目的 参照正常成年人颈椎的解剖学数据,设计出一种前路齿状突中空螺钉导向仪.方法 对上颈椎64排螺旋CT、颈椎正侧位X片检查显示正常的60例成年人影像资料进行数据分析,分为男女两组,对两组进行t或t'检验.结果 获得与导向仪相关的解剖学数据;64排CT与颈椎侧位X片测量C2/3椎间盘前缘高度分别为男性组(3.26±1.19) mm、女性组(2.54±0.83) mm与男性组(3.21±0.78) mm、女性组(3.21±0.90) mm,按性别分别行t检验,两者差异均无统计学意义(P>0.05);颈椎侧位X线片测量男性组与女性组的C5椎体前下缘中点对应皮肤至C2椎体前下缘中点的长度分别是(89.09±6.85)mm和(75.93 ±5.67) mm,两者差异有统计学意义(P<0.05).结论 根据测量数据结果设计出用于前路齿状突中空螺钉内固定术的导向仪,此导向仪的设计与人的解剖生理结构符合,易于临床实际操作应用.
目的 參照正常成年人頸椎的解剖學數據,設計齣一種前路齒狀突中空螺釘導嚮儀.方法 對上頸椎64排螺鏇CT、頸椎正側位X片檢查顯示正常的60例成年人影像資料進行數據分析,分為男女兩組,對兩組進行t或t'檢驗.結果 穫得與導嚮儀相關的解剖學數據;64排CT與頸椎側位X片測量C2/3椎間盤前緣高度分彆為男性組(3.26±1.19) mm、女性組(2.54±0.83) mm與男性組(3.21±0.78) mm、女性組(3.21±0.90) mm,按性彆分彆行t檢驗,兩者差異均無統計學意義(P>0.05);頸椎側位X線片測量男性組與女性組的C5椎體前下緣中點對應皮膚至C2椎體前下緣中點的長度分彆是(89.09±6.85)mm和(75.93 ±5.67) mm,兩者差異有統計學意義(P<0.05).結論 根據測量數據結果設計齣用于前路齒狀突中空螺釘內固定術的導嚮儀,此導嚮儀的設計與人的解剖生理結構符閤,易于臨床實際操作應用.
목적 삼조정상성년인경추적해부학수거,설계출일충전로치상돌중공라정도향의.방법 대상경추64배라선CT、경추정측위X편검사현시정상적60례성년인영상자료진행수거분석,분위남녀량조,대량조진행t혹t'검험.결과 획득여도향의상관적해부학수거;64배CT여경추측위X편측량C2/3추간반전연고도분별위남성조(3.26±1.19) mm、녀성조(2.54±0.83) mm여남성조(3.21±0.78) mm、녀성조(3.21±0.90) mm,안성별분별행t검험,량자차이균무통계학의의(P>0.05);경추측위X선편측량남성조여녀성조적C5추체전하연중점대응피부지C2추체전하연중점적장도분별시(89.09±6.85)mm화(75.93 ±5.67) mm,량자차이유통계학의의(P<0.05).결론 근거측량수거결과설계출용우전로치상돌중공라정내고정술적도향의,차도향의적설계여인적해부생리결구부합,역우림상실제조작응용.
Objective To design and product a collimator for anterior cannulated odontoid screw by measuring the data of normal adult cervical spine from the images.Methods A total of 60 normal adults who received 64 row slice CT of the upper cervical spine and the X-Ray plain film of the cervical spine were chosen and divided into male and female groups.Statistical analysis was performed.Results The main anatomy parameters for the design of collimator were obtained.There was no significant difference in the disc height of the C2 and C3 anterior edge between 64 row slice CT [male:(3.26 ± 1.19) mm,and female:(2.54 ±0.83) mm] and X-ray [male:(3.2l ±0.78) mm,and female:(3.21 ±0.90) mm] in the lat-eral cervical spine (P > 0.05).There was statistically significant difference in the length of C5 anterior lower edge midpoint corresponding skin straight pointing to the C2 front bottom edge midpoint between two groups [male:(89.09 ± 6.85) mm,and female:(75.93 ± 5.67) mm].Conclusion The collimator is designed for anterior canuulated odontoid screw.This collimator matching with human cervical anatomy is easier to be used in the clinical application.