Chinese Journal of Geriatrics
2010年
3期
200-202
,共3页
齐欣%郭继鸿%熊名琛%殷伟贤%杨茂勋
齊訢%郭繼鴻%熊名琛%慇偉賢%楊茂勛
제흔%곽계홍%웅명침%은위현%양무훈
超声心动,三维描记术%磁共振
超聲心動,三維描記術%磁共振
초성심동,삼유묘기술%자공진
Echocardiography,three-dimensional%Magnetic resonance imaging
目的 评价实时三维超声心动图测量老年患者左心室质量的准确性和重复性,并与磁共振对比.方法 选取因各种不同原因进行心脏磁共振检查的老年患者31例,同时进行实时三维超声心动图检查.实时三维超声心动图检查采用Philips iE-33型超声心动图仪,左心室质量的分析通过TomTec工作站用人工描记法完成,并与磁共振所得结果相比较.结果 实时三维超声心动图测量所得的左心室质量与磁共振测量结果比较,有良好的相关性.磁共振测量的左心室质量为74.4~208.1 g,平均(123.9±35.0)g.实时三维超声心动图测量的左心室质量为67.0~222.9 g,平均(128.6±37.9)g.二者所测结果相似(r=0.869,y=0.943 x+11.69,标准误为19.09 g),平均相差(4.7±37.7)g.在不同观察者间及观察者自身不同时间内测量的实时三维超声心动图结果显示良好的重复性.结论 实时三维超声心动图可以用于测量老年患者的左心室质量.与磁共振比较,实时三维超声心动图测量老年患者的左心室质量有较好的准确性和重复性.
目的 評價實時三維超聲心動圖測量老年患者左心室質量的準確性和重複性,併與磁共振對比.方法 選取因各種不同原因進行心髒磁共振檢查的老年患者31例,同時進行實時三維超聲心動圖檢查.實時三維超聲心動圖檢查採用Philips iE-33型超聲心動圖儀,左心室質量的分析通過TomTec工作站用人工描記法完成,併與磁共振所得結果相比較.結果 實時三維超聲心動圖測量所得的左心室質量與磁共振測量結果比較,有良好的相關性.磁共振測量的左心室質量為74.4~208.1 g,平均(123.9±35.0)g.實時三維超聲心動圖測量的左心室質量為67.0~222.9 g,平均(128.6±37.9)g.二者所測結果相似(r=0.869,y=0.943 x+11.69,標準誤為19.09 g),平均相差(4.7±37.7)g.在不同觀察者間及觀察者自身不同時間內測量的實時三維超聲心動圖結果顯示良好的重複性.結論 實時三維超聲心動圖可以用于測量老年患者的左心室質量.與磁共振比較,實時三維超聲心動圖測量老年患者的左心室質量有較好的準確性和重複性.
목적 평개실시삼유초성심동도측량노년환자좌심실질량적준학성화중복성,병여자공진대비.방법 선취인각충불동원인진행심장자공진검사적노년환자31례,동시진행실시삼유초성심동도검사.실시삼유초성심동도검사채용Philips iE-33형초성심동도의,좌심실질량적분석통과TomTec공작참용인공묘기법완성,병여자공진소득결과상비교.결과 실시삼유초성심동도측량소득적좌심실질량여자공진측량결과비교,유량호적상관성.자공진측량적좌심실질량위74.4~208.1 g,평균(123.9±35.0)g.실시삼유초성심동도측량적좌심실질량위67.0~222.9 g,평균(128.6±37.9)g.이자소측결과상사(r=0.869,y=0.943 x+11.69,표준오위19.09 g),평균상차(4.7±37.7)g.재불동관찰자간급관찰자자신불동시간내측량적실시삼유초성심동도결과현시량호적중복성.결론 실시삼유초성심동도가이용우측량노년환자적좌심실질량.여자공진비교,실시삼유초성심동도측량노년환자적좌심실질량유교호적준학성화중복성.
Objective To assess the accuracy and repeatability of real-time three-dimensional eehoeardiography (RT-3DE) for the measurements of left ventricular (LV) mass in elderly patients compared with magnetic resonance imaging (MRI).Methods A total of 31 patients (26 men) aged ≥60 years (72.1±6.6) years with various cardiovascular diseases were evaluated by MRI and RT- 3DE on the same day.RT-3DE was performed with a Philips iE-33 echocardiographic system, and LV mass was analyzed with the assistance of TomTec software.The results of LV mass obtained by manual tracing were compared with Signa 1.5-T MRI data.The acquisition of RT-3DE datasets was feasible in all 31 patients.Results A good correlation was observed between RT-3DE data with manual border detection and MRI for LV mass (r=0.869, y=0.943 x+11.69, SE= 19.09 g, P< 0.01), with a mean difference of (4.7±37.7)g.The average LV mass was (123.9±35.0)g (74.4-208.1 g) evaluated by MRI, and was (128.6±37.9)g (67.0-222.9 g) evaluated by RT-3DE.There was good inter- or intra-observer correlation between RT-3DE by two sonographers for LV mass.Conclusions The assessment of LV mass from RT-3DE data is feasible in elderly patients.The mass can be determined with high accuracy and low interobserver variability in elderly patients with adequate eehocardiographic image quality.