上海交通大学学报(医学版)
上海交通大學學報(醫學版)
상해교통대학학보(의학판)
JOURNAL OF SHANGHAI JIAOTONG UNIVERSITY(MEDICAL SCIENCE)
2010年
1期
76-79
,共4页
陆玮%季刚%陈舜杰%吴谷奋%蒋更如
陸瑋%季剛%陳舜傑%吳穀奮%蔣更如
륙위%계강%진순걸%오곡강%장경여
维持性血液透析%全身生物电阻抗分析%干体质量%细胞内液%细胞外液
維持性血液透析%全身生物電阻抗分析%榦體質量%細胞內液%細胞外液
유지성혈액투석%전신생물전조항분석%간체질량%세포내액%세포외액
maintenance hemodialysis%bioelectrical impedance analysis%dry weight%intra-cellular water%extra-cellular water
目的 采用全身生物电阻抗分析(BIA)法评估维持性血液透析(MHD)患者水平衡动态变化及干体质量.方法 选择血液透析龄≥1年且病情稳定的20例MHD患者作为研究对象(MHD组),分别于血液透析前后及透析间期(透析后24 h和48 h)采用BIA法测定细胞外液量(ECW)和细胞内液量(ICW),计算并获得相关指标;分析MHD患者0h达到干体质量与否及其平均每日尿量对ECW和ICW的影响.以20名健康志愿者作为正常对照组.结果 与正常对照组比较,MHD组透析前理想体质量纠正的ECW明显升高(P<0.05).在MHD组中,血液透析后ECW较透析前明显下降(P<0.05);与透析前和透析间期比较,血液透析后ECW/机体总水量(TBW)降低,ICW/TBW升高,差异均有统计学意义(P<0.05);各时间点ICW和ECW与患者相应体质量显著相关;达到干体质量患者0 h和24 h时间点理想体质量纠正的ICW和ECW均明显高于未达到干体质量患者(P<0.05);尿量<400 mL/d的患者透析前和透析后24 h,ECW和ICW均显著低于尿量≥400 mL/d的患者(P<0.05).结论 BIA法研究提示,MHD患者血液透析前后及透析间期水平衡动态变化以ECW为主,评价指标以ECW/TBW较为敏感;理想体质量纠正的ICW和ECW可能是评判干体质量的敏感指标.
目的 採用全身生物電阻抗分析(BIA)法評估維持性血液透析(MHD)患者水平衡動態變化及榦體質量.方法 選擇血液透析齡≥1年且病情穩定的20例MHD患者作為研究對象(MHD組),分彆于血液透析前後及透析間期(透析後24 h和48 h)採用BIA法測定細胞外液量(ECW)和細胞內液量(ICW),計算併穫得相關指標;分析MHD患者0h達到榦體質量與否及其平均每日尿量對ECW和ICW的影響.以20名健康誌願者作為正常對照組.結果 與正常對照組比較,MHD組透析前理想體質量糾正的ECW明顯升高(P<0.05).在MHD組中,血液透析後ECW較透析前明顯下降(P<0.05);與透析前和透析間期比較,血液透析後ECW/機體總水量(TBW)降低,ICW/TBW升高,差異均有統計學意義(P<0.05);各時間點ICW和ECW與患者相應體質量顯著相關;達到榦體質量患者0 h和24 h時間點理想體質量糾正的ICW和ECW均明顯高于未達到榦體質量患者(P<0.05);尿量<400 mL/d的患者透析前和透析後24 h,ECW和ICW均顯著低于尿量≥400 mL/d的患者(P<0.05).結論 BIA法研究提示,MHD患者血液透析前後及透析間期水平衡動態變化以ECW為主,評價指標以ECW/TBW較為敏感;理想體質量糾正的ICW和ECW可能是評判榦體質量的敏感指標.
목적 채용전신생물전조항분석(BIA)법평고유지성혈액투석(MHD)환자수평형동태변화급간체질량.방법 선택혈액투석령≥1년차병정은정적20례MHD환자작위연구대상(MHD조),분별우혈액투석전후급투석간기(투석후24 h화48 h)채용BIA법측정세포외액량(ECW)화세포내액량(ICW),계산병획득상관지표;분석MHD환자0h체도간체질량여부급기평균매일뇨량대ECW화ICW적영향.이20명건강지원자작위정상대조조.결과 여정상대조조비교,MHD조투석전이상체질량규정적ECW명현승고(P<0.05).재MHD조중,혈액투석후ECW교투석전명현하강(P<0.05);여투석전화투석간기비교,혈액투석후ECW/궤체총수량(TBW)강저,ICW/TBW승고,차이균유통계학의의(P<0.05);각시간점ICW화ECW여환자상응체질량현저상관;체도간체질량환자0 h화24 h시간점이상체질량규정적ICW화ECW균명현고우미체도간체질량환자(P<0.05);뇨량<400 mL/d적환자투석전화투석후24 h,ECW화ICW균현저저우뇨량≥400 mL/d적환자(P<0.05).결론 BIA법연구제시,MHD환자혈액투석전후급투석간기수평형동태변화이ECW위주,평개지표이ECW/TBW교위민감;이상체질량규정적ICW화ECW가능시평판간체질량적민감지표.
Objective To evaluate the fluid balance and dry weight with whole body bioelectrical impedance analysis (BIA) in patients with maintenance hemodialysis ( MHD). Methods Twenty patients in stable condition with MHD for more than one year were selected ( MHD group). Extra-cellular water ( ECW) and intra-cellular water (ICW) were measured by BIA before hemodialysis, after hemodialysis and at intervals of hemodialysis (24 h and 48 h after hemodialysis), and related parameters were calculated. Whether patients with MHD achieved dry weight at 0 h was analysed, and the effects of average daily urine volume on ECW and ICW were explored. Another 20 healthy volunteers were served as controls. Results Compared with control group, ECW retrieved by ideal body weight was significantly higher before hemodialysis in MHD group (P<0.05). In MHD group, ECW was significantly lower after hemodialysis than that before hemodialysis (P < 0.05). ECW/total body water (TBW) was significantly lower and ICW/TBW was significantly higher after hemodialysis than those before hemodialysis and at intervals of hemodialysis ( P < 0.05). ICW and ECW at each time point were significantly related to body weight. ICW and ECW retrieved by ideal body weight at 0 h and 24 h of those who achieved dry weight were significantly higher than those who did not achieve dry weight ( P < 0.05). ECW and ICW before hemodialysis and 24 h after hemodialysis were significantly lower in those with urine volume <400 mL/d than those with urine volume ≥400 mL/d (P <0.05). Conclusion BIA analysis indicates that changes of fluid balance before hemodialysis, after hemodialysis and at intervals of hemodialysis in patients with MHD are characterized by ECW, and ECW/TBW is a more sensitive parameter. ICW and ECW retrieved by ideal body weight may be sensitive parameters to determine the dry weight in patients with MHD.