上气道%数字化摄影%管电压%辐射剂量%图像质量
上氣道%數字化攝影%管電壓%輻射劑量%圖像質量
상기도%수자화섭영%관전압%복사제량%도상질량
Upper airway%Digital radiography%Tube voltage%Radiation dose%Image quality
目的 探讨成人上气道DR管电压选择与辐射剂量和图像质量的关系.方法 用聚甲基丙烯酸甲酯( PM MA)模拟成人上气道厚度和对比度细节体模(CDRAD2.O)组合,以不同管电压自动曝光控制系统(AEC)摄影,记录入射体表剂量(ESD)、剂量面积乘积(DAP)和曝光量(mAs),计算体模图像质量因子(IQF值),选定适宜管电压.结果 ESD、DAP、mAs随着管电压升高而下降,IQF值却增大,其间差异有统计学意义(F =45.15、26.41、29.26、56.53,P<0.05).75 kV以下ESD、DAP、mAs明显增大,75 kV以上逐渐降低,75~80 kV趋于平衡.50~75 kV之间IQF值差异无统计学意义,75~90 kV之间IQF值差异有统计学意义(F=11.35,P<0.05).不同管电压的正常人体上气道的图像质量无明显差异.结论 成人上气道DR的适宜管电压范围为75~80 kV,用IQF值对照图像质量评分可为临床评价图像质量提供依据.
目的 探討成人上氣道DR管電壓選擇與輻射劑量和圖像質量的關繫.方法 用聚甲基丙烯痠甲酯( PM MA)模擬成人上氣道厚度和對比度細節體模(CDRAD2.O)組閤,以不同管電壓自動曝光控製繫統(AEC)攝影,記錄入射體錶劑量(ESD)、劑量麵積乘積(DAP)和曝光量(mAs),計算體模圖像質量因子(IQF值),選定適宜管電壓.結果 ESD、DAP、mAs隨著管電壓升高而下降,IQF值卻增大,其間差異有統計學意義(F =45.15、26.41、29.26、56.53,P<0.05).75 kV以下ESD、DAP、mAs明顯增大,75 kV以上逐漸降低,75~80 kV趨于平衡.50~75 kV之間IQF值差異無統計學意義,75~90 kV之間IQF值差異有統計學意義(F=11.35,P<0.05).不同管電壓的正常人體上氣道的圖像質量無明顯差異.結論 成人上氣道DR的適宜管電壓範圍為75~80 kV,用IQF值對照圖像質量評分可為臨床評價圖像質量提供依據.
목적 탐토성인상기도DR관전압선택여복사제량화도상질량적관계.방법 용취갑기병희산갑지( PM MA)모의성인상기도후도화대비도세절체모(CDRAD2.O)조합,이불동관전압자동폭광공제계통(AEC)섭영,기록입사체표제량(ESD)、제량면적승적(DAP)화폭광량(mAs),계산체모도상질량인자(IQF치),선정괄의관전압.결과 ESD、DAP、mAs수착관전압승고이하강,IQF치각증대,기간차이유통계학의의(F =45.15、26.41、29.26、56.53,P<0.05).75 kV이하ESD、DAP、mAs명현증대,75 kV이상축점강저,75~80 kV추우평형.50~75 kV지간IQF치차이무통계학의의,75~90 kV지간IQF치차이유통계학의의(F=11.35,P<0.05).불동관전압적정상인체상기도적도상질량무명현차이.결론 성인상기도DR적괄의관전압범위위75~80 kV,용IQF치대조도상질량평분가위림상평개도상질량제공의거.
Objective To investigate the relationship between the tube voltage and radiation dose as well as image quality in adult upper airway digital radiography (DR).Methods We used CDRAD2.0 contrast details phantom and PMMA to simulate adult upper airway. With different tube voltages,the phantom was exposed using automatic exposure control system (AEC).The entrance surface dose( ESD),dose area product(DAP) and mAs in every exposures were recorded.The image quality factors(IQF) of all images were calculated. Results With tube voltage increasing,ESD,DAP,mAs decreased and IQF value enhanced.There were statistically significance ( F =45.15,26.41,29.26,56.53,P < 0.05 ).ESD,DAP,mAs significantly increased with tube voltage less than 75 kV,began to decrease when tube voltage more than 75 kV,tended to be on the balance in 75 -80 kV.At the same time,the fluctuation of IQF value was no statistical difference in 50 - 75 kV of tube voltage,but was statistical significance in 75 -90 kV( F =11.35,P <0.05 ).So,the image quality of adult upper airway with different tube voltage had no significant difference.Conclusions The appropriate tube voltage was 75 kV to 80 kV in the upper airway DR.The IQF value can be provided as the clinical evaluation index of image quality.