目的:评价Sysmex CA-1500全自动血凝仪PT-der法和Von-Clauss法测定血浆纤维蛋白原(Fib)的相关性和准确性。方法采用Sysmex CA-1500全自动血凝仪PT-der法和Von-Clauss法分别测定755例血液标本的血浆Fib浓度;高Fib浓度标本的稀释比例为1∶8、2∶7、3∶6、4∶5、5∶4、6∶3,以稀释比例为横坐标,以2种方法测得的Fib浓度为纵坐标,进行简单线性回归分析。结果 Fib:2.0~<6.0 g/L时,Von-Clauss法检测Fib值高于PT-der法(P<0.01);Fib:<2.0 g/L ,>6.0 g/L时, PT-der法检测Fib值高于Von-Clauss法(P<0.01)。PT-der法和Von-Clauss法检测的线性回归方程分别为:Y =4.5377 X+1.5513(R2=0.8973),Y=7.7922X+0.2900(R2=0.9805)。结论 Von-Clauss法能较好地反映人体内具有凝血功能的Fib水平。
目的:評價Sysmex CA-1500全自動血凝儀PT-der法和Von-Clauss法測定血漿纖維蛋白原(Fib)的相關性和準確性。方法採用Sysmex CA-1500全自動血凝儀PT-der法和Von-Clauss法分彆測定755例血液標本的血漿Fib濃度;高Fib濃度標本的稀釋比例為1∶8、2∶7、3∶6、4∶5、5∶4、6∶3,以稀釋比例為橫坐標,以2種方法測得的Fib濃度為縱坐標,進行簡單線性迴歸分析。結果 Fib:2.0~<6.0 g/L時,Von-Clauss法檢測Fib值高于PT-der法(P<0.01);Fib:<2.0 g/L ,>6.0 g/L時, PT-der法檢測Fib值高于Von-Clauss法(P<0.01)。PT-der法和Von-Clauss法檢測的線性迴歸方程分彆為:Y =4.5377 X+1.5513(R2=0.8973),Y=7.7922X+0.2900(R2=0.9805)。結論 Von-Clauss法能較好地反映人體內具有凝血功能的Fib水平。
목적:평개Sysmex CA-1500전자동혈응의PT-der법화Von-Clauss법측정혈장섬유단백원(Fib)적상관성화준학성。방법채용Sysmex CA-1500전자동혈응의PT-der법화Von-Clauss법분별측정755례혈액표본적혈장Fib농도;고Fib농도표본적희석비례위1∶8、2∶7、3∶6、4∶5、5∶4、6∶3,이희석비례위횡좌표,이2충방법측득적Fib농도위종좌표,진행간단선성회귀분석。결과 Fib:2.0~<6.0 g/L시,Von-Clauss법검측Fib치고우PT-der법(P<0.01);Fib:<2.0 g/L ,>6.0 g/L시, PT-der법검측Fib치고우Von-Clauss법(P<0.01)。PT-der법화Von-Clauss법검측적선성회귀방정분별위:Y =4.5377 X+1.5513(R2=0.8973),Y=7.7922X+0.2900(R2=0.9805)。결론 Von-Clauss법능교호지반영인체내구유응혈공능적Fib수평。
Objective To evaluate the relevance and accuracy of fibrinogen (Fib) detection by using PT-der assay and Von-Clauss assay on Sysmex CA-1500 Automated Coagulation Analyzer .Methods PT-der assay and Von-Clauss assay on Sysmex CA-1500 Automated Coagulation Analyzer were employed to detect the plasma Fib concentrations of 755 blood samples .The dilution ratios of samples with high Fib concentration were 1∶8 ,2∶7 ,3∶6 ,4∶5 ,5∶4 ,6∶3 ,respectively .The dilution ratios were served as the abscissa ,and the Fib concentrations measured by two methods as the ordinate ,a simple linear regression analysis was per-formed .Results When Fib concentration was in 2 .0- <6 .0 g/L ,the Fib value obtained by Von-Clauss assay was higher than that by PT-der assay(P<0 .01) .When Fib concentration was below 2 .0 g/L or above 6 .0 g/L ,the Fib value obtained by PT-der assay was higher than that by Von-Clauss assay(P<0 .01) .The linear regression equations of PT-der assay and Von-Clauss assay were Y=4 .537 7X+1 .551 3(R2 =0 .897 3) ,Y = 7 .792 2X+ 0 .290 0(R2 =0 .980 5) ,respectively .Conclusion Von-Clauss assay can better reflect the Fib level of human body which has a blood clotting function .