研究了一种极紫外波段微通道板(MCP)光子计数探测器,用于探测地球等离子体层中极微弱的30.4 nm辐射。通过改变电压、温度等参数对比了该微通道板光子计数探测器的暗噪声和分辨率的变化。结果表明:微通道板探测器的暗噪声主要来源于残余气体离子反馈和热噪声,因此要降低探测器暗噪声,应对微通道进行彻底的预处理除气,并尽量避免探测器在高温状态下工作,常温下经过预处理的微通道板光子探测系统的暗计数率仅为0.34 count/(s.cm2)。系统的分辨率主要受电压和计数率的影响,受温度影响不明显。由于不同的微通道板有不同的耐压范围,过小或过大的电压或计数率都会造成系统分辨率的降低。
研究瞭一種極紫外波段微通道闆(MCP)光子計數探測器,用于探測地毬等離子體層中極微弱的30.4 nm輻射。通過改變電壓、溫度等參數對比瞭該微通道闆光子計數探測器的暗譟聲和分辨率的變化。結果錶明:微通道闆探測器的暗譟聲主要來源于殘餘氣體離子反饋和熱譟聲,因此要降低探測器暗譟聲,應對微通道進行徹底的預處理除氣,併儘量避免探測器在高溫狀態下工作,常溫下經過預處理的微通道闆光子探測繫統的暗計數率僅為0.34 count/(s.cm2)。繫統的分辨率主要受電壓和計數率的影響,受溫度影響不明顯。由于不同的微通道闆有不同的耐壓範圍,過小或過大的電壓或計數率都會造成繫統分辨率的降低。
연구료일충겁자외파단미통도판(MCP)광자계수탐측기,용우탐측지구등리자체층중겁미약적30.4 nm복사。통과개변전압、온도등삼수대비료해미통도판광자계수탐측기적암조성화분변솔적변화。결과표명:미통도판탐측기적암조성주요래원우잔여기체리자반궤화열조성,인차요강저탐측기암조성,응대미통도진행철저적예처리제기,병진량피면탐측기재고온상태하공작,상온하경과예처리적미통도판광자탐측계통적암계수솔부위0.34 count/(s.cm2)。계통적분변솔주요수전압화계수솔적영향,수온도영향불명현。유우불동적미통도판유불동적내압범위,과소혹과대적전압혹계수솔도회조성계통분변솔적강저。
A Microchannel Plate(MCP) photon detector was developed to detect the 30.4 nm radiation in a plasma layer of the earth.By changing voltages,temperatures and other parameters,the variations of the dark noise and resolution for the MCP detector were compared.The results indicate that the dark noise of the detector comes from the ion feedback of residual gas and the thermal noises mainly.Therefore,to reduce the dark noise,the microchannel should be preprocessed by removing residual gases and the detector could not work at a high temperature state.After preprocessing at the room temperature,the dark noise counting rate of the detector has reduced to 0.34 count/s·cm2.Furthermore,the temperature has a little effect on the resolution of the detector,and the resolution dependents on the variances of voltage or counting rate obviously.With different resist pressure abilities for different MCPs,it demonstrates that the proper voltage or counting rate can provide a better resolution.