中华放射肿瘤学杂志
中華放射腫瘤學雜誌
중화방사종류학잡지
CHINESE JOURNAL OF RADIATION ONCOLOGY
2008年
3期
230-233
,共4页
徐志勇%张良安%戴光复%苑淑渝
徐誌勇%張良安%戴光複%苑淑渝
서지용%장량안%대광복%원숙투
放射性支架%再狭窄%剂量点核函数%蒙特卡罗代码
放射性支架%再狹窄%劑量點覈函數%矇特卡囉代碼
방사성지가%재협착%제량점핵함수%몽특잡라대마
Radioactive stent%Restenosis%Dose kernel function%Monte Carlo code
目的 研究103Pd放射性支架的三维剂量率分布.方法 用模拟实验、理论计算和蒙特卡罗代码(MCNP4b)3种方法估算3 mm×13 mm支架的表面剂量、表面轴向剂量和径向分布,支架活度为2.9×106 Bq.用MCNP4b计算支架的三维剂量率分布.结果 模拟实验和MCNP4b计算的支架表面剂量分别为0.109、0.106 Gy,两者差异<3%.理论计算和MCNP4b计算的轴向剂量一致.3种方法计算的径向剂量也一致.用MCNP4b估算了103Pd放射性支架三维剂量率分布.结论 3种方法估算的103Pd放射性支架的剂量都比较准确.用MCNP4b计算的支架三维剂量率分布数据可为动物实验和临床应用提供剂量学参考.
目的 研究103Pd放射性支架的三維劑量率分佈.方法 用模擬實驗、理論計算和矇特卡囉代碼(MCNP4b)3種方法估算3 mm×13 mm支架的錶麵劑量、錶麵軸嚮劑量和徑嚮分佈,支架活度為2.9×106 Bq.用MCNP4b計算支架的三維劑量率分佈.結果 模擬實驗和MCNP4b計算的支架錶麵劑量分彆為0.109、0.106 Gy,兩者差異<3%.理論計算和MCNP4b計算的軸嚮劑量一緻.3種方法計算的徑嚮劑量也一緻.用MCNP4b估算瞭103Pd放射性支架三維劑量率分佈.結論 3種方法估算的103Pd放射性支架的劑量都比較準確.用MCNP4b計算的支架三維劑量率分佈數據可為動物實驗和臨床應用提供劑量學參攷.
목적 연구103Pd방사성지가적삼유제량솔분포.방법 용모의실험、이론계산화몽특잡라대마(MCNP4b)3충방법고산3 mm×13 mm지가적표면제량、표면축향제량화경향분포,지가활도위2.9×106 Bq.용MCNP4b계산지가적삼유제량솔분포.결과 모의실험화MCNP4b계산적지가표면제량분별위0.109、0.106 Gy,량자차이<3%.이론계산화MCNP4b계산적축향제량일치.3충방법계산적경향제량야일치.용MCNP4b고산료103Pd방사성지가삼유제량솔분포.결론 3충방법고산적103Pd방사성지가적제량도비교준학.용MCNP4b계산적지가삼유제량솔분포수거가위동물실험화림상응용제공제량학삼고.
Objective To investigate three-dimensional dose distribution for 103Pd radioactive stent.Methods The surface dose,the axial dose and radial dose in surface for 103pd stent (3 mm × 13 mm) were estimated by experimental simulating method, analytic function and MCNP4b code. Three-dimensional dose distribution was calculated by MCNP4b code. Results The surface dose of 103pd stent was 0. 109 and 0. 106 Gy estimated by experimental simulating method and MCNP4b code,between which the difference was less than 3%. The axial dose calculated by analytic function and MCNP4b code was well consistent,and so was the radial dose estimated by the three methods. Dose rate table were estimated by MCNP4b code. Conclnsions Dose distribution for 103 Pd stent estimated by the three methods is relatively accurate. Three-dimensional dose table estimated by MCNP4b may be used to calculate dose for 103Pd stent in animal experiment and clinical application.