目的 探讨听觉稳态反应(ASSR)在婴幼儿听力筛查中的应用.方法 对2007年12月至2008年12月本院听力门诊32例64耳耳声发射(OAE)无反应的婴幼儿行ASSR测试.结果 32例64耳经OAE无反应的婴幼儿经ASSR 0.5、1、2、4 kHz的电位引出率分别为3.13%(2/64)、6.26%(4/64)、7.81%(5164)、9.39%(6/64),P<0.05;反应阈分别为(80.26±6.68)、(85.68±5.45)、(84.32±4.32)、(86.56±4.26)dB HL,与OAE(26.58±4.52)dB HL比较,P<0.05.结论 ASSR有助于降低婴幼儿听力筛查的假阳性,对婴幼儿听力筛查有推广价值.
目的 探討聽覺穩態反應(ASSR)在嬰幼兒聽力篩查中的應用.方法 對2007年12月至2008年12月本院聽力門診32例64耳耳聲髮射(OAE)無反應的嬰幼兒行ASSR測試.結果 32例64耳經OAE無反應的嬰幼兒經ASSR 0.5、1、2、4 kHz的電位引齣率分彆為3.13%(2/64)、6.26%(4/64)、7.81%(5164)、9.39%(6/64),P<0.05;反應閾分彆為(80.26±6.68)、(85.68±5.45)、(84.32±4.32)、(86.56±4.26)dB HL,與OAE(26.58±4.52)dB HL比較,P<0.05.結論 ASSR有助于降低嬰幼兒聽力篩查的假暘性,對嬰幼兒聽力篩查有推廣價值.
목적 탐토은각은태반응(ASSR)재영유인은력사사중적응용.방법 대2007년12월지2008년12월본원은력문진32례64이이성발사(OAE)무반응적영유인행ASSR측시.결과 32례64이경OAE무반응적영유인경ASSR 0.5、1、2、4 kHz적전위인출솔분별위3.13%(2/64)、6.26%(4/64)、7.81%(5164)、9.39%(6/64),P<0.05;반응역분별위(80.26±6.68)、(85.68±5.45)、(84.32±4.32)、(86.56±4.26)dB HL,여OAE(26.58±4.52)dB HL비교,P<0.05.결론 ASSR유조우강저영유인은력사사적가양성,대영유인은력사사유추엄개치.
Objective To investigate the application of auditory steady state responses (ASSR) in the infants hearing screening. Methods ASSR was tested in 32 cases (64 ears) of infants with no reaction to otoacoustic emission (OAE) in our hospital between December 2007 and December 2008. Results No V wave was educed in the 32 infants (64 ears) with OAE, but the positive educing rates of ASSR were respectively 3.13%(2/64), 6.26%(4/64), 7.81%(5/64), 9.39%(6/64) to the frequencies of 0.5, 1, 2 and 4 kHz (P<0.05). The corresponding thresholds with ASSR were (80.26±6.68) , (85.68±5.45) , (84.32± 4.32) and (86.56±4.26) dB HL, as compared to (26.58±4.52) dB HL with OAE (P<0.05). Conclnsion ASSR is helpful to decrease the false positive rate in the infants hearing screening, and therefore appears to be worthwhile for widespread use.