临床肺科杂志
臨床肺科雜誌
림상폐과잡지
JOUNAL OF CLINICAL PULMONARY MEDICINE
2014年
9期
1559-1562
,共4页
李春华%吕圣秀%戴欣%何颖竹%赵攀
李春華%呂聖秀%戴訢%何穎竹%趙攀
리춘화%려골수%대흔%하영죽%조반
低剂量%CT%AIDS%PCP%体质量低
低劑量%CT%AIDS%PCP%體質量低
저제량%CT%AIDS%PCP%체질량저
chest%low dose%computed tomography%AIDS%PCP
目的:探讨低剂量CT扫描在AIDS合并PCP体质量低者中的应用可行性及临床价值。方法将69例AIDS合并PCP体质量低患者,随机分成3组,分别采用管电压固定为120kV,常规剂量(自动mAs)与低剂量扫描,低剂量扫描管电流分别采用40mAs,30mAs,20mAs,对图像质量、辐射剂量进行分析,得出AIDS合并PCP体质量低患者的最优低剂量扫描方案。结果常规剂量扫描与低剂量扫描获得图像均能达到影像诊断要求。采用管电压固定为120kV,管电流为常规剂量、40 mAs、30 mAs、20 mAs组获得的平均辐射剂量CTDIvol分别为:6.65mGy、2.06 mGy、1.54 mGy、1.03 mGy,管电流20 mAs组最低,为常规剂量的15.5%。结论采用管电压kV值固定为120kV,管电流20mAs胸部低剂量CT扫描在AIDS合并PCP体质量低患者中图像质量能满足临床诊断要求,同时降低辐射剂量,值得推广。
目的:探討低劑量CT掃描在AIDS閤併PCP體質量低者中的應用可行性及臨床價值。方法將69例AIDS閤併PCP體質量低患者,隨機分成3組,分彆採用管電壓固定為120kV,常規劑量(自動mAs)與低劑量掃描,低劑量掃描管電流分彆採用40mAs,30mAs,20mAs,對圖像質量、輻射劑量進行分析,得齣AIDS閤併PCP體質量低患者的最優低劑量掃描方案。結果常規劑量掃描與低劑量掃描穫得圖像均能達到影像診斷要求。採用管電壓固定為120kV,管電流為常規劑量、40 mAs、30 mAs、20 mAs組穫得的平均輻射劑量CTDIvol分彆為:6.65mGy、2.06 mGy、1.54 mGy、1.03 mGy,管電流20 mAs組最低,為常規劑量的15.5%。結論採用管電壓kV值固定為120kV,管電流20mAs胸部低劑量CT掃描在AIDS閤併PCP體質量低患者中圖像質量能滿足臨床診斷要求,同時降低輻射劑量,值得推廣。
목적:탐토저제량CT소묘재AIDS합병PCP체질량저자중적응용가행성급림상개치。방법장69례AIDS합병PCP체질량저환자,수궤분성3조,분별채용관전압고정위120kV,상규제량(자동mAs)여저제량소묘,저제량소묘관전류분별채용40mAs,30mAs,20mAs,대도상질량、복사제량진행분석,득출AIDS합병PCP체질량저환자적최우저제량소묘방안。결과상규제량소묘여저제량소묘획득도상균능체도영상진단요구。채용관전압고정위120kV,관전류위상규제량、40 mAs、30 mAs、20 mAs조획득적평균복사제량CTDIvol분별위:6.65mGy、2.06 mGy、1.54 mGy、1.03 mGy,관전류20 mAs조최저,위상규제량적15.5%。결론채용관전압kV치고정위120kV,관전류20mAs흉부저제량CT소묘재AIDS합병PCP체질량저환자중도상질량능만족림상진단요구,동시강저복사제량,치득추엄。
Objective To investigate the application feasibility and clinical value of low-dose CT scans in AIDS patients comp-licated with low PCP body mass. Methods 69 AIDS patients with low PCP body mass were ran-domly divided into 3 groups and scanned respectively with a tube voltage fixed at 120kV, conventional dose (automat-ic mAs) and low-dose, with the low-dose scanning tube current being 40mAs, 30mAs and 20mAs respectively. The image quality and radiation dose were analyzed in order to gain the optimal low dose scanning solution in AIDS pa-tients complicated with low PCP body mass. Results The images obtained by both conventional dose scanning and low-dose scanning could meet the requirements of imaging diagnosis. The average radiation CTDIvol obtained in the group using the fixed tube voltage 120kV, with a tube current of the conventional dose, 40 mAs, 30 mAs and 20 mAs was respectively:6. 65mGy, 2. 06 mGy, 1. 54 mGy and 1. 03 mGy, with the group using tube current 20 mAs being the lowest, 15. 5% of the group using conventional dose. Conclusion The image quality of low-dose chest CT scans using a tube voltage value fixed at 120kV and a tube current 20mAs in AIDS patients combined with low PCP body mass can meet the requirements of clinical diagnosis, and at the same time, it can reduce the radiation dose.