目的:通过对同一项目、不同检测方法的测定结果进行分析比对,探讨不同尿镉分析方法存在的误差是否在允许范围内,实现检测结果互认的可比性。方法:用石墨炉原子吸收光谱法为标准方法,微分电位溶出法为参比法,选取3份样品在原子吸收分光光度计和溶出仪上进行检测,同时带有质控样。分析比对实验所得的数据,计算方法间的相关性和相对偏差。结果:两种方法的分析误差均<3.69%,加标回收率均在95.3%~104.5%,在误差允许范围之内。在各自的试验条件下质控样也均在允许的范围之内,经过计算当t值在:-0.627~2.89时,P值在0.715~0.914。结论:通过比对实验数据分析,结果一致性良好,可以满足实验的要求,溶出仪法更适合大批量的高浓度的检测。
目的:通過對同一項目、不同檢測方法的測定結果進行分析比對,探討不同尿鎘分析方法存在的誤差是否在允許範圍內,實現檢測結果互認的可比性。方法:用石墨爐原子吸收光譜法為標準方法,微分電位溶齣法為參比法,選取3份樣品在原子吸收分光光度計和溶齣儀上進行檢測,同時帶有質控樣。分析比對實驗所得的數據,計算方法間的相關性和相對偏差。結果:兩種方法的分析誤差均<3.69%,加標迴收率均在95.3%~104.5%,在誤差允許範圍之內。在各自的試驗條件下質控樣也均在允許的範圍之內,經過計算噹t值在:-0.627~2.89時,P值在0.715~0.914。結論:通過比對實驗數據分析,結果一緻性良好,可以滿足實驗的要求,溶齣儀法更適閤大批量的高濃度的檢測。
목적:통과대동일항목、불동검측방법적측정결과진행분석비대,탐토불동뇨력분석방법존재적오차시부재윤허범위내,실현검측결과호인적가비성。방법:용석묵로원자흡수광보법위표준방법,미분전위용출법위삼비법,선취3빈양품재원자흡수분광광도계화용출의상진행검측,동시대유질공양。분석비대실험소득적수거,계산방법간적상관성화상대편차。결과:량충방법적분석오차균<3.69%,가표회수솔균재95.3%~104.5%,재오차윤허범위지내。재각자적시험조건하질공양야균재윤허적범위지내,경과계산당t치재:-0.627~2.89시,P치재0.715~0.914。결론:통과비대실험수거분석,결과일치성량호,가이만족실험적요구,용출의법경괄합대비량적고농도적검측。
Objective:To discuss whether the error of different analysis method of the urine cadmium has within the scope of the permit,through a comparision between different detection methods on the same research object. In order to achieve the mutual recognition of test results comparable. Method:Graphite furnace atomic absorption spectrometry was used as the standard method which was compared by the differential potentiometric stripping method. Three samples were tested by the atomic absorption spectrophotometer and differential potentiometric stripping device and the quality control samples was prepared at the same time. Then the data was analyzed and the correlation and relative deviation were calculated. Result:The analytical errors of both methods were less than 3.69%,the standard recovery rate both locate within 95.3%-104.5%, all locate within the range of allowable error. The value of the Quality control samples under each experimental condition also accord with the range of allowable error. After calculation,when t locates within-0.627-2.89,P locates within 0.715-0.914. Conclusion:The results of the experiment show that the results consistency of the two methods are favorable,which can meet the test requirements. The differential potentiometric stripping method is more suitable treating large quantities of high concentration of detection.