化学分析计量
化學分析計量
화학분석계량
CHEMICAL ANALYSIS AND METERAGE
2013年
5期
51-54
,共4页
张进%何鑫%姚思童%刘虹辰
張進%何鑫%姚思童%劉虹辰
장진%하흠%요사동%류홍신
漫反射傅里叶变换红外光谱法%抗坏血酸%维生素C制剂%含量测定
漫反射傅裏葉變換紅外光譜法%抗壞血痠%維生素C製劑%含量測定
만반사부리협변환홍외광보법%항배혈산%유생소C제제%함량측정
diffuse reflectance fourier transform infrared spectroscopy(DRIFTS)%ascorbic acid%vitamin C tablet%determination
采用漫反射傅里叶变换红外光谱法(DRIFTS)测定维生素C制剂中抗坏血酸的含量。选择合适的分析峰1660~1680 cm-1,可排除其它成分的影响,吸光度与抗坏血酸的浓度呈良好线性关系。以镜面为背景,以75μm筛过筛后的KBr为稀释剂,用DRIFTS法分别测定了3种市售维生素C制剂中的抗坏血酸含量,测定结果分别为75.60%,9.80%,6.72%;相对标准偏差(n=6)分别为6.6%,4.9%,8.2%,与碘量法测定结果接近。样品加标回收率为94.0%。DRIFTS法测定维生素C中的抗坏血酸方法简便、准确,无试剂消耗,无废液产生,是非破坏和非污染的绿色分析方法。但其测定结果的精密度不如碘量法,而且其测定结果的准确程度受样品的性状限制,对于白色粉末或可研磨为粉末的样品分析结果准确度较高。
採用漫反射傅裏葉變換紅外光譜法(DRIFTS)測定維生素C製劑中抗壞血痠的含量。選擇閤適的分析峰1660~1680 cm-1,可排除其它成分的影響,吸光度與抗壞血痠的濃度呈良好線性關繫。以鏡麵為揹景,以75μm篩過篩後的KBr為稀釋劑,用DRIFTS法分彆測定瞭3種市售維生素C製劑中的抗壞血痠含量,測定結果分彆為75.60%,9.80%,6.72%;相對標準偏差(n=6)分彆為6.6%,4.9%,8.2%,與碘量法測定結果接近。樣品加標迴收率為94.0%。DRIFTS法測定維生素C中的抗壞血痠方法簡便、準確,無試劑消耗,無廢液產生,是非破壞和非汙染的綠色分析方法。但其測定結果的精密度不如碘量法,而且其測定結果的準確程度受樣品的性狀限製,對于白色粉末或可研磨為粉末的樣品分析結果準確度較高。
채용만반사부리협변환홍외광보법(DRIFTS)측정유생소C제제중항배혈산적함량。선택합괄적분석봉1660~1680 cm-1,가배제기타성분적영향,흡광도여항배혈산적농도정량호선성관계。이경면위배경,이75μm사과사후적KBr위희석제,용DRIFTS법분별측정료3충시수유생소C제제중적항배혈산함량,측정결과분별위75.60%,9.80%,6.72%;상대표준편차(n=6)분별위6.6%,4.9%,8.2%,여전량법측정결과접근。양품가표회수솔위94.0%。DRIFTS법측정유생소C중적항배혈산방법간편、준학,무시제소모,무폐액산생,시비파배화비오염적록색분석방법。단기측정결과적정밀도불여전량법,이차기측정결과적준학정도수양품적성상한제,대우백색분말혹가연마위분말적양품분석결과준학도교고。
Ascorbic acid in vitamin C tablet was determined by DRIFTS. The characteristic absorption peak range 1 660-1 680 cm-1 was selected as the analytical peak to eliminate the interference of subsidiary. Linear relationship between the concentration of ascorbic acid and the absorbance kept very well. The ascorbic acid contents in three vitamin C tablets in market were determined by DRIFTS. The mirror was used as the background,and smaples were diluented by KBr powder sieved with 75μm sieve. The results of the three samples by the DRIFTS were 75.60%,9.80%,6.72%;RSD(n=6) were 6.6%,4.9%,8.2%,respectively,which were consistent with the Iodometric method. The average recovery of ascorbic acid was 94.0%. The DRIFTS method is rapid,accurate and without consumption of reagents and solutions. It is a non-destructive and non-polluting green ananlysis method. But the precision of DRIFTS method is poorer than iodine quantity method, and the accuracy of the determination results is restricted by the sample’s state. The determination results for the white powder have good accuracy.