IOL Master%A超%硅油眼%眼轴%等效球镜度数
IOL Master%A超%硅油眼%眼軸%等效毬鏡度數
IOL Master%A초%규유안%안축%등효구경도수
IOL-Master%contact A-scan%silicon oil-filled eyes%axial length%spherical equivalent
目的:比较评估A超及IOL Master在硅油眼患者眼轴长度测量的精确性,探讨IOL Master在硅油眼人工晶状体度数计算中的临床应用价值。方法:分析我院2012-06/12间收治的硅油眼白内障手术患者30例30眼,分别用接触式A超及IOL Master测量术前、术后眼轴长度比较两者之间差异。并将术前 IOL Master测量自动获取IOL度数后预估患眼屈光状态与术后1 mo复查患者屈光状态(等效球镜度数)行统计学分析。结果:A超与术前IOL Master 测量眼轴有显著性差异,使用IOL Master测量术前、术后眼轴无显著差异,术后使用A超及IOL Master测量眼轴之间无显著性差异。采用IOL Master术前测量眼轴所得人工晶状体度数植入人工晶状体术后屈光状态与术前预估无明显差异。结论:IOL Master测量眼轴较A超有明显优越性,但对于某些不能使用IOL Master测量的病例仍需结合A超等其他测量方法辅助测量。
目的:比較評估A超及IOL Master在硅油眼患者眼軸長度測量的精確性,探討IOL Master在硅油眼人工晶狀體度數計算中的臨床應用價值。方法:分析我院2012-06/12間收治的硅油眼白內障手術患者30例30眼,分彆用接觸式A超及IOL Master測量術前、術後眼軸長度比較兩者之間差異。併將術前 IOL Master測量自動穫取IOL度數後預估患眼屈光狀態與術後1 mo複查患者屈光狀態(等效毬鏡度數)行統計學分析。結果:A超與術前IOL Master 測量眼軸有顯著性差異,使用IOL Master測量術前、術後眼軸無顯著差異,術後使用A超及IOL Master測量眼軸之間無顯著性差異。採用IOL Master術前測量眼軸所得人工晶狀體度數植入人工晶狀體術後屈光狀態與術前預估無明顯差異。結論:IOL Master測量眼軸較A超有明顯優越性,但對于某些不能使用IOL Master測量的病例仍需結閤A超等其他測量方法輔助測量。
목적:비교평고A초급IOL Master재규유안환자안축장도측량적정학성,탐토IOL Master재규유안인공정상체도수계산중적림상응용개치。방법:분석아원2012-06/12간수치적규유안백내장수술환자30례30안,분별용접촉식A초급IOL Master측량술전、술후안축장도비교량자지간차이。병장술전 IOL Master측량자동획취IOL도수후예고환안굴광상태여술후1 mo복사환자굴광상태(등효구경도수)행통계학분석。결과:A초여술전IOL Master 측량안축유현저성차이,사용IOL Master측량술전、술후안축무현저차이,술후사용A초급IOL Master측량안축지간무현저성차이。채용IOL Master술전측량안축소득인공정상체도수식입인공정상체술후굴광상태여술전예고무명현차이。결론:IOL Master측량안축교A초유명현우월성,단대우모사불능사용IOL Master측량적병례잉수결합A초등기타측량방법보조측량。
AIM: To evaluate and compare the accuracy of IOL-Master and A-scan in measuring the axial length of silicone oil-filled eyes, and to explore the clinical value of IOL-Master in calculating intraocular lens diopter of the silicone oil-filled eyes. METHODS: An analysis was made of 30 patients ( 30 eyes ) who received cataract surgery combined with silicone oil removal in our hospital from June 2012 to December 2012. Before the operation and one month afterwards, the axial length was measured by both IOL-Master and A-scan, and all the results were divided into two groups:A-scan group and IOL-Master group.Then we used IOL-Master to calculate the intraocular lens diopter and to estimate the postoperative refraction. Postoperative refraction was measured one month after operation and compared with the preoperative estimates. RESULTS: Before silicon oil removal, there was significant difference between the axial lengths obtained by IOL-Master and those by A-scan, with the former much smaller than the latter.After the operation, there was no significant difference between the axial lengths obtained by the two methods.The axial lengths measured by IOL -Master before and after operation were not significantly different. No significant difference was observed between the postoperative refraction and the preoperative estimates, either. CONCLUSION:IOL-Master is a more preferable method to measure axial length than A-scan.However, in cases where IOL-Master cannot be applied, A-scan together with other methods should be used.