DXI800化学发光仪%EP10-A2%性能评价
DXI800化學髮光儀%EP10-A2%性能評價
DXI800화학발광의%EP10-A2%성능평개
DXI800 chemiluminescence analyzer%EP10-A2%performance evaluation
目的:应用CLSI EP10-A2文件初步评价DXI800化学发光仪的分析性能。方法严格按照 EP10-A2文件要求,连续5 d按特定顺序测定高、中、低浓度质控血清中的人生长激素(hG H )浓度,采用 Excel软件计算hG H测定的偏差、总不精密度及每天的多元回归分析参数(即截距、斜率、非线性、携带污染率及漂移度)。结果当质控血清hGH浓度为2.15、5.47、10.80 mmol/L时,hGH测定值与靶值间线性良好(Y =0.992X+0.083,r2=0.978);偏差分别为0.066、0.265、0.055 mmol/L ;总不精密度分别为6.57%、8.53%、4.56%;采用单样本 t检验,每个多元回归分析参数5 d的值与其预测值相比,差异均无统计学意义( P>0.05)。结论 DXI800化学发光仪测定hG H ,其偏差及总不精密度均在允许范围内,线性关系良好,携带污染率较低,稳定性良好,能满足临床及科研需求。
目的:應用CLSI EP10-A2文件初步評價DXI800化學髮光儀的分析性能。方法嚴格按照 EP10-A2文件要求,連續5 d按特定順序測定高、中、低濃度質控血清中的人生長激素(hG H )濃度,採用 Excel軟件計算hG H測定的偏差、總不精密度及每天的多元迴歸分析參數(即截距、斜率、非線性、攜帶汙染率及漂移度)。結果噹質控血清hGH濃度為2.15、5.47、10.80 mmol/L時,hGH測定值與靶值間線性良好(Y =0.992X+0.083,r2=0.978);偏差分彆為0.066、0.265、0.055 mmol/L ;總不精密度分彆為6.57%、8.53%、4.56%;採用單樣本 t檢驗,每箇多元迴歸分析參數5 d的值與其預測值相比,差異均無統計學意義( P>0.05)。結論 DXI800化學髮光儀測定hG H ,其偏差及總不精密度均在允許範圍內,線性關繫良好,攜帶汙染率較低,穩定性良好,能滿足臨床及科研需求。
목적:응용CLSI EP10-A2문건초보평개DXI800화학발광의적분석성능。방법엄격안조 EP10-A2문건요구,련속5 d안특정순서측정고、중、저농도질공혈청중적인생장격소(hG H )농도,채용 Excel연건계산hG H측정적편차、총불정밀도급매천적다원회귀분석삼수(즉절거、사솔、비선성、휴대오염솔급표이도)。결과당질공혈청hGH농도위2.15、5.47、10.80 mmol/L시,hGH측정치여파치간선성량호(Y =0.992X+0.083,r2=0.978);편차분별위0.066、0.265、0.055 mmol/L ;총불정밀도분별위6.57%、8.53%、4.56%;채용단양본 t검험,매개다원회귀분석삼수5 d적치여기예측치상비,차이균무통계학의의( P>0.05)。결론 DXI800화학발광의측정hG H ,기편차급총불정밀도균재윤허범위내,선성관계량호,휴대오염솔교저,은정성량호,능만족림상급과연수구。
Objective To evaluate the detection performance of DXI800 chemiluminescence analyzer (DXI800 analyzer) according to EP10-A2 document of Clinic and Laboraroty Standard Institute (CLSI) .Methods According to EP10-A2 document ,human growth hormone (hGH) in quality control serum were continuously detected for five days with a certain order ,then the deviation ,total imprecision ,intercept ,slope rate ,non-linearity ,cross-contami-nation and drift were analyzed by using Microsoft Excellsoftware .Results The linearity between measured value and target value was relatively well (Y=0 .992X+0 .083 ,r2 =0 .978) when the concentrations of hGH were 2 .15 ,5 .47 and 10 .80 mmol/L ,with bias of 0 .066 ,0 .265 and 0 .055 mmol/L ,and total imprecision of 6 .57% ,8 .53% and 4 .56% ,respectively .Simple sample t test indicated that the difference of intercept ,slope rate ,non-linearity ,cross-contamination and drift was no statistically significant (P>0 .05) .Conclusion The deviation ,total imprecision ,in-tercept ,slope rate ,non-linearity ,cross-contamination and drift of DXI800 analyzer might be within permission lim-it ,and could satisfy the request of clinical application and scientific research of hGH .