中国基层医药
中國基層醫藥
중국기층의약
CHINESE JOURNAL OF PRIMARY MEDICINE AND PHARMACY
2014年
16期
2496-2497,2498
,共3页
数字化X线摄影%床边摄影%影像质量控制
數字化X線攝影%床邊攝影%影像質量控製
수자화X선섭영%상변섭영%영상질량공제
Digital radiography%bedside radiography%Image quality conirol
目的:通过对床边DR的影像质量的影响因素分析,总结DR床旁摄影的经验与教训,提出改进方法。方法对1265份数字化X线影像资料(1265张影像片)进行回顾性分析,并按部位、阳性率及图像质量进行统计和分析,探讨影响成像质量因素。结果1265份影像片中,甲级片1032张,甲级片率为81.6%;乙级片181张,乙级片率为14.3%;丙级片52张,丙级片率为4.1%,没有废片。对233张非甲级片进行影像质量原因统计、分析及分类。其中由设备性能问题所致为39张,占16.7%;图像噪声:15张,占6.4%;摄片技术:101张,占43.3%;图像后处理:31张,13.3%;激光相机与打印:10张,占4.3%;人为因素的影响37张,占15.9%。结论影响DR照片质量因素很多,技术人员的操作技能和水平是其主要因素。
目的:通過對床邊DR的影像質量的影響因素分析,總結DR床徬攝影的經驗與教訓,提齣改進方法。方法對1265份數字化X線影像資料(1265張影像片)進行迴顧性分析,併按部位、暘性率及圖像質量進行統計和分析,探討影響成像質量因素。結果1265份影像片中,甲級片1032張,甲級片率為81.6%;乙級片181張,乙級片率為14.3%;丙級片52張,丙級片率為4.1%,沒有廢片。對233張非甲級片進行影像質量原因統計、分析及分類。其中由設備性能問題所緻為39張,佔16.7%;圖像譟聲:15張,佔6.4%;攝片技術:101張,佔43.3%;圖像後處理:31張,13.3%;激光相機與打印:10張,佔4.3%;人為因素的影響37張,佔15.9%。結論影響DR照片質量因素很多,技術人員的操作技能和水平是其主要因素。
목적:통과대상변DR적영상질량적영향인소분석,총결DR상방섭영적경험여교훈,제출개진방법。방법대1265빈수자화X선영상자료(1265장영상편)진행회고성분석,병안부위、양성솔급도상질량진행통계화분석,탐토영향성상질량인소。결과1265빈영상편중,갑급편1032장,갑급편솔위81.6%;을급편181장,을급편솔위14.3%;병급편52장,병급편솔위4.1%,몰유폐편。대233장비갑급편진행영상질량원인통계、분석급분류。기중유설비성능문제소치위39장,점16.7%;도상조성:15장,점6.4%;섭편기술:101장,점43.3%;도상후처리:31장,13.3%;격광상궤여타인:10장,점4.3%;인위인소적영향37장,점15.9%。결론영향DR조편질량인소흔다,기술인원적조작기능화수평시기주요인소。
Objective To analyze the influence factors of quality of bedside digital radiography DR image , and summrize the experience and lessons of DR bedside photography ,and put forward some suggestions and improve-ment methods ,and continuously improve the clinical value of DR .Methods The data of 1 265 cases of digital radio-graphy DR imaging(1 265 imaging films) were retrospectively analyzed ,and were statistically analyzed according to the site,the positive rate and image quality .Influence factors of the imaging quality was explored .Results In 1 265 films,the A-level images were 1 032 cases,the rate of A-level was 81.65%,and the B-level images were 181 cases, the rate of B-level was 14.30%,and the C-level images were 52 cases,the rate of C-level was 4.10%,and there was no rejected film.The image quality of 233 non-A-level images were statistically analyzed and classified .39 cases caused by the equipment performance problems ,accounting for 16.70%;15 cases caused by the image noise ,account-ing for 6.40%;101 cases caused by radiography technology ,accounting for 43.30%;31 cases caused by image pro-cessing,13.30%;10 cases caused by laser printer and print ,accounting for 4.30%;37 cases caused by the influence of human factors,accounted for 15.90%.Conclusion It is demonstrated that there are many factors influence DR image quality and the main factor is operation skill of technician .