CHINESE JOURNAL OF PHYSICAL MEDICINE AND REHABILITATION
2012年
1期
34-37
,共4页
胡建平%伊文超%李瑞炎%蒋祥龙%易万龙
鬍建平%伊文超%李瑞炎%蔣祥龍%易萬龍
호건평%이문초%리서염%장상룡%역만룡
本体感觉%定量评估%可靠性%平均轨迹差
本體感覺%定量評估%可靠性%平均軌跡差
본체감각%정량평고%가고성%평균궤적차
Proprioception%Quantitative assessment%Reliability%Average track error
目的 研究Tecnobody康复系统定量评定本体感觉的可靠性.方法 健康人9例,其中男4例,女5例,年龄(22.8±0.68)岁,用Tecnobody按系统要求连续评定双足本体感觉3次.用重复测量标准误、简单相关系数、组内相关系数分析每次的耗时( Time)和平均轨迹差(ATE).结果 Time重复测量标准误左右分别为3.07~ 3.83和6.65~ 8.44,ATE左右分别为1.33 ~1.97和1.39 ~1.91;Time简单相关系数左右分别为0.919和0.62,ATE左右分别为0.808和0.831;Time组内相关系数左右分别为0.893和0.639,ATE左右分别为0.716和0.734.结论 该系统具备较好的可靠性,且评估不必重复多次,但其各参数设置及中国人标准参考值范围仍需进一步探究.
目的 研究Tecnobody康複繫統定量評定本體感覺的可靠性.方法 健康人9例,其中男4例,女5例,年齡(22.8±0.68)歲,用Tecnobody按繫統要求連續評定雙足本體感覺3次.用重複測量標準誤、簡單相關繫數、組內相關繫數分析每次的耗時( Time)和平均軌跡差(ATE).結果 Time重複測量標準誤左右分彆為3.07~ 3.83和6.65~ 8.44,ATE左右分彆為1.33 ~1.97和1.39 ~1.91;Time簡單相關繫數左右分彆為0.919和0.62,ATE左右分彆為0.808和0.831;Time組內相關繫數左右分彆為0.893和0.639,ATE左右分彆為0.716和0.734.結論 該繫統具備較好的可靠性,且評估不必重複多次,但其各參數設置及中國人標準參攷值範圍仍需進一步探究.
목적 연구Tecnobody강복계통정량평정본체감각적가고성.방법 건강인9례,기중남4례,녀5례,년령(22.8±0.68)세,용Tecnobody안계통요구련속평정쌍족본체감각3차.용중복측량표준오、간단상관계수、조내상관계수분석매차적모시( Time)화평균궤적차(ATE).결과 Time중복측량표준오좌우분별위3.07~ 3.83화6.65~ 8.44,ATE좌우분별위1.33 ~1.97화1.39 ~1.91;Time간단상관계수좌우분별위0.919화0.62,ATE좌우분별위0.808화0.831;Time조내상관계수좌우분별위0.893화0.639,ATE좌우분별위0.716화0.734.결론 해계통구비교호적가고성,차평고불필중복다차,단기각삼수설치급중국인표준삼고치범위잉수진일보탐구.
Objective To study the test-retest reliability in quantitative measurement of proprioception using Tecnobody rehabilitation system. Methods Nine healthy volunteers [4 males,5 females,averaged age (22.8 ±0.68) years] participated in three consecutive measurements on both feet by using Tecnobody rehabilitation system for computerized proprioceptive assessment.Standard error of measurement(SEM),correlation coefficient and intraclass correlation coefficient(ICC) obtained from the three consecutive measurements were used to analyze the "Time"used in each measuring session,and the "average track error" (ATE) of the measurements. Results SEM values of "Time" in left foot and right foot were 3.07 ~ 3.83 and 6.65 ~ 8.44 respectively.ATE values in left foot and right foot were 1.33 ~ 1.97 and 1.39 ~ 1.91 respectively.The "Time" and ATE correlation coefficients of left foot / right foot were 0.919/0.6 and 0.808/0.831,respectively.The "Time" ICC values were 0.893/0.639 for left foot / right foot,respectively ; and the "ATE" ICC values were 0.716/0.734 for left foot / right foot,respectively. Conclusion The Tecnobody rehabilitation system provided a fairly good reliability in both "relative" and " absolute" values in quantitative evaluation of proprioception in the feet.These data in a larger amount may be useful for setting up variables and the standard values of the local population for reference in proprioceptive rehabilitation.