對診斷乳腺良噁性病竈的ROC麯線下麵積分彆為0.916、0.857和0.970.③常規超聲BI-RADS分級為3級的3箇病竈經MicroPure技術診斷為4級,病理結果為噁性.結論 MicroPure輔助診斷繫統可以顯著提高微小鈣化的超聲顯示率,同時可以提高超聲診斷乳腺癌的準確率,尤其對常規超聲BI-RADS分級3~4級之間的患者,提高傾嚮性比率.
목적 평개MicroPure보조진단기술대개화초성현시솔급제고유선암진단준학솔적응용개치.방법 대아원63례환자적68개병조행상규초성、X선목파급배유MicroPure보조진단기술적초성검사,분별묘술병조개화급BI-RADS분급정황,전부병조행천자활검혹수술득이병리학진.결과 ①이목파현시병조개화작위대조,상규초성적개화현시솔위80.6%,MicroPure기술적개화현시솔위100%.②상규초성、목파、MicroPure보조진단기술대진단유선량악성병조적ROC곡선하면적분별위0.916、0.857화0.970.③상규초성BI-RADS분급위3급적3개병조경MicroPure기술진단위4급,병리결과위악성.결론 MicroPure보조진단계통가이현저제고미소개화적초성현시솔,동시가이제고초성진단유선암적준학솔,우기대상규초성BI-RADS분급3~4급지간적환자,제고경향성비솔.
Objective To evaluate the value of MicroPure software system in the display of calcification and diagnostic accuracy of breast carcinoma.Methods Sixty-three patients with 68 breast lesions were checked by conventional ultrasound,X-ray and MicroPure software technique.The calcification was described and the lesions were classified with BI-RADS.All cases were pathologically confirmed by biopsy or operation.Results ①Compared to the detection rate for calcification of X-ray,the display rate of conventional ultrasound was 80.6% ,and the display rate of MicroPure software technique was 100%.②The area under the curve of diagnosing benign and malignant of breast lesions was 0.916,0.857 and 0.970 respectively with ultrasound,X-ray and MicroPure software technique.③Three lesions was classified Ⅲ in BI-RADS with conventional ultrasound,and Ⅳ in BI-RADS with MicroPure software.The pathologic results were malignant.Conclusions MicroPure software system obviously raised the ultrasonic display rate of calcification,meanwhile,it raised the diagnostic accurate rate of breast cancer.It especially raised the tendency rate for the patients who were classified between Ⅲ and Ⅳ in BI-RADS.