胃肠灯丝预热故障%谐振式磁饱和稳压器%灯丝%高压交换闸
胃腸燈絲預熱故障%諧振式磁飽和穩壓器%燈絲%高壓交換閘
위장등사예열고장%해진식자포화은압기%등사%고압교환갑
Gastrointestinal filament preheating breakdown%Resonant type magnetic saturation manostat%Filament%The high pressure switching gear
针对XG501A型程控机开机后报E02故障的现象进行分析,其解决方案根据故障提示,对胃肠灯丝预热部分进行检修;并根据电路的工作流程进行分析查找故障所在。在检修过程中应考虑全面、逐步测量数据,不可冒然行事,避免维修时走弯路以及损坏设备。
針對XG501A型程控機開機後報E02故障的現象進行分析,其解決方案根據故障提示,對胃腸燈絲預熱部分進行檢脩;併根據電路的工作流程進行分析查找故障所在。在檢脩過程中應攷慮全麵、逐步測量數據,不可冒然行事,避免維脩時走彎路以及損壞設備。
침대XG501A형정공궤개궤후보E02고장적현상진행분석,기해결방안근거고장제시,대위장등사예열부분진행검수;병근거전로적공작류정진행분석사조고장소재。재검수과정중응고필전면、축보측량수거,불가모연행사,피면유수시주만로이급손배설비。
It reports the E02 breakdown after XG501A program controller starting. The experiences and solutions of the E02 breakdown. According to the breakdown prompt, we carry on the analysis and the overhaul to the gastrointestinal filament preheating circuit. It is able to find the breakdown surely to be at, if you carried on the analysis search according to electric circuit's work flow. You must consider that comprehensive and gradually metrical data in the overhaul process. You cannot take for granted and only then may avoid detouring and equipment's damage.