超声乳化%人工晶状体%白内障%预装式
超聲乳化%人工晶狀體%白內障%預裝式
초성유화%인공정상체%백내장%예장식
Phacoemulsification%Lenses,Intraocular%cataract%preloaded
目的 评价KS-X型丙烯酸酯预装式人工晶状体植入术的安全性、有效性.方法 对92例(100只眼)老年性白内障患者行超声乳化白内障摘除术,采用预装式推注器植入KS-X型丙烯酸酯人工晶状体.观察患者术后视力、超声乳化时间、人工晶状体植入时间、角膜内皮细胞丢失率、散光变化及手术并发症.结果 术后1周、1月、3月裸眼视力≥0.5的患者所占比例分别为91%,94%,97%;术后3个月矫正视力≥0.8为84%.超声乳化时间为(26.45±13.95)s,平均超声能量为(12.71±8.21)%,人工晶状体植入时间(12.31±3.41)s.角膜内皮细胞丢失率为1 1.92%,术后1周、1月、3月的手术源性散光分别为(1.14±0.29)D、(0.89±0.31)D、(0.76±0.36)D.所有患者均手术顺利,未出现术中并发症.结论 KS-X型丙烯酸酯预装式人工晶状体植入术是安全有效的,其植入过程具有简单、快速、安全、清洁的优点,值得推广.
目的 評價KS-X型丙烯痠酯預裝式人工晶狀體植入術的安全性、有效性.方法 對92例(100隻眼)老年性白內障患者行超聲乳化白內障摘除術,採用預裝式推註器植入KS-X型丙烯痠酯人工晶狀體.觀察患者術後視力、超聲乳化時間、人工晶狀體植入時間、角膜內皮細胞丟失率、散光變化及手術併髮癥.結果 術後1週、1月、3月裸眼視力≥0.5的患者所佔比例分彆為91%,94%,97%;術後3箇月矯正視力≥0.8為84%.超聲乳化時間為(26.45±13.95)s,平均超聲能量為(12.71±8.21)%,人工晶狀體植入時間(12.31±3.41)s.角膜內皮細胞丟失率為1 1.92%,術後1週、1月、3月的手術源性散光分彆為(1.14±0.29)D、(0.89±0.31)D、(0.76±0.36)D.所有患者均手術順利,未齣現術中併髮癥.結論 KS-X型丙烯痠酯預裝式人工晶狀體植入術是安全有效的,其植入過程具有簡單、快速、安全、清潔的優點,值得推廣.
목적 평개KS-X형병희산지예장식인공정상체식입술적안전성、유효성.방법 대92례(100지안)노년성백내장환자행초성유화백내장적제술,채용예장식추주기식입KS-X형병희산지인공정상체.관찰환자술후시력、초성유화시간、인공정상체식입시간、각막내피세포주실솔、산광변화급수술병발증.결과 술후1주、1월、3월라안시력≥0.5적환자소점비례분별위91%,94%,97%;술후3개월교정시력≥0.8위84%.초성유화시간위(26.45±13.95)s,평균초성능량위(12.71±8.21)%,인공정상체식입시간(12.31±3.41)s.각막내피세포주실솔위1 1.92%,술후1주、1월、3월적수술원성산광분별위(1.14±0.29)D、(0.89±0.31)D、(0.76±0.36)D.소유환자균수술순리,미출현술중병발증.결론 KS-X형병희산지예장식인공정상체식입술시안전유효적,기식입과정구유간단、쾌속、안전、청길적우점,치득추엄.
Objective To evaluate the safety and effectiveness of KS-X acrylic preloaded intraocular lens implantation. Methods 92 patients (100eyes) with age-related cataract were performed with phacoemulsification and KS-X acrylic preloaded intraocular lens implantation. The visual acuity, phacoemulsification time,lens implantation time, the endothelial cell loss percentage, astigmatism and operative complications were observed. Results The visual acuity equal or better than 0.5 were 91%, 94% and 97% one day, one month and three month after surgery respectively. The best corrected visual acuity equal or better than 0.8 three month after surgery was 84%. The phacoemulsification time was 26.45± 13.95sec and mean phacoemulsification energy was 12.71 ± 8.21%. The lens implantation time was 12.31±3.41 sec. The endothelial cell loss percentage was 11.92% one month after surgery. The surgery induced astigmatism were 1.14± 0.29D, 0.89± 0.31 D and 0.76± 0.36 D one week, one month and three month after surgery respectively. There was no operative complication during surgery. Conclusions The KS-X acrylic preloaded intraocular lens implantation are safe and effective. The preloaded injector system is simple, quick, safe and clean.