目的 应用前节光学相干断层扫描仪(OCT),比较WaveLight FS200飞秒激光与Moria M2 90角膜板层切开刀制作角膜瓣的厚度及形态特点.方法 前瞻性病例对照研究.对61例行双眼准分子激光原位角膜磨镶术(LASIK)的近视患者分别采用WaveLight FS200飞秒激光制瓣(WaveLight组,36例)和Moria M2 90角膜板层切开刀制瓣(Moria组,25例).术后1周,采用Visante前节OCT测量180°、90°、45°、135°轴位上距离角膜顶点0、±2、±3.5 mm处的角膜瓣厚度.运用独立样本t检验对2组数据进行统计学分析.结果 WaveLight组中央角膜瓣厚度为(105.53 ±5.86) μm,而Moria组为(132.96±13.91)μm(t=-13.16,P<0.01).WaveLight组中央角膜瓣精确度为(6.17±3.98)μm,而Moria组为(23.60±12.64)μm (t=-15.20,P<0.01).WaveLight组角膜瓣均匀性指数为(4.07±0.94)μm,Moria组为(8.48±2.35) μm (t=-12.60,P<0.01).WaveLight组角膜瓣的对称性指数[(33.81±11.28)μm]明显小于Moria组[(68.46±26.66)μm],差异有统计学意义(t=-8.67,P<0.01).结论 前节OCT显示WaveLight FS200飞秒激光制作的角膜瓣与Moria M2 90角膜板层切开刀相比,中央厚度更精确、重复性更好,角膜瓣形态更均匀.
目的 應用前節光學相榦斷層掃描儀(OCT),比較WaveLight FS200飛秒激光與Moria M2 90角膜闆層切開刀製作角膜瓣的厚度及形態特點.方法 前瞻性病例對照研究.對61例行雙眼準分子激光原位角膜磨鑲術(LASIK)的近視患者分彆採用WaveLight FS200飛秒激光製瓣(WaveLight組,36例)和Moria M2 90角膜闆層切開刀製瓣(Moria組,25例).術後1週,採用Visante前節OCT測量180°、90°、45°、135°軸位上距離角膜頂點0、±2、±3.5 mm處的角膜瓣厚度.運用獨立樣本t檢驗對2組數據進行統計學分析.結果 WaveLight組中央角膜瓣厚度為(105.53 ±5.86) μm,而Moria組為(132.96±13.91)μm(t=-13.16,P<0.01).WaveLight組中央角膜瓣精確度為(6.17±3.98)μm,而Moria組為(23.60±12.64)μm (t=-15.20,P<0.01).WaveLight組角膜瓣均勻性指數為(4.07±0.94)μm,Moria組為(8.48±2.35) μm (t=-12.60,P<0.01).WaveLight組角膜瓣的對稱性指數[(33.81±11.28)μm]明顯小于Moria組[(68.46±26.66)μm],差異有統計學意義(t=-8.67,P<0.01).結論 前節OCT顯示WaveLight FS200飛秒激光製作的角膜瓣與Moria M2 90角膜闆層切開刀相比,中央厚度更精確、重複性更好,角膜瓣形態更均勻.
목적 응용전절광학상간단층소묘의(OCT),비교WaveLight FS200비초격광여Moria M2 90각막판층절개도제작각막판적후도급형태특점.방법 전첨성병례대조연구.대61례행쌍안준분자격광원위각막마양술(LASIK)적근시환자분별채용WaveLight FS200비초격광제판(WaveLight조,36례)화Moria M2 90각막판층절개도제판(Moria조,25례).술후1주,채용Visante전절OCT측량180°、90°、45°、135°축위상거리각막정점0、±2、±3.5 mm처적각막판후도.운용독립양본t검험대2조수거진행통계학분석.결과 WaveLight조중앙각막판후도위(105.53 ±5.86) μm,이Moria조위(132.96±13.91)μm(t=-13.16,P<0.01).WaveLight조중앙각막판정학도위(6.17±3.98)μm,이Moria조위(23.60±12.64)μm (t=-15.20,P<0.01).WaveLight조각막판균균성지수위(4.07±0.94)μm,Moria조위(8.48±2.35) μm (t=-12.60,P<0.01).WaveLight조각막판적대칭성지수[(33.81±11.28)μm]명현소우Moria조[(68.46±26.66)μm],차이유통계학의의(t=-8.67,P<0.01).결론 전절OCT현시WaveLight FS200비초격광제작적각막판여Moria M2 90각막판층절개도상비,중앙후도경정학、중복성경호,각막판형태경균균.
Objective To evaluate and compare the thickness and morphology of laser in situ keratomileusis (LASIK) flaps created by either the WaveLight FS-200 femtosecond laser or the Moria microkeratome M2 90 using anterior segment optical coherence tomography (AS-OCT).Methods This was a prospective case-control study.Bilateral LASIK was performed on 61 myopic patients.Flaps were created using the WaveLight FS-200 femtosecond laser in 72 eyes and the Moria M2 90 microkeratome in 50 eyes.The thickness at 17 points across each flap,0 mm,±2 mm,and ±3.5 mm to the corneal vertex on the horizontal,vertical,45° and 135° meridians,were evaluated using a Visante AS-OCT at 1 week postoperatively.An independent t test was used.Results The central flap thickness of the WaveLight group was 105.53±5.86 μm,while it was 132.96±13.91 μm in the Moria group (t=-13.16,P<0.01).The difference between the achieved and the intended flap thickness (accuracy) was 6.17±3.98 μm and 23.60±12.64 μm,respectively (t=-15.20,P<0.01).The standard deviation of thickness in the 17 points within an individual flap (uniformity) was 4.07±0.94 μm in the WaveLight group and 8.48±2.35 μm in the Moria group (t=-12.60,P<0.01).The index S (Symmetry) of Moria group (33.81±11.28 μm) was significantly smaller in the WaveLight group (68.46±26.66 μm) (t=-8.67,P<0.01).Conclusion AS-OCT showed that the flaps created with the WaveLight FS-200 femtosecond laser were more accurate,reproducible,and uniform than those created with the Moria M2 90 microkeratome.