目的 探讨3.0T MR单源与多源发射技术对肝脏成像质量的影响.方法 回顾性分析107例行3.0 T MR肝脏扫描受试者的图像,检查序列包括频谱选择性衰减反转恢复T2WI( SPAIRT2WI序列、DWI以及平衡式快速场回波(Balanced FFE)序列.每一序列均在单源和多源的激发模式下进行双重采集,采用Student'st检验比较单多源图像的均匀性与对比度,采用Wilcoxon秩和检验比较单多源图像质量的评分.2名放射科医师独立进行阅片,采用Kappa检验比较读片者之间的诊断一致性.结果 采用单源和多源发射技术测量图像的均匀性分别为418.40±66.75和416.26±50.61,多源发射技术图像均匀性优于单源发射技术,差异有统计学意义(=2.524,P<0.05).单源和多源发射技术测得的信噪比(SNR)分别为20.36±11.21和22.03±12.16,对比噪声比(CNR)分别为15.22±8.95和18.33±10.01,差异均有统计学意义(t值分别为- 2.630和- 4.238,P值均<0.05).多源发射技术图像对比度优于单源发射技术图像.SPAIR T2WI序列单源和多源发射技术图像质量评分分别为(1.40±0.42)和(1.81±0.27)分,DWI序列分别为(1.08±0.46)和(1.63±0.36)分,Balanced FFE序列分别为(0.95±0.45)和(1.65±0.37)分,多源发射技术图像质量在上述3个序列中均优于单源发射技术图像(Z值分别为-5.894、-5.801和-6.985,P值均<0.01).2名评判医师之间的诊断一致性好(Kappa值均>0.8,P值均<0.05).结论 相对于单源发射技术来说,多源发射技术能够消除电解质阴影,提供更好的图像均匀性、更高的图像对比度和更佳的图像质量,3.0T多源技术在腹部成像上具有很明显的优势.
目的 探討3.0T MR單源與多源髮射技術對肝髒成像質量的影響.方法 迴顧性分析107例行3.0 T MR肝髒掃描受試者的圖像,檢查序列包括頻譜選擇性衰減反轉恢複T2WI( SPAIRT2WI序列、DWI以及平衡式快速場迴波(Balanced FFE)序列.每一序列均在單源和多源的激髮模式下進行雙重採集,採用Student'st檢驗比較單多源圖像的均勻性與對比度,採用Wilcoxon秩和檢驗比較單多源圖像質量的評分.2名放射科醫師獨立進行閱片,採用Kappa檢驗比較讀片者之間的診斷一緻性.結果 採用單源和多源髮射技術測量圖像的均勻性分彆為418.40±66.75和416.26±50.61,多源髮射技術圖像均勻性優于單源髮射技術,差異有統計學意義(=2.524,P<0.05).單源和多源髮射技術測得的信譟比(SNR)分彆為20.36±11.21和22.03±12.16,對比譟聲比(CNR)分彆為15.22±8.95和18.33±10.01,差異均有統計學意義(t值分彆為- 2.630和- 4.238,P值均<0.05).多源髮射技術圖像對比度優于單源髮射技術圖像.SPAIR T2WI序列單源和多源髮射技術圖像質量評分分彆為(1.40±0.42)和(1.81±0.27)分,DWI序列分彆為(1.08±0.46)和(1.63±0.36)分,Balanced FFE序列分彆為(0.95±0.45)和(1.65±0.37)分,多源髮射技術圖像質量在上述3箇序列中均優于單源髮射技術圖像(Z值分彆為-5.894、-5.801和-6.985,P值均<0.01).2名評判醫師之間的診斷一緻性好(Kappa值均>0.8,P值均<0.05).結論 相對于單源髮射技術來說,多源髮射技術能夠消除電解質陰影,提供更好的圖像均勻性、更高的圖像對比度和更佳的圖像質量,3.0T多源技術在腹部成像上具有很明顯的優勢.
목적 탐토3.0T MR단원여다원발사기술대간장성상질량적영향.방법 회고성분석107례행3.0 T MR간장소묘수시자적도상,검사서렬포괄빈보선택성쇠감반전회복T2WI( SPAIRT2WI서렬、DWI이급평형식쾌속장회파(Balanced FFE)서렬.매일서렬균재단원화다원적격발모식하진행쌍중채집,채용Student'st검험비교단다원도상적균균성여대비도,채용Wilcoxon질화검험비교단다원도상질량적평분.2명방사과의사독립진행열편,채용Kappa검험비교독편자지간적진단일치성.결과 채용단원화다원발사기술측량도상적균균성분별위418.40±66.75화416.26±50.61,다원발사기술도상균균성우우단원발사기술,차이유통계학의의(=2.524,P<0.05).단원화다원발사기술측득적신조비(SNR)분별위20.36±11.21화22.03±12.16,대비조성비(CNR)분별위15.22±8.95화18.33±10.01,차이균유통계학의의(t치분별위- 2.630화- 4.238,P치균<0.05).다원발사기술도상대비도우우단원발사기술도상.SPAIR T2WI서렬단원화다원발사기술도상질량평분분별위(1.40±0.42)화(1.81±0.27)분,DWI서렬분별위(1.08±0.46)화(1.63±0.36)분,Balanced FFE서렬분별위(0.95±0.45)화(1.65±0.37)분,다원발사기술도상질량재상술3개서렬중균우우단원발사기술도상(Z치분별위-5.894、-5.801화-6.985,P치균<0.01).2명평판의사지간적진단일치성호(Kappa치균>0.8,P치균<0.05).결론 상대우단원발사기술래설,다원발사기술능구소제전해질음영,제공경호적도상균균성、경고적도상대비도화경가적도상질량,3.0T다원기술재복부성상상구유흔명현적우세.
Objective To investigate the role of dual-source parallel radio frequence (RF) and single-source excitation in liver imaging at 3.0 T MR.Methods This study was a retrospective analysis.One hundred and seven subjects underwent a 3.0 T TX MR scanning including axial spectrally selective attenuated inversion recovery (SPAIR) T2WI,axial DWI and coronal balanced-fast field echo( Balanced FFE).Each sequence was carried out with both single-source and dual-source RF excitation.Student's t test was used to compare the differences between single-source and dual-source RF excitation in the image uniformity,signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR).Wilcoxon signed rank test was used to determine whether there was difference between conventional and parallel transmission in the score of image quality.Reader agreement was assessed using the Cohen's Kappa test.Results For the image uniformity,there was significant difference between single-source and dual-source excitation (418.40 ± 66.75 for single-source vs.416.26 ± 50.61 for dual-source,t =2.524,P < 0.05 ).There also existed significant difference between single-source and dual-source excitation in SNR and CNR,respectively.The SNR and CNR of parallel transmission (22.03 + 12.16 and 18.33 ± 10.01,respectively) were both higher than those of single transmission (20.36 ± 11.21 and 15.22 ± 8.95,respectively) ( t =- 2.630,P < 0.05 for SNR and t =- 4.238,P < 0.05 for CNR).Image quality comparisons revealed significantly better results with dual-source than single-source RF excitation at SPAIR T2 WI ( 1.40 + 0.42 vs.1.81 ± 0.27 ),DWI ( 1.08 ± 0.40 vs.1.63 ± 0.36 ) and Balanced FFE sequence ( 0.95 ± 0.45 vs.1.65 ± 0.37,Z =- 5.894,- 5.801 and - 6.985,respectively,P < 0.01 ).In the comparison of image quality,the agreement between the two readers was very good ( Kappa > 0.8,P < 0.05 ).Conclusion Dual-source parallel RF excitation MR imaging in liver enables reducing dielectric shading,improving homogeneity of the RF magnetic induction field,and increasing SNR and CNR at 3.0 T.