西北药学杂志
西北藥學雜誌
서북약학잡지
2014年
4期
368-370
,共3页
硝酸咪康唑%残留溶剂%气相色谱法
硝痠咪康唑%殘留溶劑%氣相色譜法
초산미강서%잔류용제%기상색보법
miconazole nitrate%residual solvents%GC
目的:建立硝酸咪康唑原料残留溶剂的测定方法。方法采用气相色谱法,使用 PEG-20M 毛细管(30 m ×0.25 mm ,0.25μm )色谱柱,柱温采用程序升温,初始温度60℃,保持3 min ,再以30℃· min-1升至185℃,保持20 min;进样口温度为200℃;检测器为FID ;检测器温度为250℃。结果同时测定了硝酸咪康唑原料中的2种有机溶剂,丙酮和甲苯线性范围分别为0.25~1.0mg·mL -1(r=0.9998)、0.0445~0.178mg·mL -1(r=0.9996),平均加样回收率分别为100.3%和99.9%, RSD分别为1.8%和1.3%。结论该方法灵敏、准确,可用于硝酸咪康唑中残留溶剂的检测。
目的:建立硝痠咪康唑原料殘留溶劑的測定方法。方法採用氣相色譜法,使用 PEG-20M 毛細管(30 m ×0.25 mm ,0.25μm )色譜柱,柱溫採用程序升溫,初始溫度60℃,保持3 min ,再以30℃· min-1升至185℃,保持20 min;進樣口溫度為200℃;檢測器為FID ;檢測器溫度為250℃。結果同時測定瞭硝痠咪康唑原料中的2種有機溶劑,丙酮和甲苯線性範圍分彆為0.25~1.0mg·mL -1(r=0.9998)、0.0445~0.178mg·mL -1(r=0.9996),平均加樣迴收率分彆為100.3%和99.9%, RSD分彆為1.8%和1.3%。結論該方法靈敏、準確,可用于硝痠咪康唑中殘留溶劑的檢測。
목적:건립초산미강서원료잔류용제적측정방법。방법채용기상색보법,사용 PEG-20M 모세관(30 m ×0.25 mm ,0.25μm )색보주,주온채용정서승온,초시온도60℃,보지3 min ,재이30℃· min-1승지185℃,보지20 min;진양구온도위200℃;검측기위FID ;검측기온도위250℃。결과동시측정료초산미강서원료중적2충유궤용제,병동화갑분선성범위분별위0.25~1.0mg·mL -1(r=0.9998)、0.0445~0.178mg·mL -1(r=0.9996),평균가양회수솔분별위100.3%화99.9%, RSD분별위1.8%화1.3%。결론해방법령민、준학,가용우초산미강서중잔류용제적검측。
Objective To establish a method for the determination of two residual solvents in miconazole nitrate .Methods The de-termination was performed on a GB-WAX column (30 m × 0 .25 mm ,0 .25 μm) .The column temperature was maintained at 60 ℃for 3 min ,and programmed to 185 ℃ at 30 ℃ · min-1 ,then held for 20 min .FID was used as a detector ,and the temperatures of the injector and detector were set at 200 and 250 ℃ ,respectively .Results Two organic solvents of miconazole nitrate were deter-mined simultaneously in raw materials .The calibration curves were linear in the ranges of 0 .25-1 .0 mg · mL -1 for acetone (r=0 .999 8) ,and 0 .0445-0 .178 mg · mL -1 for toluene (r=0 .999 6) .The average recoveries were 100 .3% and 99 .9% ,RSDs were 1 .8% and 1 .3% ,respectively .Conclusion The methods were proved to be accurate and sensitive ,and can be used for the detec-tion of residual solvents in miconazole nitrate .