中华放射肿瘤学杂志
中華放射腫瘤學雜誌
중화방사종류학잡지
CHINESE JOURNAL OF RADIATION ONCOLOGY
2010年
5期
465-467
,共3页
董洋%李丰彤%王境生%袁智勇%宋勇春%翁巍立%白广鑫
董洋%李豐彤%王境生%袁智勇%宋勇春%翁巍立%白廣鑫
동양%리봉동%왕경생%원지용%송용춘%옹외립%백엄흠
放射疗法,射波刀%金标数目%靶位置精度
放射療法,射波刀%金標數目%靶位置精度
방사요법,사파도%금표수목%파위치정도
Radiotherapy,cyberknife%Number of fiducials%Target localization accuracy
目的 通过硅橡胶模型模拟实验探讨置入软组织内金标数目对射波刀系统影像定位精度的影响.方法 模型内层为刚体结构,使用Accuray公司设计的球方模型;实验中用5个金标进行摆位以确定参考位置.模型外层为模拟人体软组织的非刚体性结构,使用双组分加成型硅橡胶;实验中用8个金标按不同数目的 排列组合获取影像定位数据,比较其与参考位置在平移参量和旋转参量方面差值以此探讨追踪不同数目金标对影像定位精度的影响.结果 追踪4个金标时平均每个金标对平移和旋转参量误差下降的贡献最大,分别为0.086 mm和0.033°.1~4个金标的平移参量误差下降了0.343 mm,4~8个金标的下降了0.077 mm.3~4个金标旋转参量误差下降了0.131°,4~8个金标的下降了0.090°.结论 追踪4个金标时平均每个金标对误差降低的贡献最大.金标置入4个以下时每多置1个金标对影像定位精度明显增加,置入4个以上时再增加置入金标数目对定位精度无明显改善.置入4个金标是同时兼顾降低患者痛苦和增加定位精度的一个适宜选择.
目的 通過硅橡膠模型模擬實驗探討置入軟組織內金標數目對射波刀繫統影像定位精度的影響.方法 模型內層為剛體結構,使用Accuray公司設計的毬方模型;實驗中用5箇金標進行襬位以確定參攷位置.模型外層為模擬人體軟組織的非剛體性結構,使用雙組分加成型硅橡膠;實驗中用8箇金標按不同數目的 排列組閤穫取影像定位數據,比較其與參攷位置在平移參量和鏇轉參量方麵差值以此探討追蹤不同數目金標對影像定位精度的影響.結果 追蹤4箇金標時平均每箇金標對平移和鏇轉參量誤差下降的貢獻最大,分彆為0.086 mm和0.033°.1~4箇金標的平移參量誤差下降瞭0.343 mm,4~8箇金標的下降瞭0.077 mm.3~4箇金標鏇轉參量誤差下降瞭0.131°,4~8箇金標的下降瞭0.090°.結論 追蹤4箇金標時平均每箇金標對誤差降低的貢獻最大.金標置入4箇以下時每多置1箇金標對影像定位精度明顯增加,置入4箇以上時再增加置入金標數目對定位精度無明顯改善.置入4箇金標是同時兼顧降低患者痛苦和增加定位精度的一箇適宜選擇.
목적 통과규상효모형모의실험탐토치입연조직내금표수목대사파도계통영상정위정도적영향.방법 모형내층위강체결구,사용Accuray공사설계적구방모형;실험중용5개금표진행파위이학정삼고위치.모형외층위모의인체연조직적비강체성결구,사용쌍조분가성형규상효;실험중용8개금표안불동수목적 배렬조합획취영상정위수거,비교기여삼고위치재평이삼량화선전삼량방면차치이차탐토추종불동수목금표대영상정위정도적영향.결과 추종4개금표시평균매개금표대평이화선전삼량오차하강적공헌최대,분별위0.086 mm화0.033°.1~4개금표적평이삼량오차하강료0.343 mm,4~8개금표적하강료0.077 mm.3~4개금표선전삼량오차하강료0.131°,4~8개금표적하강료0.090°.결론 추종4개금표시평균매개금표대오차강저적공헌최대.금표치입4개이하시매다치1개금표대영상정위정도명현증가,치입4개이상시재증가치입금표수목대정위정도무명현개선.치입4개금표시동시겸고강저환자통고화증가정위정도적일개괄의선택.
Objective To find a proper number of implanted fiducials in order to reconcile both more accurate image tacking and less pain for the patients.Methods The phantom was made of two parts of different materials.The inner part was a ball-cube phantom developed by Accuray Corporation, which could be regarded as a rigid body, with 5 fiducials inside using as the reference to align the whole phantom in the experiment.The outer part was made of additional silicone rubber with two components to simulate soft tissue as a non-rigid body, which was implanted with 8 fiducials inside.All combinations of different number of fiducials were tracked to aquire the target location information, which were then compared with the reference position that was set by the inner 5 fiducials to obtain the deviations of the translation and the rotation parameters.Thus the impact of the number of fiducials on image target localization accuracy could be analyzed easily.Results When 4 fiducials are used for tracking, the decline of translation and rotation errors for every fiducial is largest (0.086 mm and 0.033°).The translation error was decreased by 0.343 mm from 1 to 4 fiducials, while by only 0.077 mm from 4 to 8 fiducials;the same decline was observed for rotation errors (0.131° for 3 to 4 fiducials , and 0.09° for 4 to 8 fiducials , respectively) .Conclusions When 4 fiducials are used for tracking, the decline of errors for every fiducial is maximum.When one more fiducial is adding, the image target localization accuracy can be enhanced obviously from 1 to 4 fiducials, but not as obvious for more than 4 fiducials.