提出了一种非侵入式的心肺门控技术以改善图像分辨率.该方法基于动物的心肺运动实时地产生门控信号,并采用改进的零交叉数法实现对R波的检测,通过热电偶感受动物鼻气流的温度变化以判断其呼吸时相.该门控技术即使对0.03 kg左右的小鼠,也能准确产生其在自由呼吸状态下的门控信号.除了非侵人性,与其他现存的门控技术相比,占用更少的空间和成本,可应用于微型CT系统和其他任何需要检测动物心肺运动的系统.几组测试结果表明,所提出的心肺门控能够产生所需信号,并改善图像分辨率.
提齣瞭一種非侵入式的心肺門控技術以改善圖像分辨率.該方法基于動物的心肺運動實時地產生門控信號,併採用改進的零交扠數法實現對R波的檢測,通過熱電偶感受動物鼻氣流的溫度變化以判斷其呼吸時相.該門控技術即使對0.03 kg左右的小鼠,也能準確產生其在自由呼吸狀態下的門控信號.除瞭非侵人性,與其他現存的門控技術相比,佔用更少的空間和成本,可應用于微型CT繫統和其他任何需要檢測動物心肺運動的繫統.幾組測試結果錶明,所提齣的心肺門控能夠產生所需信號,併改善圖像分辨率.
제출료일충비침입식적심폐문공기술이개선도상분변솔.해방법기우동물적심폐운동실시지산생문공신호,병채용개진적령교차수법실현대R파적검측,통과열전우감수동물비기류적온도변화이판단기호흡시상.해문공기술즉사대0.03 kg좌우적소서,야능준학산생기재자유호흡상태하적문공신호.제료비침인성,여기타현존적문공기술상비,점용경소적공간화성본,가응용우미형CT계통화기타임하수요검측동물심폐운동적계통.궤조측시결과표명,소제출적심폐문공능구산생소수신호,병개선도상분변솔.
Non-invasive cardiac-pulmonary gating is proposed to improve the imaging resolution.It produces signals based on the cardiac-pulmonary motion of an animal in real-time.The system with the non-invasive gating consists of a digital signal processor (DSP),an electrocardiography (ECG) detection circuit and a thermocouple circuit.An enhanced R wave detection algorithm based on zero crossing counts is used to adjust the low sample frequency associated with the respiratory rate of an animal.The thermocouple recognizes the respiration phase by sensing the temperature changes of the nasal airflow of an animal.The proposed gating can accurately generate the gating signal for freely breathing mice (weight of around 0.03 kg),and its respiratory signal is too weak to be detected.Apart from non-invasiveness,compared with other existing gating techniques,it occupies minimal space at lower cost.Actually,it can be used in micro-computed tomography (CT) and other systems needed to detect the cardiac-pulmonary motion.Several tests validate that the proposed cardiac-pulmonary gating can generate the gating signal as required.By using the gating technique,the image resolution is improved.