岩矿测试
巖礦測試
암광측시
ROCK AND MINERAL ANALYSIS
2014年
3期
327-331
,共5页
宋伟娇%代世峰%赵蕾%李霄%王佩佩%李甜%王西勃
宋偉嬌%代世峰%趙蕾%李霄%王珮珮%李甜%王西勃
송위교%대세봉%조뢰%리소%왕패패%리첨%왕서발
煤%硼%微波消解%电感耦合等离子体质谱法
煤%硼%微波消解%電感耦閤等離子體質譜法
매%붕%미파소해%전감우합등리자체질보법
coal sample%boron%microwave digestion%Inductively Coupled Plasma-Mass Spectrometry
煤中硼的准确测定对于研究成煤的沉积环境具有重要意义。微波消解-电感耦合等离子体质谱( ICP-MS)可以有效地测定煤中的大部分微量元素,但由于硼易挥发等特殊的物理化学性质,致使硼的测定过程较为繁琐,测试结果不够准确。本文对微波消解ICP-MS测定煤中硼含量的分析方法进行了改进。样品中加入磷酸、硝酸和氢氟酸,用微波消解仪消解完全,置于电热板加热赶酸完毕后用硝酸提取。硼的卤化物易挥发,磷酸的加入使硼与磷酸结合生成难挥发的稳定络合物,起到了固硼的作用。ICP-MS测定中采用铍作为内标,通过在线加入的方式有效地补偿了基体效应产生的影响;用稀氨水冲洗进样系统,很好地降低了硼的记忆效应;选择高分辨率模式来测定,避免了12C和40Ar4+等质谱峰的干扰。方法的仪器检出限(0.22 ng/mL)和方法检出限(0.34 ng/mL)较低,方法精密度( RSD)小于0.6%,灵敏度高,且测定标准样品的准确度良好。该方法简单快速,适用于批量分析煤样中硼的含量。
煤中硼的準確測定對于研究成煤的沉積環境具有重要意義。微波消解-電感耦閤等離子體質譜( ICP-MS)可以有效地測定煤中的大部分微量元素,但由于硼易揮髮等特殊的物理化學性質,緻使硼的測定過程較為繁瑣,測試結果不夠準確。本文對微波消解ICP-MS測定煤中硼含量的分析方法進行瞭改進。樣品中加入燐痠、硝痠和氫氟痠,用微波消解儀消解完全,置于電熱闆加熱趕痠完畢後用硝痠提取。硼的滷化物易揮髮,燐痠的加入使硼與燐痠結閤生成難揮髮的穩定絡閤物,起到瞭固硼的作用。ICP-MS測定中採用鈹作為內標,通過在線加入的方式有效地補償瞭基體效應產生的影響;用稀氨水遲洗進樣繫統,很好地降低瞭硼的記憶效應;選擇高分辨率模式來測定,避免瞭12C和40Ar4+等質譜峰的榦擾。方法的儀器檢齣限(0.22 ng/mL)和方法檢齣限(0.34 ng/mL)較低,方法精密度( RSD)小于0.6%,靈敏度高,且測定標準樣品的準確度良好。該方法簡單快速,適用于批量分析煤樣中硼的含量。
매중붕적준학측정대우연구성매적침적배경구유중요의의。미파소해-전감우합등리자체질보( ICP-MS)가이유효지측정매중적대부분미량원소,단유우붕역휘발등특수적물이화학성질,치사붕적측정과정교위번쇄,측시결과불구준학。본문대미파소해ICP-MS측정매중붕함량적분석방법진행료개진。양품중가입린산、초산화경불산,용미파소해의소해완전,치우전열판가열간산완필후용초산제취。붕적서화물역휘발,린산적가입사붕여린산결합생성난휘발적은정락합물,기도료고붕적작용。ICP-MS측정중채용피작위내표,통과재선가입적방식유효지보상료기체효응산생적영향;용희안수충세진양계통,흔호지강저료붕적기억효응;선택고분변솔모식래측정,피면료12C화40Ar4+등질보봉적간우。방법적의기검출한(0.22 ng/mL)화방법검출한(0.34 ng/mL)교저,방법정밀도( RSD)소우0.6%,령민도고,차측정표준양품적준학도량호。해방법간단쾌속,괄용우비량분석매양중붕적함량。
Accurate determination of boron in coal is significant to the understanding of the original peat-forming environment. Inductively Coupled Plasma-Mass Spectrometry( ICP-MS)is a reliable technique to determine most trace elements in coal samples. However,the procedure for the determination of boron is complex and the results are inaccurate,due to the special chemical and physical properties of boron,such as high volatility. In this study attempts to determine boron in coal samples using an improved procedure of ICP-MS technique with microwave digestion are made. Samples were digested in a microwave oven with H3 PO4 ,HNO3 and HF,and then were heated to remove acid on the heating plate. Due to the high volatility of halogenides of boron,phosphoric acid was added to chelate the boron which minimizes boron volatilization during heating. Beryllium was added online as the internal standard,which compensated for matrix effects. Diluted ammonium hydroxide solution was used as the wash solution,which significantly reduced the memory effect of boron. Measurement was taken under high resolution mode,in order to minimize spectral interference from 12 C and 40 Ar4+. The instrument detection limit was 0. 22 ng/mL,and the method detection limit was 0. 34 ng/mL with high sensitivity and accuracy,verified by standard materials. The results show that this method is a simple and rapid procedure,and can be used for routine determination of boron for batch coal samples.