北京地区%放射诊疗%辐射危害控制%质量控制
北京地區%放射診療%輻射危害控製%質量控製
북경지구%방사진료%복사위해공제%질량공제
Beijing Municipal%Radio diagnosis and radiotherapy%Control of radiation hazard%Quality control
目的:掌握北京地区放射诊疗设备质量控制的基本状况,更好地推进和引导放射诊疗防护工作的开展。方法:选择38家医疗机构中运行的110台放射诊疗设备为研究对象,确定其关键性能指标,使用相应的仪器设备,依据国家有关技术规范与防护标准进行检测和评价。结果:110台放射诊疗设备初检时的总合格率为79.1%,其中放射诊断设备的合格率为82.5%,放射治疗设备的合格率为70.8%,核医学设备(PET和SPECT)的合格率为66.7%;经调试并复检正常后的设备总合格率为95.5%。所调查的医疗机构及其设备客观反映了本地区放射诊疗质量控制的实际状况。结论:北京地区放射诊疗设备初检时的性能状态不容乐观,加强设备的质量控制检测及辐射危害控制非常必要。
目的:掌握北京地區放射診療設備質量控製的基本狀況,更好地推進和引導放射診療防護工作的開展。方法:選擇38傢醫療機構中運行的110檯放射診療設備為研究對象,確定其關鍵性能指標,使用相應的儀器設備,依據國傢有關技術規範與防護標準進行檢測和評價。結果:110檯放射診療設備初檢時的總閤格率為79.1%,其中放射診斷設備的閤格率為82.5%,放射治療設備的閤格率為70.8%,覈醫學設備(PET和SPECT)的閤格率為66.7%;經調試併複檢正常後的設備總閤格率為95.5%。所調查的醫療機構及其設備客觀反映瞭本地區放射診療質量控製的實際狀況。結論:北京地區放射診療設備初檢時的性能狀態不容樂觀,加彊設備的質量控製檢測及輻射危害控製非常必要。
목적:장악북경지구방사진료설비질량공제적기본상황,경호지추진화인도방사진료방호공작적개전。방법:선택38가의료궤구중운행적110태방사진료설비위연구대상,학정기관건성능지표,사용상응적의기설비,의거국가유관기술규범여방호표준진행검측화평개。결과:110태방사진료설비초검시적총합격솔위79.1%,기중방사진단설비적합격솔위82.5%,방사치료설비적합격솔위70.8%,핵의학설비(PET화SPECT)적합격솔위66.7%;경조시병복검정상후적설비총합격솔위95.5%。소조사적의료궤구급기설비객관반영료본지구방사진료질량공제적실제상황。결론:북경지구방사진료설비초검시적성능상태불용악관,가강설비적질량공제검측급복사위해공제비상필요。
Objective: The basic status of quality control of apparatus of radio diagnosis and radiotherapy in Beijing were mastered to promote and guide the radiation protection work efficiently. Methods: 110 devices of radio diagnosis and radiotherapy from 38 typical medical institutions were selected as the research object, and their key performances were determined and tested by using the appropriate surveying instruments, then they were evaluated according to the national related technical criterions and protection standards. Results:The initial total qualified rate of the 110 sets of radio diagnosis and radiotherapy apparatus was 79.1%, and the initial qualified rates for radio diagnosis equipment, radiotherapy equipment and nuclear medicine equipment (PET and SPECT) were 82.5%, 70.8%and 66.7%respectively. The total qualified rate of all the devices was 95.5%after being debugged and rechecked. The local practical quality control conditions of radio diagnosis and radiotherapy can be reflected from the 38 medical institutions and their medical radiation equipments surveyed in Beijing. Conclusion:The initial examination of performances of the 110 medical radiation devices in Beijing is not optimistic, it is very necessary to strengthen the quality control testing of radio diagnosis and radiotherapy devices.