Tomography,X-ray computed,cone beam%Tomography,X-ray computed%Esophageal neoplasms%Dosimetry
目的 探讨千伏级锥形束CT图像(KVCBCT)对食管癌放疗剂量计算的可行性.方法 比较Trilogy加速器采集的Catphan(R)600模体KVCBCT图像CT值稳定性及离轴曲线重合度并建立CT值-密度转换曲线用于剂量计算.在模体和10例食管癌患者常规CT上设计逆向调强计划,并将计划移植到相对应的KVCBCT图像上.基于KVCBCT的转换曲线计算靶区及周围组织受量,并与常规CT计算结果行配对t检验.结果 KVCBCT的CT值变化范围均在 1.6%内,且未发现有随时间变化趋势.常规CT和KVCBCT的CT值离轴曲线重合度较高,大部分成像区域差别在1%内.模体调强计划常规CT与KVCBCT所得剂量统计参数最大差别为1.33%,对食管癌患者凋强计划两者剂量差别在3.65%内.等中心层面剂量分布有较好一致性.结论 使用食管癌KVCBCT图像进行剂量计算是准确可行的,在临床上具有实际应用意义.
目的 探討韆伏級錐形束CT圖像(KVCBCT)對食管癌放療劑量計算的可行性.方法 比較Trilogy加速器採集的Catphan(R)600模體KVCBCT圖像CT值穩定性及離軸麯線重閤度併建立CT值-密度轉換麯線用于劑量計算.在模體和10例食管癌患者常規CT上設計逆嚮調彊計劃,併將計劃移植到相對應的KVCBCT圖像上.基于KVCBCT的轉換麯線計算靶區及週圍組織受量,併與常規CT計算結果行配對t檢驗.結果 KVCBCT的CT值變化範圍均在 1.6%內,且未髮現有隨時間變化趨勢.常規CT和KVCBCT的CT值離軸麯線重閤度較高,大部分成像區域差彆在1%內.模體調彊計劃常規CT與KVCBCT所得劑量統計參數最大差彆為1.33%,對食管癌患者凋彊計劃兩者劑量差彆在3.65%內.等中心層麵劑量分佈有較好一緻性.結論 使用食管癌KVCBCT圖像進行劑量計算是準確可行的,在臨床上具有實際應用意義.
목적 탐토천복급추형속CT도상(KVCBCT)대식관암방료제량계산적가행성.방법 비교Trilogy가속기채집적Catphan(R)600모체KVCBCT도상CT치은정성급리축곡선중합도병건립CT치-밀도전환곡선용우제량계산.재모체화10례식관암환자상규CT상설계역향조강계화,병장계화이식도상대응적KVCBCT도상상.기우KVCBCT적전환곡선계산파구급주위조직수량,병여상규CT계산결과행배대t검험.결과 KVCBCT적CT치변화범위균재 1.6%내,차미발현유수시간변화추세.상규CT화KVCBCT적CT치리축곡선중합도교고,대부분성상구역차별재1%내.모체조강계화상규CT여KVCBCT소득제량통계삼수최대차별위1.33%,대식관암환자조강계화량자제량차별재3.65%내.등중심층면제량분포유교호일치성.결론 사용식관암KVCBCT도상진행제량계산시준학가행적,재림상상구유실제응용의의.
Objective To investigate the dose calculation accuracy and feasibility of using kilovoltage cone-beam CT (KVCBCT) for esophageal cancer radiotherapy.Methods Hounsfield unit (HU) values and profile along the horizontal line of Catphan (R) 600 phantom in KVCBCT images acquired on Trilogy linear accelerator were compared to those in the planning CT.The KVCBCT value-density calibration curve was established.The intensity modulated radiotherapy plans were created on the planning CT images and copied to KVCBCT images.The dose distribution was recalculated by means of the KVCBCT value-density calibration curve in the treatment planning system.The dosimetric comparisons were performed between the KVCBCT and planning CT plans on the phantom and 10 patients with esophageal cancer.Results The KVCBCT value was stable with a maximum variation of 1.6%,and there was no significant time trend.CT value profiles showed good agreement within 1% variation except the peripheral regions.The dosimetric differences were less than 1.33% and 3.65% for the phantom case and the patient ones,respectively.The dose distribution comparison was also in good agreements.Conclusions The accurate dose caleulation based on KVCBCT for esophageal cancer is feasible.The KVCBCT images can be used for monitoring the dosimetric changes during the treatment.