Lenstar 900%眼生物学测量%白内障
Lenstar 900%眼生物學測量%白內障
Lenstar 900%안생물학측량%백내장
Lenstar 900%eye biometry%cataract
目的:比较Lenstar 900与A型超声和角膜曲率计在白内障患者眼生物学测量中测量眼轴长度( AL )、前房深度( ACD)及角膜曲率( K1,K2,Km)上的差异,并分析其一致性,为Lenstar 900临床应用提供参考。<br> 方法:选取接受白内障手术的患者36例50眼,晶状体核均在Ⅳ级以下,术前分别采用Lenstar 900、A超和角膜曲率计完成AL,ACD和K1,K2,Km的测量,比较测量结果的差异采用配对t检验,测量结果的相关性采用 Pearson相关分析,测量结果的一致性采用Bland-Ahamn统计分析。<br> 结果:Lenstar 900与A型超声测量AL和ACD值之间的差异无统计学意义( P>0.05)。 Lenstar 900与角膜曲率计测量Kl,K2和 Km 值的差异均无统计学意义(P>0.05)。Lenstar 900与A型超声和角膜曲率计测量的所有参数均具有密切的线性相关(r>0.9,P<0.05),Bland-Ahamn统计分析一致性良好。<br> 结论:Lenstar 900与A型超声和角膜曲率计对白内障患者术前生物学测量结果准备可靠,可以相互替代,但是Lenstar 900不仅可以一次完成对AL、ACD和角膜曲率的测量,而且同时可以获得角膜厚度、晶状体厚度、角膜缘白到白距离、瞳孔大小、光轴的偏心率及视网膜厚度等参数,具有测量过程非接触、方便快捷、节约时间等优点,具有广泛的应用前景。
目的:比較Lenstar 900與A型超聲和角膜麯率計在白內障患者眼生物學測量中測量眼軸長度( AL )、前房深度( ACD)及角膜麯率( K1,K2,Km)上的差異,併分析其一緻性,為Lenstar 900臨床應用提供參攷。<br> 方法:選取接受白內障手術的患者36例50眼,晶狀體覈均在Ⅳ級以下,術前分彆採用Lenstar 900、A超和角膜麯率計完成AL,ACD和K1,K2,Km的測量,比較測量結果的差異採用配對t檢驗,測量結果的相關性採用 Pearson相關分析,測量結果的一緻性採用Bland-Ahamn統計分析。<br> 結果:Lenstar 900與A型超聲測量AL和ACD值之間的差異無統計學意義( P>0.05)。 Lenstar 900與角膜麯率計測量Kl,K2和 Km 值的差異均無統計學意義(P>0.05)。Lenstar 900與A型超聲和角膜麯率計測量的所有參數均具有密切的線性相關(r>0.9,P<0.05),Bland-Ahamn統計分析一緻性良好。<br> 結論:Lenstar 900與A型超聲和角膜麯率計對白內障患者術前生物學測量結果準備可靠,可以相互替代,但是Lenstar 900不僅可以一次完成對AL、ACD和角膜麯率的測量,而且同時可以穫得角膜厚度、晶狀體厚度、角膜緣白到白距離、瞳孔大小、光軸的偏心率及視網膜厚度等參數,具有測量過程非接觸、方便快捷、節約時間等優點,具有廣汎的應用前景。
목적:비교Lenstar 900여A형초성화각막곡솔계재백내장환자안생물학측량중측량안축장도( AL )、전방심도( ACD)급각막곡솔( K1,K2,Km)상적차이,병분석기일치성,위Lenstar 900림상응용제공삼고。<br> 방법:선취접수백내장수술적환자36례50안,정상체핵균재Ⅳ급이하,술전분별채용Lenstar 900、A초화각막곡솔계완성AL,ACD화K1,K2,Km적측량,비교측량결과적차이채용배대t검험,측량결과적상관성채용 Pearson상관분석,측량결과적일치성채용Bland-Ahamn통계분석。<br> 결과:Lenstar 900여A형초성측량AL화ACD치지간적차이무통계학의의( P>0.05)。 Lenstar 900여각막곡솔계측량Kl,K2화 Km 치적차이균무통계학의의(P>0.05)。Lenstar 900여A형초성화각막곡솔계측량적소유삼수균구유밀절적선성상관(r>0.9,P<0.05),Bland-Ahamn통계분석일치성량호。<br> 결론:Lenstar 900여A형초성화각막곡솔계대백내장환자술전생물학측량결과준비가고,가이상호체대,단시Lenstar 900불부가이일차완성대AL、ACD화각막곡솔적측량,이차동시가이획득각막후도、정상체후도、각막연백도백거리、동공대소、광축적편심솔급시망막후도등삼수,구유측량과정비접촉、방편쾌첩、절약시간등우점,구유엄범적응용전경。
AIM:To investigate the differences among Lenstar 900, A-scan ultrasound and keratometer in measurement of axial length ( AL ) , anterior chamber depth ( ACD ) and corneal curvature ( K1 , K2 , Km ) , and evaluate the consistency of the instruments, with the purpose providing references for the clinical application of Lenstar 900. <br> METHODS: In this study we picked up 36 patients ( 50 eyes ) underwent cataract surgery, and lens nucleus hardness were under level IV. Before the operation, AL, ACD and K1 , K2 , Km were measured by Lenstar 900, A-scan ultrasound and keratometer respectively. The differences between the results were compared by the paired t-test. The correlation of the results was analyzed by Pearson correlation analysis, and the consistency was measured by Bland-Ahamn method. <br> RESULTS: The mean AL and ACD values measured by Lenstar 900 and A-scan ultrasound had no significantly statistic differences (P>0. 05). The K1, K2, Km measured by Lenstar 900 and keratometer were not significantly statistical different (P>0. 05). The results measured by these three instruments had close linearity correlation ( r>0.9, P<0. 01). The consistency of the results was well in Bland-Ahamn analysis. <br> CONCLUSION:The preoperatively biometric result of Lenstar 900, A - scan ultrasound and keratometer in patients with cataract are all reliable, and they can be substituted by each other. However, Lenstar 900 can not only measure AL, ACD and corneal curvature at the same time, but also cornal thickness, lens thickness, white to white, pupil size, optical axis eccentricity, retinal thickness and so on. It has a number of advantages such as non-touching, convenient and efficient, and can be recommended to use widely.