OCULUS角膜生物力学分析仪%Goldmann压平眼压计%眼压%中央角膜厚度%一致性分析
OCULUS角膜生物力學分析儀%Goldmann壓平眼壓計%眼壓%中央角膜厚度%一緻性分析
OCULUS각막생물역학분석의%Goldmann압평안압계%안압%중앙각막후도%일치성분석
OCULUS Corvis ST corneal analyzer%Goldmann applanation tonomery%Intraocular Pressure%Central corneal thickness%Bland-Altman analysis
目的 探讨在不同的中央角膜厚度(CCT)下OCULUS Corvis ST角膜生物力学分析仪(OCS)与Goldmann压平眼压计(GAT)眼压(IOP)测量值的相关性,评价两者测量值的一致性及临床上两种眼压计测量值相互替换的可能性,为临床应用提供参考.方法 非青光眼病例32例(64只眼)分别用OCS和GAT进行眼压测量,两种仪器的测量顺序随机.同时用OCS测量CCT.OCS与GAT眼压测量值的相关性采用配对t检验,Bland-Altman分析法评价两种仪器IOP测量值的一致性.结果 在不同角膜厚度情况下,OCS与GAT的IOP测量值差异无统计学意义(CCT≤520.μm,n=12,t =-0.15,P>0.05;520 μm<CCT≤580 μm, n=44,t=0.12,P>0.05;CCT>580 μm,n=8,t =0.59,P>0.05).结论 两种方法测量的眼压值差异无统计学意义,且具有较好的相关性,在临床上两种仪器测得的IOP值可以相互替换.
目的 探討在不同的中央角膜厚度(CCT)下OCULUS Corvis ST角膜生物力學分析儀(OCS)與Goldmann壓平眼壓計(GAT)眼壓(IOP)測量值的相關性,評價兩者測量值的一緻性及臨床上兩種眼壓計測量值相互替換的可能性,為臨床應用提供參攷.方法 非青光眼病例32例(64隻眼)分彆用OCS和GAT進行眼壓測量,兩種儀器的測量順序隨機.同時用OCS測量CCT.OCS與GAT眼壓測量值的相關性採用配對t檢驗,Bland-Altman分析法評價兩種儀器IOP測量值的一緻性.結果 在不同角膜厚度情況下,OCS與GAT的IOP測量值差異無統計學意義(CCT≤520.μm,n=12,t =-0.15,P>0.05;520 μm<CCT≤580 μm, n=44,t=0.12,P>0.05;CCT>580 μm,n=8,t =0.59,P>0.05).結論 兩種方法測量的眼壓值差異無統計學意義,且具有較好的相關性,在臨床上兩種儀器測得的IOP值可以相互替換.
목적 탐토재불동적중앙각막후도(CCT)하OCULUS Corvis ST각막생물역학분석의(OCS)여Goldmann압평안압계(GAT)안압(IOP)측량치적상관성,평개량자측량치적일치성급림상상량충안압계측량치상호체환적가능성,위림상응용제공삼고.방법 비청광안병례32례(64지안)분별용OCS화GAT진행안압측량,량충의기적측량순서수궤.동시용OCS측량CCT.OCS여GAT안압측량치적상관성채용배대t검험,Bland-Altman분석법평개량충의기IOP측량치적일치성.결과 재불동각막후도정황하,OCS여GAT적IOP측량치차이무통계학의의(CCT≤520.μm,n=12,t =-0.15,P>0.05;520 μm<CCT≤580 μm, n=44,t=0.12,P>0.05;CCT>580 μm,n=8,t =0.59,P>0.05).결론 량충방법측량적안압치차이무통계학의의,차구유교호적상관성,재림상상량충의기측득적IOP치가이상호체환.
Objective To investigate the correlation of intraocular pressure (IOP) obtained by OCULUS Corvis ST comeal biomechanics analyzer (OCS) and Goldmann planation tonometry (GAT)in non-glaucomatous for central corneal thickness (CCT) and to evaluate the agreement and the interchangeablity of the IOP values obtained by the two instrument.Methods In a randomized order two or three consecutive IOP measurements were performed on 64 eyes of 32 non-glaucomatous subjects by one examiner using the two instruments.Furthermore,the CCT was obtained by OCULUS Corvis ST corneal analyzer.The matched t-test was determined to compare IOP readings between OCS and GAT.CCT and OCS/GAT IOP values.Bland-Altman analysis was applied to evaluate the agreement between the instruments.Results Mean IOP values obtained by both instruments were no significantly difference in different central corneal thickness (CCT<520m,n =12,t =-0.15,P >0.05; 520m<CCT<580m,n =44,t =0.12,P >0.05; CCT>580□m,n =8,t =0.59,P >0.05).Conclusions There are no significantly difference among the two instruments,and the correlation of them are high.They may replace each other in some kind of degree.