孟庆乐%杨瑞%曹艳%王自正%李少华%邵国强
孟慶樂%楊瑞%曹豔%王自正%李少華%邵國彊
맹경악%양서%조염%왕자정%리소화%소국강
SPECT/CT融合显像%99Tcm-MIBI双时相显像%甲状腺结节%体层摄影术%放射性核素
SPECT/CT融閤顯像%99Tcm-MIBI雙時相顯像%甲狀腺結節%體層攝影術%放射性覈素
SPECT/CT융합현상%99Tcm-MIBI쌍시상현상%갑상선결절%체층섭영술%방사성핵소
SPECT-CT fusion imaging%dual-phase 99Tcm-MIBI thyroid scintigraphy%thyroid nodules%tomography%radionuclide
目的:探讨SPECT/CT融合显像在99Tcm-甲氧基异丁基异腈(MIBI)甲状腺亲肿瘤显像中的临床价值。方法103例甲状腺冷结节患者先行99Tcm-MIBI双时相(15 min与2 h)显像,后行SPECT/CT同机断层融合,结果与病理检查结果进行比较。结果99Tcm-MIBI双时相法对甲状腺结节性质良恶性鉴别的灵敏度、特异度和准确性分别为59.1%、49.4%、51.4%,SPECT/CT融合显像诊断的灵敏度、特异度和准确性分别为81.8%、76.5%、77.7%。结论 SPECT/CT融合显像在甲状腺99Tcm-MIBI亲肿瘤显像中可提高诊断准确性,为临床提供可靠信息。
目的:探討SPECT/CT融閤顯像在99Tcm-甲氧基異丁基異腈(MIBI)甲狀腺親腫瘤顯像中的臨床價值。方法103例甲狀腺冷結節患者先行99Tcm-MIBI雙時相(15 min與2 h)顯像,後行SPECT/CT同機斷層融閤,結果與病理檢查結果進行比較。結果99Tcm-MIBI雙時相法對甲狀腺結節性質良噁性鑒彆的靈敏度、特異度和準確性分彆為59.1%、49.4%、51.4%,SPECT/CT融閤顯像診斷的靈敏度、特異度和準確性分彆為81.8%、76.5%、77.7%。結論 SPECT/CT融閤顯像在甲狀腺99Tcm-MIBI親腫瘤顯像中可提高診斷準確性,為臨床提供可靠信息。
목적:탐토SPECT/CT융합현상재99Tcm-갑양기이정기이정(MIBI)갑상선친종류현상중적림상개치。방법103례갑상선랭결절환자선행99Tcm-MIBI쌍시상(15 min여2 h)현상,후행SPECT/CT동궤단층융합,결과여병리검사결과진행비교。결과99Tcm-MIBI쌍시상법대갑상선결절성질량악성감별적령민도、특이도화준학성분별위59.1%、49.4%、51.4%,SPECT/CT융합현상진단적령민도、특이도화준학성분별위81.8%、76.5%、77.7%。결론 SPECT/CT융합현상재갑상선99Tcm-MIBI친종류현상중가제고진단준학성,위림상제공가고신식。
Objective To explore the clinical value of SPECT-CT fusion imaging in 99Tcm-methoxy isobutyl isonitrile (MIBI) thyroid scintigraphy. Methods 103 patients with nonfunctional thyroid nodules were examined by using dual-phase 99Tcm-MIBI thyroid scintigraphy (15min and 2h) and then examined by using SPECT-CT fusion imaging. The results of 103 patients achieved through dual-phase 99Tcm-MIBI thyroid scintigraphy and SPECT-CT fusion imaging were compared with those of pathological examination. Results The sensitivity, speciifcity and accuracy of dual-phase 99Tcm-MIBI thyroid scintigraphy in diagnosing benign and malignant of nonfunctional thyroid nodules were 59.1%, 49.4%and 51.4%, respectively while those of SPECT-CT fusion imaging were 81.8%, 76.5%and 77.7%, respectively. Conclusion The diagnosis accuracy of 99Tcm-MIBI thyroid scintigraphy in nonfunctional thyroid nodules can be improved with the application of SPECT-CT fusion imaging, who can provide reliable information for clinical diagnosis.