海南医学
海南醫學
해남의학
HAINAN MEDICAL JOURNAL
2014年
15期
2245-2246
,共2页
自动血沉仪%魏氏法%红细胞沉降率
自動血沉儀%魏氏法%紅細胞沉降率
자동혈침의%위씨법%홍세포침강솔
Automated ESR Analyzer%Weishi method%Erythrocyte sedimentation rate
目的:评价ESR-2040动态血沉分析仪检测红细胞沉降率(ESR)的重复性、准确性。方法随机抽取160例住院患者静脉血样本,每个样本均采用ESR-2040动态血沉分析仪及魏氏法同时测定ESR,同时抽取ESR测定值从低到高的标本5例,采用ESR-2040动态血沉分析仪重复测定10次,并对测定的结果进行统计分析。结果 ESR-2040动态血沉分析仪检测红细胞沉降率具有较好的重复性(CV<5%),与魏氏法具有良好的相关性。两种方法的测定结果在正常值范围内结果基本一致;血沉结果大于20 mm/h时,魏氏法测定的结果高于仪器法,并且随着测定值的升高二者之间的差异越显著。结论自动血沉仪与魏氏法血沉测定结果在20 mm/h以内时有良好的相关性和准确性,>20 mm/h时应用魏氏法复检以提高检测值的准确性。
目的:評價ESR-2040動態血沉分析儀檢測紅細胞沉降率(ESR)的重複性、準確性。方法隨機抽取160例住院患者靜脈血樣本,每箇樣本均採用ESR-2040動態血沉分析儀及魏氏法同時測定ESR,同時抽取ESR測定值從低到高的標本5例,採用ESR-2040動態血沉分析儀重複測定10次,併對測定的結果進行統計分析。結果 ESR-2040動態血沉分析儀檢測紅細胞沉降率具有較好的重複性(CV<5%),與魏氏法具有良好的相關性。兩種方法的測定結果在正常值範圍內結果基本一緻;血沉結果大于20 mm/h時,魏氏法測定的結果高于儀器法,併且隨著測定值的升高二者之間的差異越顯著。結論自動血沉儀與魏氏法血沉測定結果在20 mm/h以內時有良好的相關性和準確性,>20 mm/h時應用魏氏法複檢以提高檢測值的準確性。
목적:평개ESR-2040동태혈침분석의검측홍세포침강솔(ESR)적중복성、준학성。방법수궤추취160례주원환자정맥혈양본,매개양본균채용ESR-2040동태혈침분석의급위씨법동시측정ESR,동시추취ESR측정치종저도고적표본5례,채용ESR-2040동태혈침분석의중복측정10차,병대측정적결과진행통계분석。결과 ESR-2040동태혈침분석의검측홍세포침강솔구유교호적중복성(CV<5%),여위씨법구유량호적상관성。량충방법적측정결과재정상치범위내결과기본일치;혈침결과대우20 mm/h시,위씨법측정적결과고우의기법,병차수착측정치적승고이자지간적차이월현저。결론자동혈침의여위씨법혈침측정결과재20 mm/h이내시유량호적상관성화준학성,>20 mm/h시응용위씨법복검이제고검측치적준학성。
Objective To evaluate the repeatability and accuracy of 2040 Automated Erythrocyte Sedimenta-tion Rate (ESR) Analyzer. Methods Using ERS-2040 Automated ESR Analyzer and traditiona1 Weishi method to measure 160 random venous blood samples from hospitalized patient respectively, and extract five samples that the low-er to higher from the results of ESR were repeated measurement (n=10) with ERS-2040 Automated ESR Analyzer. The results of ESR were statistically analyzed. Results In detecting ESR, it showed that the ESR-2040 analyzer had a good repeatability (CV<5%), The method of ESR-2040 analyzer, compared with traditional Weishi method, showed a satisfac-tory correlation. At the normal range of ESR, there were no significant difference between the results of ESR from two methods. However, when the results were beyond the range of 20 mm/h,the results from Weishi method were higher than those from analyzer, and the higher the results, the more significant the difference. Conclusion At ESR within 20 mm/h, there were significant correlation and accuracy between ESR-2040 analyzer with traditional Weishi method. At ESR above 20 mm/h, we need to use the traditional Weishi method detecting recheck to improve accuracy of results.