国际检验医学杂志
國際檢驗醫學雜誌
국제검험의학잡지
INTERNATIONAL JOURNAL OF LABORATORY MEDICINE
2014年
5期
605-606
,共2页
六西格玛%偏倚%实验室技术和方法
六西格瑪%偏倚%實驗室技術和方法
륙서격마%편의%실험실기술화방법
six sigma%bias%laboratory techniques and procedures
目的:评价Beckman Coulter L H750检测系统的性能。方法根据CLSI EP9确定分析仪的真实度,根据CLSI EP5确定分析仪的精密度,通过六西格玛度量方法计算血细胞分析仪各项目的性能并进行评估。结果 L H750白细胞计数(WBC )的六西格玛水平在5.5~11.1,红细胞计数(RBC)在5.5~9.4,血红蛋白含量(Hb)在6.0~14.6,血小板计数(PLT )在4.1~10.0,红细胞压积(HCT)的六西格玛值低于了3.0。结论 Beckman Coulter LH750检测系统基本满足六西格玛的质量管理要求, HCT这个项目还需要进行质量改进。六西格玛质量理论可以反映检测过程的性能。
目的:評價Beckman Coulter L H750檢測繫統的性能。方法根據CLSI EP9確定分析儀的真實度,根據CLSI EP5確定分析儀的精密度,通過六西格瑪度量方法計算血細胞分析儀各項目的性能併進行評估。結果 L H750白細胞計數(WBC )的六西格瑪水平在5.5~11.1,紅細胞計數(RBC)在5.5~9.4,血紅蛋白含量(Hb)在6.0~14.6,血小闆計數(PLT )在4.1~10.0,紅細胞壓積(HCT)的六西格瑪值低于瞭3.0。結論 Beckman Coulter LH750檢測繫統基本滿足六西格瑪的質量管理要求, HCT這箇項目還需要進行質量改進。六西格瑪質量理論可以反映檢測過程的性能。
목적:평개Beckman Coulter L H750검측계통적성능。방법근거CLSI EP9학정분석의적진실도,근거CLSI EP5학정분석의적정밀도,통과륙서격마도량방법계산혈세포분석의각항목적성능병진행평고。결과 L H750백세포계수(WBC )적륙서격마수평재5.5~11.1,홍세포계수(RBC)재5.5~9.4,혈홍단백함량(Hb)재6.0~14.6,혈소판계수(PLT )재4.1~10.0,홍세포압적(HCT)적륙서격마치저우료3.0。결론 Beckman Coulter LH750검측계통기본만족륙서격마적질량관리요구, HCT저개항목환수요진행질량개진。륙서격마질량이론가이반영검측과정적성능。
Objective To evaluate the performance of Beckman Coulter LH750 detection system .Methods The truth degree of the analyzer was detected according to the CLSI EP9 and the precision was determined according to the CLSI EP5 analyzer ,the per-formance of various items in the blood cell analyzer was calculated and evaluated by the Six Sigma metrics .Results The Six Sigma level of white blood cell (WBC) count in LH750 was 5 .5-11 .1 ,which of the red blood cell (RBC) count in 5 .5-9 .4 ,which of he-moglobin (Hb) in 6 .0-14 .6 and which of the platelet (PLT) count in 4 .1-10 .0 ,the Six Sigma value of hematocrit (HCT) was lower than 3 .0 .Conclusion The Beckman Coulter LH750 detection system basically meets the requirements of the Six Sigma quali-ty management ,the HCT item also needs to be improved in the quality .The Six Sigma quality theory can reflect the performance of the detection process .