陈忠余%周爱娥%赵世巧%刘海波%甘晓玲
陳忠餘%週愛娥%趙世巧%劉海波%甘曉玲
진충여%주애아%조세교%류해파%감효령
化学发光测定法%C反应蛋白质%偏倚%试剂评价
化學髮光測定法%C反應蛋白質%偏倚%試劑評價
화학발광측정법%C반응단백질%편의%시제평개
chemiluminescent measurements%C-reactive protein%bias%drug evaluation
目的:评价快速化学发光超敏C反应蛋白(hs-CRP)定量检测试剂盒的性能。方法按照美国临床实验室标准化委员会(NCCLS )EP10-A2文件,采用快速化学发光hs-CRP定量检测试剂盒对CRP低、中、高值质控血清进行检测,计算其偏倚、总不精密度及其斜率、截距、非线性、携带污染率和漂移性,并判断其临床可接受性。结果快速化学发光hs-CRP定量检测试剂盒的偏倚和总不精密度均在允许范围内,其斜率、截距、携带污染率、非线性、漂移性平均值分别为1.0057、0.5378、0.7896%、0.0192、0.0360,差异无统计学意义(P>0.05)。结论快速化学发光hs-CRP试剂盒准确度和精密度良好,性能稳定,符合临床检测要求。
目的:評價快速化學髮光超敏C反應蛋白(hs-CRP)定量檢測試劑盒的性能。方法按照美國臨床實驗室標準化委員會(NCCLS )EP10-A2文件,採用快速化學髮光hs-CRP定量檢測試劑盒對CRP低、中、高值質控血清進行檢測,計算其偏倚、總不精密度及其斜率、截距、非線性、攜帶汙染率和漂移性,併判斷其臨床可接受性。結果快速化學髮光hs-CRP定量檢測試劑盒的偏倚和總不精密度均在允許範圍內,其斜率、截距、攜帶汙染率、非線性、漂移性平均值分彆為1.0057、0.5378、0.7896%、0.0192、0.0360,差異無統計學意義(P>0.05)。結論快速化學髮光hs-CRP試劑盒準確度和精密度良好,性能穩定,符閤臨床檢測要求。
목적:평개쾌속화학발광초민C반응단백(hs-CRP)정량검측시제합적성능。방법안조미국림상실험실표준화위원회(NCCLS )EP10-A2문건,채용쾌속화학발광hs-CRP정량검측시제합대CRP저、중、고치질공혈청진행검측,계산기편의、총불정밀도급기사솔、절거、비선성、휴대오염솔화표이성,병판단기림상가접수성。결과쾌속화학발광hs-CRP정량검측시제합적편의화총불정밀도균재윤허범위내,기사솔、절거、휴대오염솔、비선성、표이성평균치분별위1.0057、0.5378、0.7896%、0.0192、0.0360,차이무통계학의의(P>0.05)。결론쾌속화학발광hs-CRP시제합준학도화정밀도량호,성능은정,부합림상검측요구。
Objective To evaluate the performance of the hypersensitive C-reactive protein(hs-CRP) rapid quantitative chemilu-minescent detection kit .Methods According to National Committee for Clinical Laboratory Standards (NCCLS) EP10-A2 docu-ment ,hs-CRP rapid quantitative chemiluminescent detection kit was employed to measure the CRP at low ,medium and high con-centration levels of quality control serum .Bias ,total imprecision and their slope rates ,intercepts ,carryover ,non-linearity and drift were calculated ,and its clinical acceptability was evaluated .Results Bias and total imprecision of hs-CRP rapid quantitative chemi-luminescent detection kit were within the allowable ranges ,the average values of slope rates ,intercepts ,carryover ,non-linearity and drift were 1 .005 7 ,0 .537 8 ,0 .789 6% ,0 .019 2 ,0 .036 0 ,respectively ,the differences showed no statistically significance ( P>0 .05) .Conclusion hs-CRP rapid quantitative chemiluminescent detection kit has good accuracy and precision ,stable perform-ance ,and consistent with the clinical testing requirements .