重力仪%恒温%畸变%故障
重力儀%恆溫%畸變%故障
중력의%항온%기변%고장
gravitometer%constant temperature%distortion%fault
通过对成都地震台GS-15型重力仪记录固体潮曲线畸变特征分析,找出记录曲线畸变的变化规律,并对引起该变化的因素、物理机制进行分析。仪器加温系统供电电源性能变差,导致仪器内部恒温系统破坏是引起记录固体潮曲线畸变原因。
通過對成都地震檯GS-15型重力儀記錄固體潮麯線畸變特徵分析,找齣記錄麯線畸變的變化規律,併對引起該變化的因素、物理機製進行分析。儀器加溫繫統供電電源性能變差,導緻儀器內部恆溫繫統破壞是引起記錄固體潮麯線畸變原因。
통과대성도지진태GS-15형중력의기록고체조곡선기변특정분석,조출기록곡선기변적변화규률,병대인기해변화적인소、물리궤제진행분석。의기가온계통공전전원성능변차,도치의기내부항온계통파배시인기기록고체조곡선기변원인。
Through the analysis on the aberrance of solid tide that recorded by GS-1 5 gravimeter at Chengdu Seismic Station,we found aberration law,and made a study on influencing factors and physical mechanisms.The reason that leads to the solid tide record curve distorted was that the performance of heating power becomes poor,damage the temperature system inside and outside the instrument.