目的:比较光学相干断层扫描( optical coherence tomography ,OCT)、角膜内皮计与超声测厚仪测量准分子激光上皮下角膜磨镶术( laser subepithelial keratomileusis , LASEK)术后患者中央角膜厚度( CCT )的差异性和相关性。<br> 方法:选取LASEK术后6 mo复诊的近视患者49例96眼,对每位患者分别用OCT、角膜内皮计、超声测厚仪三种仪器测量其CCT,数据统计分析采用配对t 检验和Pearson相关性分析。<br> 结果:OCT、角膜内皮计、超声测厚仪测量CCT的平均值为分别为475.15±33.67,447.80±34.11,465.18±34.23μm。手术预留CCT平均值为431.22±35.32μm。三种仪器所测量结果之间两两比较差异均有统计学意义(P<0.05)。线性相关显示超声测厚仪与OCT、角膜内皮计两种仪器之间存在正相关( r=0.949,0.983, P<0.01)。 LASEK术后实际CCT较手术预留角膜厚度要厚。<br> 结论:三种仪器的中央角膜厚度测量值不可互换。
目的:比較光學相榦斷層掃描( optical coherence tomography ,OCT)、角膜內皮計與超聲測厚儀測量準分子激光上皮下角膜磨鑲術( laser subepithelial keratomileusis , LASEK)術後患者中央角膜厚度( CCT )的差異性和相關性。<br> 方法:選取LASEK術後6 mo複診的近視患者49例96眼,對每位患者分彆用OCT、角膜內皮計、超聲測厚儀三種儀器測量其CCT,數據統計分析採用配對t 檢驗和Pearson相關性分析。<br> 結果:OCT、角膜內皮計、超聲測厚儀測量CCT的平均值為分彆為475.15±33.67,447.80±34.11,465.18±34.23μm。手術預留CCT平均值為431.22±35.32μm。三種儀器所測量結果之間兩兩比較差異均有統計學意義(P<0.05)。線性相關顯示超聲測厚儀與OCT、角膜內皮計兩種儀器之間存在正相關( r=0.949,0.983, P<0.01)。 LASEK術後實際CCT較手術預留角膜厚度要厚。<br> 結論:三種儀器的中央角膜厚度測量值不可互換。
목적:비교광학상간단층소묘( optical coherence tomography ,OCT)、각막내피계여초성측후의측량준분자격광상피하각막마양술( laser subepithelial keratomileusis , LASEK)술후환자중앙각막후도( CCT )적차이성화상관성。<br> 방법:선취LASEK술후6 mo복진적근시환자49례96안,대매위환자분별용OCT、각막내피계、초성측후의삼충의기측량기CCT,수거통계분석채용배대t 검험화Pearson상관성분석。<br> 결과:OCT、각막내피계、초성측후의측량CCT적평균치위분별위475.15±33.67,447.80±34.11,465.18±34.23μm。수술예류CCT평균치위431.22±35.32μm。삼충의기소측량결과지간량량비교차이균유통계학의의(P<0.05)。선성상관현시초성측후의여OCT、각막내피계량충의기지간존재정상관( r=0.949,0.983, P<0.01)。 LASEK술후실제CCT교수술예류각막후도요후。<br> 결론:삼충의기적중앙각막후도측량치불가호환。
AIM: To compare the differences and the relationships of central corneal thickness ( CCT) measurements among optical coherence tomography ( OCT ) , Tomey pachymetry and ultrasound pachymetry after laser assisted subepithelial keratomileusis ( LASEK) . <br> METHODS:Forty-nine patients ( 96 eyes ) with myopia after 6mo of LASEK were assigned to measure the CCT by OCT, Tomey pachymetry and ultrasound pachymetry. Paired t-test and Pearson correlation analysis was used for statistical analysis. <br> RESULTS: The mean CCT measured by OCT, Tomey pachymetry and ultrasound pachymetry was ( 475.15 ± 33.67μm), (447.80 ±34.11μm), (465.18 ±34.23μm), respectively.Surgery reserved mean CCT was ( 431.22 ± 35.32 )μm.There were statistically differences between two comparative measured results (P<0.05).Ultrasound pachymetry had a positive correlation with OCT and Tomey pachymetry ( P <0.01 ), and the correlation coefficient was 0.949 and 0.983, respectively.The actual CCT of 6mo post-operation was thicker than surgery reserved CCT. <br> CONCLUSION: The CCT measurements of three instruments were not interchangeable.